NMT Module IntelliVue NMT ModuleRecalls & reported problems

Philips Medizin Systeme Böblingen GmbH · IntelliVue NMT Module is a quantitative measurement of the level of muscle relaxation as a consistent part of the IntelliVue patient monitoring. · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00884838029583), received between Oct 8, 2020 and Oct 29, 2024 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Failure to Calibrate1
  • Overheating of Device1
  • Patient Device Interaction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Blister1
  • Burn(s)1
  • Insufficient Information1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 29, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THE MEASUREMENT WAS UNRELIABLE AND THE ANESTHESIOLOGIST ADMINISTERED MEDICATION TO REVERSE PARALYTIC AGENTS. NO OTHER CLINICAL INFORMATION OR MEDICAL INTERVENTION WAS REPORTED; THEREFORE, GOOD FAITH EFFORTS ARE ONGOING.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20560247 ↗
  • InjuryJan 4, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    THE CUSTOMER REPORTED THAT THE SYSTEM TRANSDUCER HAS HAD EXCESSIVE LEAKAGE CURRENT CAUSING INJURY TO THE PATIENT'S THUMB. THE DEVICE WAS IN USE AT THE TIME OF THE EVENT, THERE WAS AN INJURY TO THE PATIENT DURING THE EVENT REPORTED.

    Patient problems: Blister

    MAUDE report 18445604 ↗
  • InjuryNov 6, 2023Overheating of Device; Overheating of Device

    IT WAS REPORTED THE PATIENT DEVELOPED A BURN AT THE NMT ELECTRODE SITE, POSSIBLY FROM THE MODULE TEMPERATURE BEING SET TOO HIGH.

    Patient problems: Burn(s)

    MAUDE report 18074697 ↗
  • MalfunctionOct 8, 2020Incorrect, Inadequate or Imprecise Result or Readings; Failure to Calibrate

    THE CUSTOMER CALLED PHILIPS TO REPORT DIFFICULTIES WITH NMT MODULE CALIBRATION, AND IF THE MODULES DO CALIBRATE, INACCURATE READINGS HAVE BEEN OBSERVED. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10652658 ↗