Impress® 00884450009437Recalls & reported problems
Merit Medical Systems, Inc. · Class 2
All models of Impress (145)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00884450009437), received between May 6, 2021 and Jun 1, 2022 — classified as 5 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture6
- Material Separation4
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Foreign Body In Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryJun 1, 2022Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT DURING AN ENDOVASCULAR PROCEDURE, THE CATHETER TIP DETACHED WITHIN THE PATIENT. THE PHYSICIAN HAD ACQUIRED VASCULAR ACCESS AND DURING CATHETER MANIPULATIONS OVER A GUIDEWIRE UNDER FLUOROSCOPY THE CATHETER TIP DETACHED. THE PHYSICIAN WAS ABLE TO SUCCESSFULLY EXTERNALIZED THE FOREIGN BODY FROM THE PATIENT WITH A VASCULAR SNARE DEVICE THUS LIBERATING THE VESSEL. THE PATIENT TOLERATED THE PROCEDURE WELL. NO ADDITIONAL PATIENT CONSEQUENCES TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14569837 ↗InjuryJun 1, 2022Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT DURING AN ENDOVASCULAR PROCEDURE, THE CATHETER TIP DETACHED WITHIN THE PATIENT. THE PHYSICIAN HAD ACQUIRED VASCULAR ACCESS AND DURING CATHETER MANIPULATIONS UNDER FLUOROSCOPY THE CATHETER TIP DETACHED AND MIGRATED INTO THE PATIENTS LUNG. THE PATIENT WAS TRANSFERRED TO IR DEPARTMENT FOR FOREIGN BODY REMOVAL. THE PATIENT TOLERATED THE PROCEDURE WELL. NO ADDITIONAL PATIENT CONSEQUENCES TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14569801 ↗InjuryJul 13, 2021Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT DURING THE PROCEDURE THE TIP OF THE IMPRESS CATHETER BROKE OFF. THE TIP HAD TO BE RETRIEVED USING A SNARE DEVICE.
Patient problems: Foreign Body In Patient
MAUDE report 12160296 ↗InjuryJul 13, 2021Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT DURING THE PROCEDURE THE TIP OF THE IMPRESS CATHETER BROKE OFF. NO PATIENT INJURY TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12160294 ↗MalfunctionMay 13, 2021Fracture; Fracture
THE ACCOUNT ALLEGES WHILE MANIPULATING THE CATHETER WITHIN THE PATIENT'S BLADDER, THE TIP OF THE DEVICE STARTED TO SHEAR AWAY FROM THE REST OF THE DEVICE. THE CATHETER WAS REMOVED COMPLETELY WITHOUT AN ADDITIONAL PROCEDURE OR INTERVENTIONAL DEVICE REQUIRED. NO PATIENT INJURY TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11816073 ↗InjuryMay 6, 2021Fracture; Fracture
THE ACCOUNT ALLEGES DURING A PROCEDURE THE BLACK RADIOPAQUE TIP OF THE CATHETER DETACHED FROM THE REST OF THE DEVICE. THE PHYSICIAN USED AN ENSNARE DEVICE TO RETRIEVE THE DETACHED TIP. THE PROCEDURE WAS SUCCESSFUL AND THERE ARE NO ADDITIONAL PATIENT CONSEQUENCES TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11787812 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H749085263001Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page