Impress® 00884450038093Recalls & reported problems
Merit Medical Systems, Inc. · Class 2
All models of Impress (145)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00884450038093), received between Oct 31, 2023 and Jun 25, 2024 — classified as 5 injuries.
The most frequently reported specific device problem is “Fracture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture5
- Material Separation5
Patient problems
- Foreign Body In Patient3
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJun 25, 2024Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT DURING THE USE OF THE UAC CATHETER, THE BLACK HYDROPHILIC PORTION OF THE DEVICE DISCONNECTED FROM THE REST OF THE CATHETER. AN ATTEMPT WAS MADE TO REMOVE IT WITH A SNARE, BUT THE DISCONNECTED PART FRAGMENTED FURTHER. THE FRAGMENTS BECAME LODGED WITHIN THE RIGHT HYPOGASTRIC AND RIGHT POPLITEAL ARTERIES. NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: Foreign Body In Patient
MAUDE report 19605870 ↗InjuryFeb 19, 2024Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT DURING THE BENDING MANEUVER IN THE AORTA FOR CONTRALATERAL CATHETERIZATION OF THE LEFT INTERNAL ILIAC ARTERY, THE CATHETER DETACHED. IN AN ATTEMPT TO CAPTURE THE COMPONENTS WITH A SNARE LOOP, THEY FRAGMENTED AND WERE THEN STABILIZED IN THE INTERNAL ILIAC ARTERIES, LEFT DEEP FEMORAL ARTERY, AND LEFT PERONEAL ARTERY. NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: Foreign Body In Patient
MAUDE report 18736019 ↗InjuryNov 20, 2023Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT TIP OF THE CATHETER GOT DETACHED WHEN IT WAS PULLED INTO A SHEATH DURING TEVAR PROCEDURE. NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18172750 ↗InjuryNov 20, 2023Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT DURING AN ABDOMINAL STENTGRAFT PLACEMENT PROCEDURE (EVAR), THE TIP OF THE CATHETER BROKE OFF. THIS OCCURRED DURING MANIPULATION WITHIN THE DEPLOYED STENTGRAFT. WHEN BRINGING THE CATHETER TO THE RIGHT INTERNAL ILIAC ARTERY, IT WAS THEN BROUGHT TO THE AORTA AND THE CATHETER TIP WAS RETURNED TO A MODIFIED HOOK. WHEN THE CATHETER WAS MOVED DOWN WITHIN THE LOWER EXTREMITY, THAT IS WHEN THE TIP DETACHED. THE TIP WAS NOT ABLE TO BE REMOVED FROM THE PATIENT AND WAS LEFT WITHIN THE PATIENT'S VESSEL. THE GRAFT WAS DEPLOYED AND THE TIP OF THE CATHETER REMAINS BETWEEN THE GRAFT AND THE VESSEL.
Patient problems: Foreign Body In Patient
MAUDE report 18172747 ↗InjuryOct 31, 2023Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT THE DISTAL END OF THE CATHETER BROKE OFF WITHIN THE PATIENT DURING THE PROCEDURE. NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18038422 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page