Vue PACS Vue PACS - 12Recalls & reported problems
Philips Medical Systems Nederland B.V. · The Vue PACS is an image management system whose intended use is to provide completely scalable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve and display images and data from all hospital modalities and information systems. The system is to be used by trained professionals including, but not limited to, physicians and medical technicians. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00884838100336), received between Apr 11, 2024 and Apr 11, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
50 other recalls cover devices cleared under the same FDA submission (K110919); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 5, 2026Class IIOpen, Classified
Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence.
Z-1692-2026 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Application Program Freezes or Fails to Launch3
- Patient Data Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 11, 2024Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
IT HAS BEEN REPORTED TO PHILIPS BY (B)(4) ((B)(6)) THAT A CUSTOMER REPORTED THAT DURING SURGICAL OPERATION VUE PACS WAS ON STANDBY AND SHUTDOWN, THIS PREVENTED THE SURGEON TO VIEW THE NECESSARY X-RAY IMAGES FOR SURGERY. THE SYSTEM WAS IN CLINICAL USE WHEN THIS OCCURRED. THERE WAS NO REPORT OF HARM. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19091057 ↗MalfunctionApr 11, 2024Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
IT HAS BEEN REPORTED TO PHILIPS BY SWISSMEDIC ((B)(4)) THAT A CUSTOMER REPORTED THAT DURING SURGICAL OPERATION VUE PACS WAS ON STANDBY AND SHUTDOWN, THIS PREVENTED THE SURGEON TO VIEW THE NECESSARY X-RAY IMAGES FOR SURGERY. THE SYSTEM WAS IN CLINICAL USE WHEN THIS OCCURRED. THERE WAS NO REPORT OF HARM. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19089840 ↗MalfunctionApr 11, 2024Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
IT HAS BEEN REPORTED TO PHILIPS BY SWISSMEDIC (GHOL-SIG-2024-042) THAT A CUSTOMER REPORTED THAT DURING SURGICAL OPERATION VUE PACS WAS ON STANDBY AND SHUTDOWN, THIS PREVENTED THE SURGEON TO VIEW THE NECESSARY X-RAY IMAGES FOR SURGERY. THE SYSTEM WAS IN CLINICAL USE WHEN THIS OCCURRED. THERE WAS NO REPORT OF HARM. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19089524 ↗MalfunctionApr 11, 2024Patient Data Problem; Patient Data Problem
IT HAS BEEN REPORTED TO PHILIPS BY SWISSMEDIC (GHOL-SIG-2024-055) THAT A CUSTOMER REPORTED THAT VUE PACS DISPLAYS PATIENT X, BUT WHEN DOUBLE-CLICKED PATIENT Y APPEARS IN THE WORKLIST. THE SYSTEM WAS IN CLINICAL USE WHEN THIS OCCURRED. THERE WAS NO REPORT OF HARM. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19088522 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K110919. One submission often covers many models, so check the product and lot details in each record.
- Z-0497-2025 · Oct 10, 2024 · Open, ClassifiedVue PACS Diagnostic Viewer with Vue PACS client 12.1.5 to 12.2.8.454 software Model Number: 1149418 Intended use is to provide completely scalable local and wide area PACS (Pictu
- Z-0235-2014 · Sep 16, 2013 · TerminatedCarestream Vue PACS; MX WEB SVR W/3 MX-O LIC< 7.5K EXAMS The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area
- Z-0259-2014 · Sep 16, 2013 · TerminatedCarestream Vue PACS; UPG MX-RAD SVR 50K EX/YR UNLIMITED The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area
- Z-0219-2014 · Sep 16, 2013 · TerminatedCarestream Vue PACS; Virtual Enterprise SW StandAlone The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area P
- Z-0216-2014 · Sep 16, 2013 · TerminatedCarestream Vue PACS; Virtual PWR Reading Basic SW 8 Pack The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area
- Z-0256-2014 · Sep 16, 2013 · TerminatedCarestream Vue PACS; UPG MX-RAD SVR 7.5K EX/YR UNLIMITED The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area
- Z-0233-2014 · Sep 16, 2013 · TerminatedCarestream Vue PACS; MX RAD SW ADV CR/DR LAYOUT & VIEW The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area P
- Z-0231-2014 · Sep 16, 2013 · TerminatedCarestream Vue PACS; MX-O SA WORKSTATION SW LICENSE The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS
- Z-0215-2014 · Sep 16, 2013 · TerminatedCarestream Vue PACS; Virtual PWR Reading Basic SW Workgroup The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide
- Z-0248-2014 · Sep 16, 2013 · TerminatedCarestream Vue PACS; MX WEB SVR W/3 MX-RAD LIC<75K EX/YR The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area
- Z-0222-2014 · Sep 16, 2013 · TerminatedCarestream Vue PACS; Unlimited Reading LIC<7.5K EX/YR The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PA
- Z-0220-2014 · Sep 16, 2013 · TerminatedCarestream Vue PACS; Virtual PWR Reading Basic SW StandAlone The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide
- Z-0237-2014 · Sep 16, 2013 · TerminatedCarestream Vue PACS; MX WEB SVR W/3 MX-O LIC < 25K EXAMS The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area
- Z-0253-2014 · Sep 16, 2013 · TerminatedCarestream Vue PACS; MX WEB SW/10 MX-RAD CONCUR USER LIC The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area
- Z-0240-2014 · Sep 16, 2013 · TerminatedCarestream Vue PACS; UPG MX-O SVR 7.5K EX/YR UNLIMITED The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area P
…and 35 more.
Other System, image processing, radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page