Skip to content

Pinnacle Radiation Therapy Planning SystemRecalls & reported problems across all models

Philips Medical Systems (cleveland), Inc. · 2 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

3 open recalls name at least one Pinnacle Radiation Therapy Planning System model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls
3

We found 2 models of Pinnacle Radiation Therapy Planning System by Philips Medical Systems (cleveland), Inc. in the FDA's device registry (GUDID). Together they are named in 7 adverse event reports in the FDA's MAUDE database, received between Jun 28, 2022 and Oct 18, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Program or Algorithm Execution Failure” (5 reports).

3 FDA recall records name at least one of these models by its device identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Pinnacle Radiation Therapy Planning System recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Aug 5, 2025Open, Classified

    Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.

    Z-2570-2025 · Root cause: Software design · Recall details · FDA record ↗

  2. Jul 17, 2025Open, Classified

    Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,

    Z-2239-2025 · Root cause: Software change control · Recall details · FDA record ↗

  3. Aug 30, 2022Open, Classified

    When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.

    Z-1744-2022 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

Most reported Pinnacle Radiation Therapy Planning System problems

Device problems

  • Program or Algorithm Execution Failure5
  • Application Program Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Pinnacle Radiation Therapy Planning System models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Pinnacle Radiation Therapy Planning System adverse event reports

The 7 most recent reports across all models, in the reporter's own words as published by the FDA.

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page