Skip to content

Comprehensive® Reverse Shoulder 180561Recalls & reported problems

Biomet Orthopedics, LLC · Class 2

All models of Comprehensive Reverse Shoulder (57)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00880304677289), received between Nov 5, 2019 and Nov 5, 2024 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

19 other recalls cover devices cleared under the same FDA submission (K080642, K130390, K132239); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Adverse Event Without Identified Device or Use Problem1
  • Difficult to Insert1
  • Migration or Expulsion of Device1
  • Packaging Problem1
  • Tear, Rip or Hole in Device Packaging1

Patient problems

  • Failure of Implant3
  • No Clinical Signs, Symptoms or Conditions2
  • Pain2
  • Bacterial Infection1
  • Foreign Body In Patient1
  • Loss of Range of Motion1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 5, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Failure of Implant

    MAUDE report 20611586 ↗
  • InjuryOct 8, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Failure of Implant; Loss of Range of Motion; Pain

    MAUDE report 20397692 ↗
  • InjuryMar 6, 2023Fracture; Migration or Expulsion of Device; Fracture; Migration or Expulsion of Device

    IT WAS REPORTED THAT THE PATIENT HAD AN INITIAL SURGERY APPROXIMATELY 3 YEARS AGO, SUBSEQUENTLY THEY THEN HAD A REVISION FOR THE REMOVAL OF THE GLENOSPHERE IMPLANT AND BASEPLATE PULLOUT. DURING THE REVISION THERE WAS A BROKEN SCREW THAT WAS PARTIALLY REMOVED, HOWEVER SOME THE SCREW REMAINED IN THE GLENOID. ALSO, THE BACTERIOLOGY WAS POSITIVE FOR CUTI ACNES, BUT THERE WERE NO INTRAOPERATIVE SEPTIC SIGNS, NOR PAIN BEFORE THE BREAKAGE OF THE MATERIAL. ATTEMPTS HAVE BEEN MADE AND THERE IS NO FURTHER INFORMATION AT THIS TIME.

    Patient problems: Bacterial Infection; Failure of Implant; Foreign Body In Patient

    MAUDE report 16487226 ↗
  • MalfunctionAug 4, 2022Tear, Rip or Hole in Device Packaging; Packaging Problem; Tear, Rip or Hole in Device Packaging; Packaging Problem

    IT WAS REPORTED THAT A DEVICE WAS RETURNED WITH PACKAGING DAMAGE AND THE STERILITY BARRIER POTENTIALLY COMPROMISED. THERE IS NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15163111 ↗
  • MalfunctionMay 20, 2022Fracture; Difficult to Insert; Fracture; Difficult to Insert

    IT WAS REPORTED WHEN INSERTING THE INFERIOR SCREW, THE SCREW COULD NOT BE FULLY COUNTERSUNK. REMOVING THE SCREW WAS ALSO DIFFICULT. PLIERS FROM THE OPERATING ROOM HAD TO BE USED BECAUSE THE SCREWDRIVERS DID NOT GRIP THE SCREW HEAD. A DEFECT IN THE SCREWDRIVER COULD HAVE BEEN THE CAUSE. WHEN INSERTING THE SUPERIOR SCREW, THE SCREW HEAD GOT BROKEN. THE REST OF THE SCREW WAS LEFT IN PLACE AND THE REMAINING 3 PERIPHERAL SCREW HOLES WERE FILLED. ALL SCREW HOLES WERE ADEQUATE IN DEPTH. THE SCREW WAS RETURNED FRACTURED. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THIS MALFUNCTION, ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14462277 ↗
  • InjuryNov 5, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Pain

    MAUDE report 9279476 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K080642, K130390, K132239. One submission often covers many models, so check the product and lot details in each record.

  • Z-1131-2021 · Jan 15, 2021 · Terminated
    Central Screw 6.5x25mm - Product Usage: used as a sterile screw that is intended for implantable shoulder prosthesis assembly and/or fixation.
  • Z-0619-2021 · Nov 25, 2020 · Terminated
    Comprehensive Reverse Shoulder System, Item numbers 115394, 115395, 115396, 115398, 180550, 180551, 180552, 180553, 180554, 180555, 180556, 180557, 180558, 180559, 180560, 1805
  • Z-2144-2020 · Apr 17, 2020 · Terminated
    Extremities Item Number: 1)113954 Hybrid Glenoid Glenoid Base, 4 MM 2) XL-115366 Comprehensive Reverse Shoulder Humeral Bearing, ArComXL, 44 MM X 41 MM, Standard 3) XL-115364 Comp
  • Z-1750-2020 · Mar 13, 2020 · Terminated
    Comprehensive Humeral Tray Taper Extraction Pliers - Replacement Tips, Model Number 110028522 - Product Usage: Intended to be used in surgery to relieve pain and to replace the gle
  • Z-1859-2017 · Feb 17, 2017 · Terminated
    various polyethylene implants Extremities, Shoulders and Elbows BioModular Shoulder System; Comprehensive Shoulder System; Comprehensive Reverse Shoulder System Product Usage: Sh
  • Z-1103-2017 · Dec 20, 2016 · Terminated
    Comprehensive Reverse Shoulder System Humeral Tray Model 115340
  • Z-1548-2013 · May 15, 2013 · Terminated
    Superior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain.
  • Z-2546-2011 · Apr 22, 2011 · Terminated
    Custom bi-angular to reverse shoulder humeral tray 44 mm sterile, REF CP561548, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder products are indicated fo
  • Z-2550-2011 · Apr 22, 2011 · Terminated
    Custom comprehensive reverse shoulder bio-modular base 44 mm humeral tray standard, REF CP561510, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder product
  • Z-2549-2011 · Apr 22, 2011 · Terminated
    Custom comprehensive reverse humeral tray set w/ lock rings 44 mm + 15 mm & + 18 mm sterile, REF CP561705, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder
  • Z-2543-2011 · Apr 22, 2011 · Terminated
    Comprehensive reverse shoulder humeral tray with locking ring, 44 mm +10 mm, sterile, single use, REF 115348, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Should
  • Z-2545-2011 · Apr 22, 2011 · Terminated
    Comprehensive reverse shoulder humeral tray with locking ring, 44 mm standard, sterile, single use, REF 115340, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Sho
  • Z-2541-2011 · Apr 22, 2011 · Terminated
    Bio-modular reverse shoulder humeral tray with lock ring, 44 mm 10 degree angle, sterile, single use, REF 11-113685, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse S
  • Z-2548-2011 · Apr 22, 2011 · Terminated
    Custom biomodular humeral tray 44 mm sterile, REF CP561686, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder products are indicated for use in patients who
  • Z-2542-2011 · Apr 22, 2011 · Terminated
    Bio-modular reverse shoulder humeral tray with lock ring, 44 mm standard, sterile, single use, REF 11-113680, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder

…and 4 more.

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page