Impress® 00884450668122Recalls & reported problems

Merit Medical Systems, Inc. · Class 2

All models of Impress (145)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00884450668122), received between Jun 27, 2023 and Aug 6, 2025 — classified as 5 injuries, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture6
  • Material Separation5

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Foreign Body In Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2025Fracture; Fracture

    A RADIOLOGIST WAS PERFORMING AN ILIAC ANGIOPLASTY. WHILE ATTEMPTING TO PASS CATHETER THROUGH STENOSED ILIAC ARTERY, AROUND 5CM OF TIP OF CATHETER BROKE OFF AND HAD TO BE RETRIEVED. THE RADIOLOGIST WAS ABLE TO RETRIEVE THE TIP BY USING AN ANGIOPLASTY BALLOON AND STENT TO OPEN VESSELS AROUND FRAGMENT. NO REPORTED INJURY OR ADDITIONAL PATIENT DETAILS WERE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22731505 ↗
  • InjuryMay 27, 2025Fracture; Material Separation; Fracture; Material Separation

    THE ACCOUNT ALLEGES AFTER INSERTING THE DEVICES INTO THE PATIENT AND ATTEMPTING TO REMOVE THE RIM PROBE FROM THE INTRODUCER, IT FRACTURED, LEAVING THE BASE IN THE ILIAC ARTERY. THE FOREIGN BODY WAS SUCCESSFULLY REMOVED. NO ADDITIONAL PATIENT CONSEQUENCE TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22091829 ↗
  • InjuryNov 19, 2024Fracture; Material Separation; Fracture; Material Separation

    THE ACCOUNT ALLEGES THAT DURING THE PROCEDURE THE BLACK TIP OF THE CATHETER SNAPPED WHILST TRYING TO REMOVE THE CATHETER OVER AN AMPLTAZ GUIDEWIRE. THE TIP WAS SUCCESSFULLY RETRIEVED. NO ADDITIONAL PATIENT CONSEQUENCES TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20726829 ↗
  • InjurySep 18, 2024Fracture; Material Separation; Fracture; Material Separation

    THE ACCOUNT ALLEGES THAT DURING ARTERIAL ACCESS FOR A PERIPHERAL VASCULAR INTERVENTION, THE CATHETER TIP DETACHED AT THE BLACK DISTAL TIP. THE DETACHED DEVICE WAS SUCCESSFULLY RECOVERED USING AN ENDOVASCULAR SNARE REMOVAL. NO ADDITIONAL PATIENT CONSEQUENCE TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20248170 ↗
  • InjuryAug 27, 2024Fracture; Material Separation; Fracture; Material Separation

    THE ACCOUNT ALLEGES THE TIP OF A RIM CATHETER BROKE OFF IN A PATIENT'S VESSEL DURING A ROUTINE AV FISTULOGRAM. IT WAS RETRIEVED WITH A SNARE DEVICE AND THE PATIENT WAS NOT HARMED. NO ADDITIONAL PATIENT CONSEQUENCE TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20081833 ↗
  • InjuryJun 27, 2023Fracture; Material Separation; Fracture; Material Separation

    THE ACCOUNT ALLEGES THAT DURING AN ANGIOGRAM PROCEDURE THE TIP OF THE CATHETER DETACHED WITHIN THE PATIENT. WHILE ATTEMPTING TO CROSS THROUGH THE COMMON ILIAC ARTERY, THE TIP OF THE DEVICE GOT STUCK WITHIN THE HIGHLY CALCIFIED VESSEL AND DETACHED WITHIN THE PATIENT. THERE WAS AN ATTEMPT TO REMOVE THE FOREIGN BODY USING A SNARE, AND WAS UNSUCCESSFUL. THE PATIENT WAS SENT TO SURGERY TO REMOVE THE DETACHED TIP. NO ADDITIONAL PATIENT CONSEQUENCE TO REPORT.

    Patient problems: Foreign Body In Patient

    MAUDE report 17213892 ↗

Other Catheter, intravascular, diagnostic devices

All Catheter, intravascular, diagnostic devices →