Impress® 00884450043769Recalls & reported problems
Merit Medical Systems, Inc. · Class 2
All models of Impress (145)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00884450043769), received between Apr 4, 2023 and May 23, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Separation2
- Break1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 23, 2023Break; Material Separation; Break; Material Separation
THE ACCOUNT ALLEGES THAT DURING A SURGICAL ABDOMINAL AORTIC STENT IMPLANT PROCEDURE [AAA] ON A FEMALE PATIENT, THE PROXIMAL HYDROPHILIC SEGMENT OF THE 5F CATHETER DETACHED WITHIN THE PATIENT. THE PHYSICIAN HAD ACQUIRED RETROGRADE FEMORAL ARTERY ACCESS TO DELIVER A COIL-TYPE EMBOLIC DEVICES, DURING CATHETER MANIPULATIONS OVER A GUIDEWIRE UNDER FLUOROSCOPY, THE CATHETER TIP DETACHED WITHIN THE PATIENT'S COMMON ILIAC ARTERY. THE PHYSICIAN USED A VASCULAR SNARE DEVICE TO SUCCESSFULLY EXTERNALIZE THE FRAGMENTED FOREIGN BODIES, BUT THE FOREIGN BODY AGAIN WAS FRAGMENTED. A THROMBECTOMY CATHETER WAS EVENTUALLY USED TO REMOVE ALL FRAGMENTED PARTS LIBERATING THE PATIENT'S VESSEL. NO ADDITIONAL PATIENT CONSEQUENCES TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16986522 ↗InjuryApr 4, 2023Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT AT THE BEGINNING OF THE PROCEDURE, THE DOCTOR REQUESTED A PUNCTURE NEEDLE AND 5F INTRODUCER, LATER HE REQUESTED A HYDROPHILIC GUIDEWIRE AND A 5F CATHETER. AS IT WAS A FIBROID EMBOLIZATION, THE DOCTOR SUGGESTED USING A UAC CATHETER, BUT HE REJECTED IT. AFTER INTRODUCING THE 5F CATHETER INTO THE PATIENT, DURING ITS MANIPULATION BY THE DOCTOR, THE HYDROPHILIC TIP DISLODGED, THE DOCTOR REMOVED IT WITH AN ENDOVASCULAR LOOP, THE DOCTOR MANAGED TO REMOVE THE FRAGMENTS LODGED INSIDE THE PATIENT AND THE INTERVENTION ENDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16675233 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page