GE HD T/R Quad Extremity CoilRecalls & reported problems
Philips Ds North America LLC · HD T/R Quad Extremity Coil · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00884838067110), received between Feb 10, 2020 and Mar 31, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
- Connection Problem1
- Defective Device1
- Therapeutic or Diagnostic Output Failure1
Patient problems
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 31, 2020Defective Device; Connection Problem; Therapeutic or Diagnostic Output Failure
THE COIL WAS PREVIOUSLY IN USE AT TURKISH SITE: BILIM OZEL TIP MERKEZI. THE END USER CALLED THE DISTRIBUTOR (GE) TO REPORT THAT COIL HAD "NO CONNECTIVITY". GE FOUND THAT A MCQA TEST FAILED AND A CABLE REPLACEMENT DID NOT RESOLVE THE ISSUE. GE REPLACED THE COIL AND SENT THE DEFECTIVE COIL TO INVIVO FOR SERVICE AND REPAIR. UPON ARRIVAL AT INVIVO, THE SERVICE AND REPAIR DEPARTMENT IDENTIFIED WARPED HOUSING ON THE QLE COIL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9907676 ↗MalfunctionFeb 10, 2020Break
COIL WAS DELIVERED TO INVIVO SERVICE AND REPAIR. UPON RECEIPT, THERMAL DAMAGE WAS NOTED ON CHIMNEY HOUSING. NOTE: THE END USER HAS CONFIRMED THAT THERE WAS NO PATIENT INVOLVEMENT WITH RESPECT TO THIS ISSUE.
Patient problems: No Patient Involvement
MAUDE report 9689805 ↗
Other Coil, Magnetic Resonance, Specialty devices
- NA SIGNA PremierGE Medical Systems, L.L.C. · Coil, magnetic resonance, specialty3 recalls0 reports
- NA SIGNA PremierGE Medical Systems, L.L.C. · Coil, magnetic resonance, specialty3 recalls0 reports
- MAGNETOM Skyra 10849580Siemens Healthcare GmbH · Coil, magnetic resonance, specialty2 recalls2 reports
- MAGNETOM Altea 11410371Siemens Healthcare GmbH · Coil, magnetic resonance, specialty1 recall2 reports
- MAGNETOM Spectra 10837643Siemens Shenzhen Magnetic Resonance Ltd. · Coil, magnetic resonance, specialty1 recall1 report
- MR Coils dS Breast 16ch 3.0TPhilips Medical Systems Nederland B.V. · Coil, Magnetic Resonance, Specialty1 recall0 reports
- Invivo 45353028072Philips Ds North America LLC · Coil, magnetic resonance, specialty1 recall0 reports
- Philips dS Breast 7ch 3.0TPhilips Ds North America LLC · Coil, magnetic resonance, specialty1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page