Skip to content

EPIQ 7 DIAGNOSTIC ULTRASOUND SYSTEM 795202Recalls & reported problems

Philips Ultrasound LLC · Class 2

All models of EPIQ 7 DIAGNOSTIC ULTRASOUND SYSTEM (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00884838097889), received between Dec 5, 2022 and Apr 9, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Integrity Problem1
  • Material Separation1
  • Patient Data Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 9, 2025Material Integrity Problem; Material Integrity Problem

    AS PART OF CAPA 5677695, IT WAS IDENTIFIED THAT RETROSPECTIVE REVIEW ACTIVITIES WERE WARRANTED FOR INTRACAVITY TRANSDUCER DAMAGE IF PHYSICALLY SPLIT, FRACTURED, OR CONTAINS SHARP EDGES ON THE TRANSDUCER, REGARDLESS OF WHETHER THE DEVICE IS IN CLINICAL USE. PHILIPS ULTRASOUND IS REPORTING THIS COMPLAINT AS PART OF THE RETROSPECTIVE REVIEW COMPLETED FOR INTRACAVITY TRANSDUCER DAMAGE. IT WAS REPORTED THE C10-3V TRANSDUCER HAD A LARGE HOLE IN THE LENS AND ABLE TO SEE INSIDE. THIS TRANSDUCER WAS ASSOCIATED WITH AN EPIQ 7W ULTRASOUND SYSTEM. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE SERVICE ENGINEER REPLACED THE TRANSDUCER TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21804939 ↗
  • MalfunctionFeb 13, 2025Material Separation; Material Separation

    A CUSTOMER REPORTED THEIR 3D9-3V TRANSDUCER BROKE IN HALF DURING THE CLEANING PROCESS. THERE WAS NO HARM OR INJURY REPORTED TO THE PATIENT OR USER, THE DEVICE THE WAS NOT IN CLINICAL USE. ANALYSIS WAS PERFORMED BY ONSITE FIELD SERVICE ENGINEER WHO CONFIRMED THE REPORTED ISSUE. THE RESULTS OF THE ANALYSIS CONFIRMED THE TRANSDUCER WAS BROKEN IN HALF NEAR THE BASE. THE ISSUE WAS RESOLVED BY REPLACING THE TRANSDUCER AND THE PRODUCT IS BACK IN SERVICE. PHILIPS HAS STARTED AN INVESTIGATION INTO THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21380449 ↗
  • MalfunctionDec 5, 2022Patient Data Problem; Patient Data Problem

    A CUSTOMER REPORTED ENCOUNTERING AN INCIDENT WHERE A PATIENT'S DATA WAS MIXED WITH ANOTHER EXAM ON THE EPIQ ULTRASOUND SYSTEM. THE DATA MIX-UP WAS IDENTIFIED BY THE USER AND THERE WAS NO ALLEGATION OF ALTERED PATIENT OUTCOME AS A RESULT OF THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15915357 ↗

Other System, imaging, pulsed echo, ultrasonic devices

All System, imaging, pulsed echo, ultrasonic devices →