Skip to content

Patient Information Center iX (REF 866389)Recalls & reported problems

Philips Medizin Systeme Böblingen GmbH · Patient Information Center iX · Class 2

All models of Patient Information Center iX (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Philips Patient Information Center iX is a central station monitor that tracks patients' vital signs and triggers alarms. There are 24 reports naming it, including 2 deaths and 4 injuries, and most describe alarms failing to sound—such as for low oxygen or slow heart rate—sometimes with reports of delayed response to a patient's decline. Four recalls are open, involving missed push notifications, silent custom ringtones, screen crashes requiring reboots, and connectivity drops with bedside monitors.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
2
Injuries reported
4
Recalls naming it
4

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00884838104594), received between Apr 13, 2023 and Jul 16, 2026 — classified as 2 deaths, 4 injuries, 18 malfunctions.

The most frequently reported specific device problem is “No Audible Alarm” (13 reports).

4 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 3 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 28, 2024Class IIOpen, Classified

    It was found that the Care Assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when iOS device screens were locked with a passcode.

    Z-1386-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Nov 28, 2023Class IIOpen, Classified

    Push notifications may fail to send to the user under certain conditions. This could potentially result in patient harm due to delay in detection of a change in patient condition.

    Z-0731-2024 · Root cause: Software design · Recall details · FDA record ↗

  3. Aug 17, 2023Class IIOpen, Classified

    Patient Information Center (PIC iX) Release 4.x Surveillance Crash Caused by Intel Graphics Driver Error-with a blank screen and subsequently require a manual reboot of the hardware to restart and continue central patient monitoring. Potential for a delay in the detection of a change or deterioration in the condition of one or more patients

    Z-2608-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Apr 19, 2023Class IIOpen, Classified

    During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails.

    Z-0035-2025 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • No Audible Alarm13
  • Application Network Problem2
  • Application Program Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Alarm Not Visible1
  • Communication or Transmission Problem1
  • Defective Alarm1
  • Device Alarm System1

Patient problems

  • No Clinical Signs, Symptoms or Conditions17
  • Insufficient Information3
  • Bradycardia1
  • Cardiac Arrest1
  • Low Oxygen Saturation1
  • Tachycardia1

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 16, 2026No Audible Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25835125 ↗
  • InjuryMar 25, 2026No Audible Alarm

    IT WAS REPORTED, A BRADYCARDIA ALARM HAD NOT BEEN TRIGGERED ON THE MONITOR AND THE CONTROL PANEL. A DELAY IN RESPONDING TO CARDIOPULMONARY ARREST OF A BABY WAS REPORTED. NO OTHER CLINICAL INFORMATION OR MEDICAL INTERVENTION WAS REPORTED; THEREFORE, GOOD FAITH EFFORTS ARE BEING PERFORMED.

    Patient problems: Cardiac Arrest

    MAUDE report 24686658 ↗
  • InjuryJan 28, 2026Unexpected Shutdown

    PHILIPS RECEIVED A COMPLAINT ON A PATIENT INFORMATION CENTER IX INDICATING THAT THE CUSTOMER IS REQUESTING A LOG REVIEW FOR A REBOOT ISSUE WHICH OCCURRED FOR MX800 / X2/ ETCO2 & PIICIX. IT WAS REPORTED THIS ISSUE OCCURRED FOR 8 MINUTES WHILE THE PATIENT CODED. NO OTHER CLINICAL INFORMATION OR MEDICAL INTERVENTION WAS REPORTED, HOWEVER IT WAS CONFIRMED THE PATIENT IS FINE.

    Patient problems: Insufficient Information

    MAUDE report 24202854 ↗
  • MalfunctionJan 21, 2026No Audible Alarm; Incorrect Measurement

    THE CUSTOMER REPORTED THAT RHYTHM ANALYSIS IS INCORRECT AND THE UNIT DID NOT DETECT A-FLUTTER. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24132904 ↗
  • DeathJan 6, 2026No Audible Alarm

    IT WAS REPORTED THE MONITOR DID NOT GENERATE AN ALARM FOR AN OXYGEN DESATURATION OF 30% AND THE PATIENT PASSED AWAY. NO OTHER CLINICAL INFORMATION OR MEDICAL INTERVENTION WAS REPORTED; THEREFORE, GOOD FAITH EFFORTS ARE BEING PERFORMED. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT.

