Affiniti 30 Diagnostic Ultrasound SystemRecalls & reported problems across all models
Philips Ultrasound LLC · 2 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 1
We found 2 models of Affiniti 30 Diagnostic Ultrasound System by Philips Ultrasound LLC in the FDA's device registry (GUDID). Together they are named in 9 adverse event reports in the FDA's MAUDE database, received between Oct 14, 2020 and Sep 30, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Unintended System Motion” (3 reports).
1 FDA recall record name at least one of these models by its device identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Affiniti 30 Diagnostic Ultrasound System recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Aug 6, 2021Class IICompleted
Due to a software defect that can intermittently cause the system to lock-up which exiting Review Mode while performing an exam.
Z-2484-2021 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗
Most reported Affiniti 30 Diagnostic Ultrasound System problems
Device problems
- Unintended System Motion3
- Detachment of Device or Device Component2
- Break1
- Material Integrity Problem1
- Material Separation1
- Poor Quality Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Known Impact Or Consequence To Patient1
Affiniti 30 Diagnostic Ultrasound System models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Affiniti 30 Diagnostic Ultrasound System 989605457591Transducer, ultrasonic, diagnostic1 recall9 reports
- Affiniti 30 Diagnostic Ultrasound System 795218System, imaging, pulsed doppler, ultrasonic1 recall0 reports
Recent Affiniti 30 Diagnostic Ultrasound System adverse event reports
The 9 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Material Integrity Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950965 ↗MalfunctionNov 17, 2025Material Separation; Material Separation
IT WAS REPORTED THE C9-4V TRANSDUCER HAD SEPARATION OF THE HEAD FROM THE BODY OF THE TRANSDUCER. THIS WAS ASSOCIATED WITH AN AFFINITI 30 ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THERE WAS NO FURTHER INFORMATION AVAILABLE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23573137 ↗MalfunctionMar 28, 2025Break; Break
IT WAS REPORTED THAT THERE WAS DAMAGE TO THE MEMBRANE OF AN INTRACAVITY TRANSDUCER, EXPOSING THE METAL UNDERNEATH THE MEMBRANE, RESULTING IN A FAILED LEAKAGE CURRENT TEST. THE C9-4V TRANSDUCER WAS ASSOCIATED WITH THE AFFINITI 30 ULTRASOUND SYSTEM AT THE CUSTOMER'S SITE. THE ISSUE WAS DISCOVERED WHILE THE DEVICE WAS OUTSIDE OF CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE PHILIPS SERVICE ENGINEER EVALUATED THE TRANSDUCER AND CONFIRMED THE MEMBRANE WAS DAMAGED. THE TRANSDUCER WAS REPLACED TO ADDRESS THE CUSTOMER'S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21721584 ↗MalfunctionJun 18, 2024Poor Quality Image; Poor Quality Image
AN NMPA REPORT 1335143062024000113 WAS RECEIVED AND REPORTED A MISDIAGNOSIS, MISJUDGMENT, AND UNABLE TO RECEIVE TIMELY TREATMENT TO A PATIENT. THE AFFINITI 30 ULTRASOUND SYSTEM WAS BEING USED FOR AN UNSPECIFIED PATIENT EXAMINATION. THE 2D GAIN OF THE SYSTEM COULD NOT BE ADJUSTED, RESULTING IN IMAGE DISTORTION, AND AFFECTING THE DIAGNOSIS OF THE PATIENT. THERE WAS NO REPORTED PATIENT OR USER HARM AS A RESULT OF THIS ISSUE. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME; ATTEMPTS TO GATHER MORE DETAILS ARE UNDERWAY. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT. IF FURTHER INFORMATION BECOMES AVAILABLE OR UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19557955 ↗MalfunctionSep 5, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
A CUSTOMER REPORTED THE MONITOR ARM DETACHED FROM THE ARTICULATION ARM ASSEMBLY OF THEIR AFFINITI 30 ULTRASOUND SYSTEM. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. AN INVESTIGATION IS UNDERWAY TO DETERMINE THE ROOT CAUSE OF THE ISSUE. RESULTS OF THE INVESTIGATION WILL BE INCLUDED IN A FOLLOW-UP REPORT UPON ITS COMPLETION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17685904 ↗MalfunctionMar 8, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
A CUSTOMER REPORTED THE MONITOR FELL FROM THEIR AFFINITI 30 ULTRASOUND SYSTEM WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE MONITOR WIRES PREVENTED THE MONITOR FROM COMPLETELY DETACHING FROM THE SYSTEM. A PHILIPS SERVICE ENGINEER INSTALLED A NEW TILT SWIVEL ARM KIT TO REPAIR THE SYSTEM. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE SYSTEM HAS BEEN RETURNED TO SERVICE WITH NO ADDITIONAL ISSUES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13690895 ↗MalfunctionMay 13, 2021Unintended System Motion; Unintended System Motion
THE CUSTOMER REPORTED AN INCIDENT WHERE THE SWIVEL MECHANISM OF THEIR AFFINITI ULTRASOUND SYSTEM¿S CONTROL PANEL WOULD NOT LOCK INTO POSITION WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE ULTRASOUND SYSTEM HAS BEEN RETURNED TO SERVICE WITH NO ADDITIONAL ISSUES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11821868 ↗MalfunctionApr 14, 2021Unintended System Motion; Unintended System Motion
A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR AFFINITI 30 ULTRASOUND SYSTEM¿S CONTROL PANEL WOULD NOT LOCK INTO POSITION. IT IS UNKNOWN IF THE INCIDENT OCCURRED WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE CONTROL PANEL CABLES WERE ADJUSTED TO REPAIR THE SYSTEM. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11665892 ↗MalfunctionOct 14, 2020Unintended System Motion; Unintended System Motion
THE CUSTOMER REPORTED AN INCIDENT WHERE THE SWIVEL MECHANISM OF THEIR AFFINITI ULTRASOUND SYSTEM¿S CONTROL PANEL WOULD NOT FUNCTION PROPERLY. IT COULD NOT BE DETERMINED IF THE PROBLEM WAS IDENTIFIED WHILE THE SYSTEM WAS STATIONARY OR IN TRANSIT. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10680974 ↗
Other Philips Ultrasound LLC products
- Transducer C5-13 models · 0 reports · 12 recalls
- EPIQ Diagnostic Ultrasound System9 models · 281 reports · 5 recalls
- Transducer X5-13 models · 0 reports · 10 recalls
- EPIQ 7 DIAGNOSTIC ULTRASOUND SYSTEM5 models · 330 reports · 3 recalls
- Transducer C10-3v2 models · 0 reports · 7 recalls
- Transducer X7-2t4 models · 26 reports · 6 recalls
- Transducer C5-210 models · 0 reports · 6 recalls
- Transducer S5-12 models · 0 reports · 6 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page