Skip to content

Affiniti 30 Diagnostic Ultrasound System 989605457591Recalls & reported problems

Philips Ultrasound LLC · Class 2

All models of Affiniti 30 Diagnostic Ultrasound System (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00884838074187), received between Oct 14, 2020 and Sep 30, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Unintended System Motion” (3 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 6, 2021Class IICompleted

    Due to a software defect that can intermittently cause the system to lock-up which exiting Review Mode while performing an exam.

    Z-2484-2021 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Unintended System Motion3
  • Detachment of Device or Device Component2
  • Break1
  • Material Integrity Problem1
  • Material Separation1
  • Poor Quality Image1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Material Integrity Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950965 ↗
  • MalfunctionNov 17, 2025Material Separation; Material Separation

    IT WAS REPORTED THE C9-4V TRANSDUCER HAD SEPARATION OF THE HEAD FROM THE BODY OF THE TRANSDUCER. THIS WAS ASSOCIATED WITH AN AFFINITI 30 ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THERE WAS NO FURTHER INFORMATION AVAILABLE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23573137 ↗
  • MalfunctionMar 28, 2025Break; Break

    IT WAS REPORTED THAT THERE WAS DAMAGE TO THE MEMBRANE OF AN INTRACAVITY TRANSDUCER, EXPOSING THE METAL UNDERNEATH THE MEMBRANE, RESULTING IN A FAILED LEAKAGE CURRENT TEST. THE C9-4V TRANSDUCER WAS ASSOCIATED WITH THE AFFINITI 30 ULTRASOUND SYSTEM AT THE CUSTOMER'S SITE. THE ISSUE WAS DISCOVERED WHILE THE DEVICE WAS OUTSIDE OF CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE PHILIPS SERVICE ENGINEER EVALUATED THE TRANSDUCER AND CONFIRMED THE MEMBRANE WAS DAMAGED. THE TRANSDUCER WAS REPLACED TO ADDRESS THE CUSTOMER'S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21721584 ↗
  • MalfunctionJun 18, 2024Poor Quality Image; Poor Quality Image

    AN NMPA REPORT 1335143062024000113 WAS RECEIVED AND REPORTED A MISDIAGNOSIS, MISJUDGMENT, AND UNABLE TO RECEIVE TIMELY TREATMENT TO A PATIENT. THE AFFINITI 30 ULTRASOUND SYSTEM WAS BEING USED FOR AN UNSPECIFIED PATIENT EXAMINATION. THE 2D GAIN OF THE SYSTEM COULD NOT BE ADJUSTED, RESULTING IN IMAGE DISTORTION, AND AFFECTING THE DIAGNOSIS OF THE PATIENT. THERE WAS NO REPORTED PATIENT OR USER HARM AS A RESULT OF THIS ISSUE. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME; ATTEMPTS TO GATHER MORE DETAILS ARE UNDERWAY. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT. IF FURTHER INFORMATION BECOMES AVAILABLE OR UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19557955 ↗
  • MalfunctionSep 5, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    A CUSTOMER REPORTED THE MONITOR ARM DETACHED FROM THE ARTICULATION ARM ASSEMBLY OF THEIR AFFINITI 30 ULTRASOUND SYSTEM. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. AN INVESTIGATION IS UNDERWAY TO DETERMINE THE ROOT CAUSE OF THE ISSUE. RESULTS OF THE INVESTIGATION WILL BE INCLUDED IN A FOLLOW-UP REPORT UPON ITS COMPLETION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17685904 ↗
  • MalfunctionMar 8, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    A CUSTOMER REPORTED THE MONITOR FELL FROM THEIR AFFINITI 30 ULTRASOUND SYSTEM WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE MONITOR WIRES PREVENTED THE MONITOR FROM COMPLETELY DETACHING FROM THE SYSTEM. A PHILIPS SERVICE ENGINEER INSTALLED A NEW TILT SWIVEL ARM KIT TO REPAIR THE SYSTEM. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE SYSTEM HAS BEEN RETURNED TO SERVICE WITH NO ADDITIONAL ISSUES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13690895 ↗
  • MalfunctionMay 13, 2021Unintended System Motion; Unintended System Motion

    THE CUSTOMER REPORTED AN INCIDENT WHERE THE SWIVEL MECHANISM OF THEIR AFFINITI ULTRASOUND SYSTEM¿S CONTROL PANEL WOULD NOT LOCK INTO POSITION WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE ULTRASOUND SYSTEM HAS BEEN RETURNED TO SERVICE WITH NO ADDITIONAL ISSUES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11821868 ↗
  • MalfunctionApr 14, 2021Unintended System Motion; Unintended System Motion

    A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR AFFINITI 30 ULTRASOUND SYSTEM¿S CONTROL PANEL WOULD NOT LOCK INTO POSITION. IT IS UNKNOWN IF THE INCIDENT OCCURRED WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE CONTROL PANEL CABLES WERE ADJUSTED TO REPAIR THE SYSTEM. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11665892 ↗
  • MalfunctionOct 14, 2020Unintended System Motion; Unintended System Motion

    THE CUSTOMER REPORTED AN INCIDENT WHERE THE SWIVEL MECHANISM OF THEIR AFFINITI ULTRASOUND SYSTEM¿S CONTROL PANEL WOULD NOT FUNCTION PROPERLY. IT COULD NOT BE DETERMINED IF THE PROBLEM WAS IDENTIFIED WHILE THE SYSTEM WAS STATIONARY OR IN TRANSIT. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10680974 ↗

Other Transducer, ultrasonic, diagnostic devices

All Transducer, ultrasonic, diagnostic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page