    Patient problems: Low Oxygen Saturation

    MAUDE report 23983571 ↗
  • InjuryDec 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PHILIPS RECEIVED A COMPLAINT ON THE PATIENT INFORMATION CENTER IX INDICATING THERE WAS AN 'INTER-BEDSIDE DYSFUNCTION'. THE INFANT PATIENT WAS TRANSFERRED FROM THE NEONATOLOGY DEPARTMENT TO ICU ON THE X3 BETWEEN 12-1PM. UPON ARRIVAL TO THE ICU, THE X3 WAS DOCKED ON THE MX400 MONITOR. IT WAS REPORTED THE PATIENT HAD A HEART ATTACK AND WAS INTUBATED AT 12:15PM AND THEN STABILIZED. THE FATHER OF THE PATIENT WAS PRESENT IN THE ROOM AND ALERTED THE NURSING STAFF. IT WAS ALSO NOTED ' AN ALARM WOULD HAVE SOUNDED ON THE MONITOR AND ON THE CONTROL PANEL BUT THERE WOULD HAVE BEEN NO BEDSIDE STOP'. ADDITIONAL INFORMATION INDICATED THE ALARM WAS GENERATED AT THE BEDSIDE AND CENTRAL STATION; HOWEVER, THE STAFF DID NOT HEAR THE ALARM AND THE FATHER OF THE PATIENT ALERTED NURSING STAFF TO THE CONDITION OF THE PATIENT. THERE WAS NO BEDSIDE OVERVIEW FROM THE MONITOR IN THE PATIENT ROOM. THE PATIENT WAS STABILIZED AND 'OK' PER BIOMED. LOGS WERE RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION RECEIVED, THE EVENT WAS LIFE-THREATENING AND REQUIRED MEDICAL INTERVENTION.

    Patient problems: Insufficient Information

    MAUDE report 23888743 ↗
  • MalfunctionFeb 25, 2025No Audible Alarm; No Audible Alarm

    A FIELD SERVICE ENGINEER REPORTED THAT THE PATIENT INFORMATION CENTER IX DID NOT SOUND FOR SPO2 NO PULSE INOP ALARMS. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21456073 ↗
  • MalfunctionFeb 5, 2025No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THAT THE IBP ALARMS DID NOT SOUND AT THE PATIENT INFORMATION CENTER IX. THE DEVICE WAS IN USE ON A PATIENT. NO PATIENT OR USER HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21313281 ↗
  • MalfunctionDec 6, 2024No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THAT THE IBP ALARMS DID NOT SOUND AT THE PATIENT INFORMATION CENTER IX. THE DEVICE WAS IN USE ON A PATIENT. NO PATIENT OR USER HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20862202 ↗
  • MalfunctionNov 20, 2024No Audible Alarm; No Audible Alarm

    PHILIPS RECEIVED A COMPLAINT ON THE PATIENT INFORMATION CENTER IX SYSTEM INDICATING THAT THERE IS NO SOUND WHEN THERE ARE MESSAGES ROOM- TO- ROOM IS ACTIVATED. THE DEVICE WAS IN USE MONITORING A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT INVOLVING A PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20736345 ↗
  • DeathNov 15, 2024Defective Alarm; Defective Alarm

    IT WAS REPORTED A PATIENT INCIDENT OCCURRED, RESULTING IN DEATH. THE USER INDICATED THE ALARMS FOR VENTRICULAR TACHYCARDIA AND VENTRICULAR FIBRILLATION WERE MISSING, WHICH RESULTED IN A DELAY IN CARE. NO OTHER CLINICAL INFORMATION OR MEDICAL INTERVENTION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 20698451 ↗
  • MalfunctionOct 21, 2024Communication or Transmission Problem; Communication or Transmission Problem

    PHILIPS RECEIVED A COMPLAINT ON THE PATIENT INFORMATION CENTER IX SYSTEM INDICATING THE SYSTEM SHOWS THE WRONG CONFIGURATION OF THE DHCP SERVER. THE DEVICE WAS IN USE MONITORING A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT INVOLVING A PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20493552 ↗
  • MalfunctionJun 6, 2024No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THE MONITOR GENERATED A VISUAL RED ALARM FOR THE PATIENT'S TACHYCARDIA BUT NO AUDIBLE ALARM WAS GENERATED. A TECHNICIAN WENT ONSITE TO PERFORM TESTING ON THE DEVICE, REVEALING THE SYSTEM FUNCTIONED PER SPECIFICATION WITH NO MALFUNCTION. THE CLINICAL AUDIT LOGS WERE ALSO REVIEWED, REVEALING A RED VENT FIB/TACHY ALARM WAS GENERATED AT THE MONITOR AND THE CENTRAL STATION AT 1256. A SHORT TIME LATER AT 1302, THE ALARM WAS ACKNOWLEDGED AT THE CENTRAL STATION. INFORMATION INDICATES THERE WAS A DELAY IN RESUSCITATION; HOWEVER, THE INFORMATION ALSO INDICATES THE RESUSCITATION WAS NOT RELATED TO MONITORING. IT WAS NOTED DURING THE RESUSCITATION THERE WERE NO ALARMS FROM THE MONITOR AND THE VENTRICULAR TACHYCARDIA WAS NOT RECOGNIZED AS A 3-STAR ALARM. THE OUTCOME FOR THE PATIENT REMAINS UNKNOWN AT THE TIME OF THIS REVIEW. BASED ON THE INFORMATION RECEIVED, THERE WAS A FAILURE TO ALARM WHICH LED TO A DELAY IN RESUSCITATION, WHICH CAN BE CONSIDERED LIFE-THREATENING AND IS THUS, A SERIOUS INJURY.

