Affiniti 30 Diagnostic Ultrasound System 989605457591Recalls & reported problems
Philips Ultrasound LLC · Class 2
All models of Affiniti 30 Diagnostic Ultrasound System (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00884838074187), received between Oct 14, 2020 and Sep 30, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Unintended System Motion” (3 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 6, 2021Class IICompleted
Due to a software defect that can intermittently cause the system to lock-up which exiting Review Mode while performing an exam.
Z-2484-2021 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Unintended System Motion3
- Detachment of Device or Device Component2
- Break1
- Material Integrity Problem1
- Material Separation1
- Poor Quality Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Material Integrity Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950965 ↗MalfunctionNov 17, 2025Material Separation; Material Separation
IT WAS REPORTED THE C9-4V TRANSDUCER HAD SEPARATION OF THE HEAD FROM THE BODY OF THE TRANSDUCER. THIS WAS ASSOCIATED WITH AN AFFINITI 30 ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THERE WAS NO FURTHER INFORMATION AVAILABLE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23573137 ↗MalfunctionMar 28, 2025Break; Break
IT WAS REPORTED THAT THERE WAS DAMAGE TO THE MEMBRANE OF AN INTRACAVITY TRANSDUCER, EXPOSING THE METAL UNDERNEATH THE MEMBRANE, RESULTING IN A FAILED LEAKAGE CURRENT TEST. THE C9-4V TRANSDUCER WAS ASSOCIATED WITH THE AFFINITI 30 ULTRASOUND SYSTEM AT THE CUSTOMER'S SITE. THE ISSUE WAS DISCOVERED WHILE THE DEVICE WAS OUTSIDE OF CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE PHILIPS SERVICE ENGINEER EVALUATED THE TRANSDUCER AND CONFIRMED THE MEMBRANE WAS DAMAGED. THE TRANSDUCER WAS REPLACED TO ADDRESS THE CUSTOMER'S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21721584 ↗MalfunctionJun 18, 2024Poor Quality Image; Poor Quality Image
AN NMPA REPORT 1335143062024000113 WAS RECEIVED AND REPORTED A MISDIAGNOSIS, MISJUDGMENT, AND UNABLE TO RECEIVE TIMELY TREATMENT TO A PATIENT. THE AFFINITI 30 ULTRASOUND SYSTEM WAS BEING USED FOR AN UNSPECIFIED PATIENT EXAMINATION. THE 2D GAIN OF THE SYSTEM COULD NOT BE ADJUSTED, RESULTING IN IMAGE DISTORTION, AND AFFECTING THE DIAGNOSIS OF THE PATIENT. THERE WAS NO REPORTED PATIENT OR USER HARM AS A RESULT OF THIS ISSUE. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME; ATTEMPTS TO GATHER MORE DETAILS ARE UNDERWAY. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT. IF FURTHER INFORMATION BECOMES AVAILABLE OR UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19557955 ↗MalfunctionSep 5, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
A CUSTOMER REPORTED THE MONITOR ARM DETACHED FROM THE ARTICULATION ARM ASSEMBLY OF THEIR AFFINITI 30 ULTRASOUND SYSTEM. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. AN INVESTIGATION IS UNDERWAY TO DETERMINE THE ROOT CAUSE OF THE ISSUE. RESULTS OF THE INVESTIGATION WILL BE INCLUDED IN A FOLLOW-UP REPORT UPON ITS COMPLETION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17685904 ↗MalfunctionMar 8, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
A CUSTOMER REPORTED THE MONITOR FELL FROM THEIR AFFINITI 30 ULTRASOUND SYSTEM WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE MONITOR WIRES PREVENTED THE MONITOR FROM COMPLETELY DETACHING FROM THE SYSTEM. A PHILIPS SERVICE ENGINEER INSTALLED A NEW TILT SWIVEL ARM KIT TO REPAIR THE SYSTEM. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE SYSTEM HAS BEEN RETURNED TO SERVICE WITH NO ADDITIONAL ISSUES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13690895 ↗MalfunctionMay 13, 2021Unintended System Motion; Unintended System Motion
THE CUSTOMER REPORTED AN INCIDENT WHERE THE SWIVEL MECHANISM OF THEIR AFFINITI ULTRASOUND SYSTEM¿S CONTROL PANEL WOULD NOT LOCK INTO POSITION WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE ULTRASOUND SYSTEM HAS BEEN RETURNED TO SERVICE WITH NO ADDITIONAL ISSUES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11821868 ↗MalfunctionApr 14, 2021Unintended System Motion; Unintended System Motion
A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR AFFINITI 30 ULTRASOUND SYSTEM¿S CONTROL PANEL WOULD NOT LOCK INTO POSITION. IT IS UNKNOWN IF THE INCIDENT OCCURRED WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE CONTROL PANEL CABLES WERE ADJUSTED TO REPAIR THE SYSTEM. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11665892 ↗MalfunctionOct 14, 2020Unintended System Motion; Unintended System Motion
THE CUSTOMER REPORTED AN INCIDENT WHERE THE SWIVEL MECHANISM OF THEIR AFFINITI ULTRASOUND SYSTEM¿S CONTROL PANEL WOULD NOT FUNCTION PROPERLY. IT COULD NOT BE DETERMINED IF THE PROBLEM WAS IDENTIFIED WHILE THE SYSTEM WAS STATIONARY OR IN TRANSIT. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10680974 ↗
Other Transducer, ultrasonic, diagnostic devices
- Site~Rite 8 Ultrasound System 9770603Bard Access Systems, Inc. · Transducer, ultrasonic, diagnostic620 reports
- ULTRASONIC PROBE UM-S20-17SOlympus Medical Systems Corp. · Transducer, ultrasonic, diagnostic2 deaths344 reports
- Transducer X7-2t 989605414122Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic6 recalls21 reports
- EPIQ Diagnostic Ultrasound System 989605488741Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic3 recalls165 reports
- bk3000, bk3500, bk5000, bkActiv 2300Bk Medical Aps · Transducer, Ultrasonic, Diagnostic6 recalls1 report
- Transducer L12-3 989605412072Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls1 report
- Transducer L15-7io 989605387852Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
- Transducer eL18-4 EMT 989605431531Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page