Impress® 00884450020098Recalls & reported problems
Merit Medical Systems, Inc. · Class 2
All models of Impress (145)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00884450020098), received between May 12, 2021 and Feb 21, 2023 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture4
- Material Separation4
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Foreign Body In Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 21, 2023Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT THE BLACK PORTION AT THE TIP OF THE CATHETER GOT DETACHED WITHIN THE VESSEL. AS THE TIP WAS ALONG A GUIDEWIRE, IT COULD BE SUCCESSFULLY WITHDRAWN OUT FROM THE VESSEL. NO FRAGMENT IS REMAINING IN THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16415038 ↗InjuryDec 20, 2022Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT WHILE USING THE ANGIOGRAPHY CATHETER, IT FRACTURED WITHIN THE PATIENT'S CAROTID ARTERY. THE PHYSICIAN CUT DOWN THE ARTERY TO REMOVE THE FOREIGN BODY.
Patient problems: Foreign Body In Patient
MAUDE report 16018634 ↗InjuryAug 9, 2022Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT DURING A SELECTIVE PERIPHERAL ANGIOGRAM PROCEDURE, THE CATHETER TIP DETACHED WITHIN THE PATIENT'S SUPERFICIAL FEMORAL ARTERY [SFA]. THE PHYSICIAN HAD ACQUIRED RETROGRADE FEMORAL ARTERY ACCESS AND HAD NEGOTIATED THE PATIENT ILIAC BIFURCATION TO ACCESS THE PATIENTS CONTRALATERAL SFA. WHILE MANIPULATING THE GUIDE CATHETER OVER AN INTERVENTIONAL GUIDEWIRE, UNDER FLUOROSCOPY, THE CATHETER TIP DETACHED WITHIN THE PATIENTS SFA. SOME VASCULAR CALCIFICATION WAS APPRECIATED. THE PHYSICIAN USED A VASCULAR SNARE DEVICE TO SUCCESSFULLY EXTERNALIZE THE FOREIGN BODY LIBERATING THE VESSEL. NO ADDITIONAL PATIENT CONSEQUENCES TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15196295 ↗InjuryMay 12, 2021Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT DURING A PERIPHERAL VASCULAR PROCEDURE, THE CATHETER TIP DETACHED WITHIN THE PATIENT. THE PHYSICIAN HAD SUCCESSFULLY ACQUIRED RETROGRADE ARTERIAL ACCESS AND DURING CATHETER MANIPULATIONS WITHIN THE VASCULATURE, OVER A GUIDE WIRE, UNDER FLUOROSCOPY THE CATHETER TIP DETACHED. THE FOREIGN BODY WAS SUCCESSFULLY RETRIEVED FORM THE PATIENT WITH VASCULAR SNARE DEVICES. A NEW CATHETER WAS USED TO COMPLETE THE PROCEDURE FOR THE PATIENT. NO ADDITIONAL CONSEQUENCES TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11813232 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page