Impress® 00884450660942Recalls & reported problems
Merit Medical Systems, Inc. · Class 2
All models of Impress (145)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00884450660942), received between Jul 9, 2024 and Dec 5, 2024 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Material Separation2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 5, 2024Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT DURING USE OF THE CATHETER, THERE WAS MUCH RESISTANCE DURING WITHDRAWAL, UPON REMOVAL THE DISTAL PORTION OF THE CATHETER THE TIP OF THE CATHETER BROKE OFF WITHIN THE ILLIAC VESSEL. NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20854104 ↗MalfunctionNov 15, 2024
THERE WAS SO MUCH RESISTANCE THAT, ON WITHDRAWING ONE OF THE 5-FRENCH BERN CATHETERS, A DISTAL PORTION OF THE CATHETER SHEARED OFF ON THE END OF THE ILIAC LIMB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20693934 ↗InjuryJul 9, 2024Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT DURING THE PTA PROCEDURE, THE "BERN" CATHETER WAS USED IN AN ATTEMPT TO CANNULATE A CALCIFIED LEFT SUBCLAVIAN ARTERY AND LAA. DURING CATHETER MANIPULATIONS, THE CATHETER TIP DETACHED AND MIGRATED WITHIN THE PATIENT'S RENAL ARTERY. AN UNSUCCESSFUL ATTEMPT WAS MADE TO REMOVE THE FOREIGN BODY. CONSIDERING THE PATIENT'S AGE, HISTORY, AND CONTRAST MEDIUM USED DURING THE PROCEDURE, THE DECISION WAS MADE BY SURGERY TO LEAVE THE FOREIGN BODY WITHIN THE PATIENT. THE PATIENT'S VITAL SIGNS WERE STABLE THROUGHOUT THE PROCEDURE. THE ASYMPTOMATIC PATIENT WAS DISCHARGED TO HOME. NO ADDITIONAL PATIENT CONSEQUENCE TO REPORT.
Patient problems: Foreign Body In Patient
MAUDE report 19705317 ↗
Other Catheter, intravascular, diagnostic devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page