Skip to content

InQwire® 00884450023563Recalls & reported problems

Merit Medical Systems, Inc. · Class 2

All models of InQwire (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00884450023563), received between Jul 27, 2022 and May 21, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Material Too Rigid or Stiff” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Material Too Rigid or Stiff1

Patient problems

  • Memory Loss/Impairment2
  • Muscle Weakness/Atrophy2
  • Hemorrhage/Blood Loss/Bleeding1
  • Low Blood Pressure/ Hypotension1
  • Speech Disorder1
  • Visual Impairment1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 21, 2025Material Too Rigid or Stiff; Material Too Rigid or Stiff

    THE ACCOUNT ALLEGES THAT DURING A PULMONARY VEIN ISOLATION PROCEDURE WITH THE PATIENT UNDER SEDATION WITH PROPOFOL, A TAMPONADE OCCURRED REQUIRING THORACIC SURGERY. AFTER SUCCESSFUL TRANSSEPTAL PUNCTURE, THE LEFT SUPERIOR AND INFERIOR PULMONARY VEIN WERE ISOLATED. AFTER DEFLATING AND RETRACTING THE BALLOON CATHETER FROM THE LEFT INFERIOR PULMONARY VEIN, THERE WAS A DROP IN BLOOD PRESSURE. A TRANSTHORACIC ECHOGRAPHY REVEALED ACCUMULATING PERICARDIAL BLOOD. WITH A PERICARDIAL PUNCTURE >750ML OF BLOOD WAS WITHDRAWN AND IT WAS DECIDED TO DO EMERGENCY THORACIC SURGERY. THIS REVEALED A 2 CM ANTERO-SUPERIOR TEAR OF THE LEFT INFERIOR PULMONARY VEIN. THE TEAR WAS SUTURED AND THE PATIENT STABILIZED. THE PATIENT WAS MOVED TO THE ICU FOR FURTHER OBSERVATION. IT IS ALLEGED THAT WHILE SWITCHING TO THE LEFT INFERIOR PULMONARY VEIN, THE GUIDE WIRE PROTRUDED OUT OF THE BALLOON CATHETER CAUSING A TEAR. THIS WAS NOT NOTICED AT FIRST AND THE GUIDE WIRE WAS ADVANCED IN THE DIRECTION OF THE LEFT INFERIOR PULMONARY VEIN. THE BALLOON WAS THEN ADVANCED AND INFLATED. AFTER DEFLATING, THE BALLOON NO LONGER BLOCKED THE TEAR AND THE TAMPONADE WAS THEN NOTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Low Blood Pressure/ Hypotension

    MAUDE report 22061204 ↗
  • InjuryJul 27, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE ACCOUNT ALLEGES THAT THE PATIENT WAS EVALUATED AND TREATED FOR CAD. POST PCI PROCEDURE, THE PATIENT COMPLAINED OF LEFT ARM WEAKNESS AND WAS POSITIVELY DIAGNOSED WITH AN EMBOLIC STROKE IN RECOVERY UNIT. THE PATIENT WAS SUCCESSFULLY TREATED AND DISCHARGED TO HOME WITH LOSS OF PERIPHERAL VISION AND WEAKNESS IN THE ARM.

    Patient problems: Muscle Weakness/Atrophy; Visual Impairment

    MAUDE report 15115088 ↗
  • InjuryJul 27, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE ACCOUNT ALLEGES THAT A PATIENT WAS ADMITTED FOR CAD. POST LHC PROCEDURE THE PATIENT WAS EVALUATED IN RECOVERY UNIT. THE PATIENT EXPERIENCED EXPRESSIVE APHASIA AND WAS POSITIVELY DIAGNOSED WITH AN EMBOLIC STROKE. THE PATIENT WAS TREATED AND DISCHARGED TO HOME WITH SLIGHT APHASIA.

    Patient problems: Memory Loss/Impairment; Speech Disorder

    MAUDE report 15115081 ↗
  • InjuryJul 27, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE ACCOUNT ALLEGES THAT A PATIENT WAS EVALUATED AND TREATED FOR CAD AND AORTIC VALVE STENOSIS. POST RHC AND LHC PROCEDURE, DURING RECOVERY IN THE UNIT, THE PATIENT ALLEGED TO HAVE SLIGHT MEMORY LOSS AND ARM WEAKNESS AND WAS POSITIVELY DIAGNOSED WITH AN EMBOLIC STROKE. THE PATIENT WAS SUCCESSFULLY TREATED AND DISCHARGED TO HOME WITH SLIGHT MEMORY LOSS AND LEFT ARM WEAKNESS.

    Patient problems: Memory Loss/Impairment; Muscle Weakness/Atrophy

    MAUDE report 15114971 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →