ReSolve® 00884450009482Recalls & reported problems
Merit Medical Systems, Inc. · Class 1
All models of ReSolve (16)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 1
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00884450009482), received between Sep 30, 2020 and Dec 14, 2021 — classified as 1 death, 3 injuries.
The most frequently reported specific device problem is “Crack” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Crack2
- Fracture2
- Material Separation2
Patient problems
- Foreign Body In Patient2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
DeathDec 14, 2021Fracture; Material Separation; Fracture; Material Separation
THE PATIENT WAS ADMITTED AND COMMENCED ON INTRAVENOUS ANTIBIOTICS FOR SUSPECTED CELLULITIS TO BILATERAL LEGS WITH REFERRAL FOR INSERTION OF ASCITIC DRAIN (ABDOMINAL CAVITY) TO REDUCE FLUID RETENTION. DURING THE DRAINAGE TUBE PLACEMENT PROCEDURE, AN 8-FRENCH GAUGE PIGTAIL CATHETER WAS SUCCESSFULLY INSERTED AND SECURED TO THE ANTERIOR ABDOMINAL WALL ON THE RIGHT SIDE IN THE LOWER ILIAC FOSSA. THE CATHETER WAS PLACED UNDER ULTRASOUND GUIDANCE IN THE MEDICAL IMAGING (MI) DEPARTMENT PRIOR TO RETURNING TO THE WARD FOR MONITORING FOR A PLANNED DRAINAGE TIME OF SIX HOURS PRIOR TO REMOVAL (SCHEDULED FOR 20:30 HOURS). THE PATIENT WAS RETURNED TO THE WARD FOR OBSERVATION WITHOUT NOTED COMPLICATIONS FOLLOWING THE PROCEDURE. DURING THE DRAINAGE TUBE REMOVAL PROCEDURE, THE MEDICAL STAFF REPORTED THAT THEY WERE FAMILIAR WITH THIS TYPE OF DRAINAGE CATHETER AND HOW TO REMOVE THE DEVICE FROM THE PATIENT. WHEN THE HUB WAS CUT OFF, A LARGE AMOUNT OF FLUID GUSHED OUT WHICH STARTLED THE NURSE, RESULTING IN LOSS OF GRIP ON THE DRAINAGE CATHETER. THE PATIENT INHALED AT THE SAME TIME, AND THE DRAINAGE CATHETER WITHDREW INTO THE PATIENT'S ABDOMEN RESULTING IN THE UNINTENDED RETENTION. THE HUB WAS NOT KEPT F
Patient problems: Foreign Body In Patient
MAUDE report 13005428 ↗InjuryOct 5, 2021Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT THEY TRIED TO EXCHANGE THE 8 FRENCH NEPHROSTOMY TUBE, HOWEVER THE TIP OF THE CATHETER BECAME STUCK ON THE KIDNEY CAPSULE AND A SMALL PIECE WAS RETAINED WITHIN THE TRACT NEAR THE KIDNEY. DESPITE MULTIPLE ATTEMPTS, THE RETAINED PRODUCT COULD NOT BE RETRIEVED FLUOROSCOPICALLY. THE PATIENT IS SCHEDULED TO HAVE UROLOGY REMOVE IT SURGICALLY.
Patient problems: Foreign Body In Patient
MAUDE report 12576875 ↗InjurySep 30, 2020Crack; Crack
THE ACCOUNT ALLEGES THAT POST-DRAINAGE CATHETER PLACEMENT WITHIN THE PATIENT'S BILIARY SYSTEM, THE DRAINAGE CATHETER BECAME DAMAGED AT THE BEND 15 DAYS LATER AND WOULD NOT CONTINUE TO DRAIN. THE DRAINAGE CATHETER WAS REPLACED FOR THIS PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10609999 ↗InjurySep 30, 2020Crack; Crack
THE ACCOUNT ALLEGES THAT POST-DRAINAGE CATHETER PLACEMENT WITHIN THE PATIENT'S BILIARY SYSTEM, THE DRAINAGE CATHETER BECAME DAMAGED AT THE BEND 7 DAYS LATER AND WOULD NOT CONTINUE TO DRAIN. THE DRAINAGE CATHETER WAS REPLACED FOR THIS PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10609993 ↗
Other CATHETER, IRRIGATION devices
- Navarre Opti-Drain NOD8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION901 reports
- Navarre NNU8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION736 reports
- BLAKE 2227Ethicon Inc. · CATHETER, IRRIGATION331 reports
- Navarre NNU6LPTBard Access Systems, Inc. · CATHETER, IRRIGATION211 reports
- Navarre Opti-Drain NOD10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION178 reports
- Navarre NNU10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION127 reports
- BLAKE 2233Ethicon Inc. · CATHETER, IRRIGATION114 reports
- Vasoshield VH-5001Maquet Cardiovascular LLC · Catheter, Irrigation1 recall27 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page