Skip to content

ReSolve® 00884450362976Recalls & reported problems

Merit Medical Systems, Inc. · Class 1

All models of ReSolve (16)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00884450362976), received between Mar 25, 2021 and Sep 25, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Detachment of Device or Device Component1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 25, 2024Fracture; Fracture

    THE ACCOUNT ALLEGES THAT A PATIENT HAD AN 8.5FR RESOLVE DRAINAGE CATHETER PLACED ON AN UNKNOWN DATE. THE PATIENT RETURNED FOR A CATHETER EXCHANGE ON (B)(6) 2024. DURING REMOVAL OF THE RESOLVE DRAIN IT WAS NOTED TO NOT BE INTACT. THE PHYSICIAN REMOVED THREE SEPARATE PIECES OF THE DRAIN WHICH THEN WAS ABLE TO CONFIRM ENTIRE DRAIN WAS RETRIEVED.

    Patient problems: Foreign Body In Patient

    MAUDE report 20300491 ↗
  • MalfunctionApr 21, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    THE ACCOUNT ALLEGES THAT WHILE CHANGING A PATIENT'S DRESSING, ON A DRAINAGE CATHETER, THE PLASTIC FITTING DISCHARGED FROM THE PROBE. A NEW DEVICE WAS IMPLANTED WITH NO PATIENT INJURY TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11701300 ↗
  • InjuryMar 25, 2021Fracture; Material Separation; Fracture; Material Separation

    THE ACCOUNT ALLEGES THAT DURING A NEPHROSTOGRAM PROCEDURE THE DRAINAGE CATHETER THAT WAS PLACED ON (B)(6) 2021 WAS NOTED TO BE BROKEN. A NEW DRAINAGE CATHETER WOULD HAVE BEEN PLACED FOR THIS PATIENT. MEDICAL INTERVENTION WOULD HAVE BEEN NECESSARY. NO ADDITIONAL PATIENT CONSEQUENCES TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11572742 ↗

Other CATHETER, IRRIGATION devices

All CATHETER, IRRIGATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page