Pinnacle Radiation Therapy Planning System 18.0Recalls & reported problems

Philips Medical Systems (cleveland), Inc. · Class 2

All models of Pinnacle Radiation Therapy Planning System (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00884838103566), received between Jun 30, 2022 and Oct 18, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Application Program Problem” (2 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 5, 2025Class IIOpen, Classified

    Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.

    Z-2570-2025 · Root cause: Software design · Recall details · FDA record ↗

  2. Jul 17, 2025Class IIOpen, Classified

    Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,

    Z-2239-2025 · Root cause: Software change control · Recall details · FDA record ↗

  3. Aug 30, 2022Class IIOpen, Classified

    When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.

    Z-1744-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Application Program Problem2
  • Program or Algorithm Execution Failure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 18, 2024Application Program Problem; Application Program Problem

    DURING INTERNAL TESTING AN ISSUE WAS FOUND IN WHICH USERS ARE ABLE TO SELECT DATASETS FROM DIFFERENT SCANS FOR STOPPING POWER RATIO (SPR) GENERATION. THERE IS NO ALLEGATION OF A DEATH OR SERIOUS INJURY. A DEFECT WAS FILED AND HAS BEEN EVALUATED BY THE DEFECT REVIEW BOARD (DRB) AND SCORED AS AN UNACCEPTABLE SAFETY DEFECT. AN ISSUE IMPACT ASSESSMENT (IIA) HAS BEEN TRIGGERED AND PHILIPS IS INVESTIGATING THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20481319 ↗
  • MalfunctionOct 18, 2024Application Program Problem; Application Program Problem

    DURING INTERNAL TESTING AN ISSUE WAS FOUND IN WHICH THE ADVANCED REGION OF INTEREST (ROI) EXPANSION TOOL EXPANDS ROIS INCORRECTLY ON HEADFIRST PRONE (HFP), FEET FIRST SUPINE (FFS), AND FEET FIRST PRONE (FFP) DATASETS. THERE IS NO ALLEGATION OF A DEATH OR SERIOUS INJURY. A DEFECT WAS FILED AND HAS BEEN EVALUATED BY THE DEFECT REVIEW BOARD (DRB) AND SCORED AS AN UNACCEPTABLE SAFETY DEFECT. AN ISSUE IMPACT ASSESSMENT (IIA) HAS BEEN TRIGGERED AND PHILIPS IS INVESTIGATING THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20481113 ↗
  • MalfunctionAug 24, 2022Program or Algorithm Execution Failure; Program or Algorithm Execution Failure

    THE DOSE FILE THAT WAS EXPORTED TO MIM AND TO MOBIUS FOR 3 PATIENT PLANS IS SHOWING A CALCULATED DOSE VALUE THAT IS DIFFERENT FROM WHAT IS IN PINNACLE 18.0.1. THIS BEHAVIOR WAS IDENTIFIED AT A CUSTOMER SITE AFTER THEY TRANSFERRED MULTIPLE PATIENT PLANS TO MOBIUS FOR PLAN QA PURPOSES. THE PLANS FAILED QA IN MOBIUS. UPON FURTHER INVESTIGATION THE USER DISCOVERED THAT THE DOSE INFORMATION EXPORTED OUT OF PINNACLE DID NOT MATCH THE DOSE IN THE PLANS IN PINNACLE. THIS ISSUE IS CURRENTLY UNDER INVESTIGATION BY THE MANUFACTURER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15285745 ↗
  • MalfunctionJun 30, 2022Program or Algorithm Execution Failure; Program or Algorithm Execution Failure

    EXPORTED DOSE FROM A PLAN IS WRONG WHEN IMPORTED BACK INTO PINNACLE AND WHEN EXPORTING TO 3RD PARTY SYSTEMS. THIS BEHAVIOR WAS IDENTIFIED BY THE CUSTOMER WHILE DOING DOSE COMPARISON TESTING BETWEEN SOFTWARE VERSION 18.0.1 AND SOFTWARE VERSION 9.8. AFTER EXPORTING PLANS TO 3RD PARTY SYSTEM MIM IT WAS DISCOVERED THAT THE DOSE WAS DIFFERENT BETWEEN THE TWO PLANS. THIS ISSUE IS CURRENTLY UNDER INVESTIGATION BY THE MANUFACTURER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14875278 ↗

Other System, planning, radiation therapy treatment devices

All System, planning, radiation therapy treatment devices →