    Patient problems: Tachycardia

    MAUDE report 19477282 ↗
  • MalfunctionApr 19, 2024No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THAT THE AUDIO ALARMS AT THE CENTRAL STATION COULD NOT BE HEARD. PATIENT INVOLVEMENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19141528 ↗
  • MalfunctionMar 5, 2024No Audible Alarm; No Audible Alarm

    THE CUSTOMER REPORTED THERE WAS "ALARM LIMIT SET WITH THE PATIENT WAS NOT REACHED". THE DEVICE WAS IN USE ON PATIENT AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18839365 ↗
  • MalfunctionMar 5, 2024No Audible Alarm; No Audible Alarm

    THE CUSTOMER REPORTED A SPEAKER MALFUNCTION NO SOUND WAS COMING FROM THE DEVICE. THE DEVICE WAS USE AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18836661 ↗
  • MalfunctionJan 30, 2024Alarm Not Visible; Alarm Not Visible

    THE DEVICE WAS IN USE AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18606797 ↗
  • InjuryDec 4, 2023No Audible Alarm; No Audible Alarm

    IT WAS REPORTED ON (B)(6) 2023, THERE WAS A FAILURE TO ALARM FOR BRADYCARDIA AT THE CENTRAL STATION AND THE BEDSIDE MONITOR. THE DEVICE WAS IN USE AND THE MONITOR DISPLAYED A HEART RATE OF 54 BPM, WHICH WAS OUTSIDE THE SET PARAMETERS (HIGH LIMIT-200 BPM AND LOW LIMIT-100 BPM). IT WAS INDICATED THE BRADYCARDIA WAS NOT MANAGED IN TIME.

    Patient problems: Bradycardia

    MAUDE report 18256471 ↗
  • MalfunctionJul 11, 2023Application Network Problem; Application Network Problem

    A CUSTOMER BIOMEDICAL ENGINEER (BIOMED) REPORTED THAT ALL OF PATIENT MONITORING KEEPS REBOOTING AND DISCONNECTING FROM SERVER IN ALL FLOORS AND DEPARTMENTS. THE BIOMED CALLED BACK AND SPOKE TO A PHILIPS REMOTE SERVICE ENGINEER (RSE); SOME OF SERVERS AND SWITCHES FOR MULTIPLE DEPARTMENTS ARE STILL NOT WORKING OR IS INTERMITTENTLY. AN ADDITIONAL FOLLOW-UP CALL INDICATED THE SITE WAS RUNNING ACCORDING TO THE BIOMED. THE CAUSE OF THE ISSUE IS UNKNOWN AND HOSPITAL INFORMATION TECHNOLOGY (IT) IS ALSO WORKING WITH PHILIPS ON THE ISSUE. FURTHER INFORMATION WILL BE REVIEWED TO DETERMINE THE ROOT CAUSE. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17301639 ↗
  • MalfunctionJun 6, 2023Application Program Problem; Application Program Problem

    IT WAS REPORTED WHEN YOU CHANGE AN ALARM LIMIT, THE CHANGE IS NOT TAKEN BY THE MX40. THE BIOMEDICAL ENGINEER (BIOMED) WAS TRYING TO GENERATE RED ALARMS TO TEST THE RECORDING FUNCTION, AND NONE GENERATED WHEN THE ALARM LIMIT WAS CHANGED TO 40 BPM. WHEN THE BIOMED SHOWED STAFF HOW TO TURN OFF RESPIRATORY ALARMS OR PARAMETER, IT DID NOT TURN OFF UNTIL THE PIC WAS REBOOTED. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO THE PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17071637 ↗
  • MalfunctionMay 22, 2023Application Program Problem; Application Program Problem

    IT WAS REPORTED THAT ALARM LIMITS ARE SET ON THE PATIENT INFORMATION CENTER IX (PIC IX) BUT THEY ARE NOT TRANSFERRING TO THE MX40 DEVICES. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO THE PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16979858 ↗
  • MalfunctionApr 28, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT PATIENT MONITOR AND CENTRAL STATION ARE GIVING FALSE PATIENT VITALS. IT IS UNKNOWN IF THE DEVICE WAS IN USE AT TIME OF EVENT AND NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16829098 ↗
  • MalfunctionApr 21, 2023Application Network Problem; Application Network Problem

    IT WAS REPORTED THAT THE LIMITS THAT ARE SET ON THE CONTROL CENTER DO NOT ALARM. THE REASON FOR THIS IS THAT THE MX40 TELEMETRY DEVICE DOES NOT TAKE OVER THE SET LIMITS OF THE CONTROL CENTER. THE PROBLEM IS THAT THE CONTROL CENTER INDICATES THAT THE LIMITS HAVE BEEN SET. IT IS UNKNOWN IF THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT, NO ADVERSE EVENT OR PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16788217 ↗
  • MalfunctionApr 13, 2023Device Alarm System; Device Alarm System

    THE CUSTOMER REPORTED A SPO2 FAILURE TO ALARM, NOT RED ALARM APPEARED FROM THE DEVICE. THE DEVICE WAS IN USE AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16730514 ↗

Other Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) devices

All Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page