Skip to content

Embosphere® 00884450403426Recalls & reported problems

Biosphere Medical · Class 2 · Implant

All models of Embosphere (18)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00884450403426), received between Apr 14, 2026 and Jun 4, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K181300); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Appropriate Device Problem Term/Code Not Available1
  • Illegible Information1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2026Break; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE TIP WAS OPEN AND THERE WAS NOT ENOUGH FLUID IN THE PARTICLES. THERE WAS NO MEDICAL INTERVENTION OR PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25393245 ↗
  • MalfunctionJun 2, 2026Break

    THE CUSTOMER REPORTS THE SYRINGE BROKE DURING USE DURING PREP. THERE WAS NO PATIENT INVOLVEMENT. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25363735 ↗
  • MalfunctionApr 17, 2026Leak/Splash

    THE CUSTOMER REPORTED THAT BEFORE THE PROCEDURE, THEY FOUND THE SALINE HAD LEAKED. THE PROCEDURE WAS COMPLETED WITH NEW PRODUCT. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24909586 ↗
  • MalfunctionApr 14, 2026Illegible Information

    THE CUSTOMER REPORTED THE EXPIRY DATE (YEAR) ON THE PRODUCT LABEL CONFUSED THE CUSTOMER BECAUSE IT LOOKED LIKE 2026 INSTEAD OF 2025, HENCE THE HOSPITAL REJECTED AND RETURNED THE UNUSED PRODUCT TO DISTRIBUTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24870658 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K181300. One submission often covers many models, so check the product and lot details in each record.

  • Z-3305-2026 · Aug 10, 2026 · Open, Classified
    Embosphere 500-700 micrometer 2 mL Syringe Sterile Steam, REF: S420GH/EUA, S620GH/EUA; Embosphere 700-900 micrometer 2 mL Syringe Sterile Steam, REF: S820GH/CAB; Embosphere 900-

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page