IntelliSpace Cardiovascular 3.2Recalls & reported problems
Philips Medical Systems Nederland B.V. · Cardiovascular Image and Information Management Solution (CVIS) · Class 2
All models of IntelliSpace Cardiovascular (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00884838091429), received between May 26, 2026 and Sep 30, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K153022); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings2
- Application Program Problem1
- Failure to Transmit Record1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
Philips identified a software defect in ISCV (IntelliSpace Cardiovascular) system wherein the user was reported that DICOM import crashed occasionally as there was a PDFClinicalContentExecuter error.. The device was in clinical use at the time of event.Based on the new information that Philips became aware, this has been determined to be a reportable malfunction. Under specific circumstances, it could potentially lead to delayed diagnosis. Given the defect¿s nature and its potential impact on clinical decision-making if it were to recur, it is determined that this constitutes a reportable malfunction.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944390 ↗MalfunctionSep 21, 2026Incorrect, Inadequate or Imprecise Result or Readings
Philips identified a software defect in ISCV (IntelliSpace Cardiovascular) system wherein user could not retrieve a study. The device was in clinical use at the time of event.Based on the new information that Philips became aware, this has been determined to be a reportable malfunction. Under specific circumstances, it could potentially lead to delayed diagnosis. Given the defect¿s nature and its potential impact on clinical decision-making if it were to recur, it is determined that this constitutes a reportable malfunction.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50841476 ↗MalfunctionSep 17, 2026Failure to Transmit Record
Philips identified a software defect in ISCV (IntelliSpace Cardiovascular) system wherein thumbnails bar said user was unable to retrieve the lost study. The device was in clinical use at the time of event. Based on the new information that Philips became aware, this has been determined to be a reportable malfunction. Under specific circumstances, it could potentially lead to delayed diagnosis. Given the defect¿s nature and its potential impact on clinical decision-making if it were to recur, it is determined that this constitutes a reportable malfunction.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50799227 ↗MalfunctionSep 16, 2026Output Problem
Philips identified a software defect in ISCV (IntelliSpace Cardiovascular) system wherein the user found a vulnerability in ISCV 3.2.1 that permits anyone to login to ISCV without a password through a Universal Resource Locator (URL) . The device was in clinical use at the time of event. Based on the new information that Philips became aware, this has been determined to be a reportable malfunction. Under specific circumstances, it could potentially lead to delayed diagnosis. Given the defect¿s nature and its potential impact on clinical decision-making if it were to recur, it is determined that this constitutes a reportable malfunction.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50786948 ↗MalfunctionMay 26, 2026Application Program Problem
IT WAS REPORTED TO PHILIPS THAT A USER WAS UNABLE TO SAVE CHANGES TO A REPORT AN ERROR WAS RECEIVED. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT, AND NO ADVERSE EVENTS OR HARM WERE REPORTED IN THE COMPLAINT. BASED ON THE RESULTS OF THE ANALYSIS, IT WAS DETERMINED THAT THE PRODUCT HAS MALFUNCTIONED AND THE CAUSE OF THE REPORTED PROBLEM WAS A 'NAN' (NOT A NUMBER) VALUE IN THE XML FILE GENERATED BY DCMIMPORT. INTELLISPACE CARDIOVASCULAR (ISCV) GENERATED AN ERROR AS IT WAS UNABLE TO PROCESS THIS VALUE. AS AN IMMEDIATE ACTION, THE ISSUE WAS RESOLVED BY EDITING OUT THE NAN VALUE IN D:\ANALYSYSBOX FOLDER. ALTHOUGH NO ADVERSE EVENTS OR PATIENT HARM WERE REPORTED IN THE ASSOCIATED COMPLAINT, THIS MALFUNCTION HAS BEEN DETERMINED TO BE REPORTABLE. UNDER SPECIFIC CIRCUMSTANCES, IT COULD POTENTIALLY LEAD TO DELAYED DIAGNOSIS OR MISDIAGNOSIS. GIVEN THE DEFECT¿S NATURE AND ITS POTENTIAL IMPACT ON CLINICAL DECISION-MAKING IF IT WERE TO RECUR, PHILIPS HAS DETERMINED THAT THIS CONSTITUTES A REPORTABLE MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25274026 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K153022. One submission often covers many models, so check the product and lot details in each record.
- Z-2543-2025 · Aug 8, 2025 · Open, ClassifiedIntelliSpace Cardiovascular, Software 8.0.0.4.
- Z-1740-2025 · Apr 7, 2025 · Open, ClassifiedIntelliSpace Cardiovascular software versions 6.x, 7.x, and 8.x; Model Number: 830089.
- Z-1252-2025 · Apr 4, 2024 · Open, ClassifiedPhilips IntelliSpace Cardiovascular software, Model 830089.
Other System, image processing, radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page