Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04931921011072), received between Mar 6, 2024 and Mar 6, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Alarm Not Visible” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Alarm Not Visible1
- No Audible Alarm1
- Protective Measures Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 6, 2024No Audible Alarm; Alarm Not Visible; Protective Measures Problem; No Audible Alarm; Alarm Not Visible; Protective Measures Problem
NIHON KOHDEN CLINICAL APPLICATION SPECIALIST (NKCAS) WAS ON-SITE WHEN THE ISSUE OCCURRED THAT THE G9 MONITOR DID NOT ALARM FOR THE TOF ALARM (TRAIN OF FOUR) WHEN USING THIS NMT SMART POD, TOF DEVICE WITH A G9 BEDSIDE MONITOR (BSM) DURING A SURGICAL PROCEDURE. NK CAS CONFIRMED THE NMT SMART POD'S VALUE REACHED THE ALARM LEVEL LIMIT VALUE, BUT NO ALARM WAS TRIGGERED EVEN IN ALARM EVENTS. NO PATIENT HARM WAS REPORTED. NK CAS COPIED ALL CLINICAL SETTINGS FROM A KNOWN WORKING NMT SMART POD, TOF DEVICE BUT THE ISSUE PERSISTED. NK CAS CONCLUDED THAT IT WAS NOT THE G9 BEDSIDE MONITOR (BSM) THAT HAD THE ISSUE, IT WAS AN ISSUE WITH THE NMT SMART POD, TOF AND FURTHER INVESTIGATION IS ON-GOING BY NK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18847703 ↗
Other Stimulator, Nerve, Peripheral, Electric devices
- NMT cable IntelliVue NMT Patient CablePhilips Medizin Systeme Böblingen GmbH · STIMULATOR, NERVE, PERIPHERAL, ELECTRIC6 reports
- NMT Module IntelliVue NMT ModulePhilips Medizin Systeme Böblingen GmbH · STIMULATOR, NERVE, PERIPHERAL, ELECTRIC4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page