Patient Information Center iX recallZ-2608-2023 · Philips North America Llc · initiated Aug 17, 2023

Class II Open, Classified All recalls posted in September 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Patient Information Center (PIC iX) Release 4.x Surveillance Crash Caused by Intel Graphics Driver Error-with a blank screen and subsequently require a manual reboot of the hardware to restart and continue central patient monitoring. Potential for a delay in the detection of a change or deterioration in the condition of one or more patients

FDA root cause category: Under Investigation by firm

What the company is doing

Philips issued Urgent field Safety Notice to to U.S. customers via certified mail on 8/17/23. Philips Market Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take: If central monitoring is disrupted and blank screen displays: 1) Initiate local downtime procedures to evaluate patients at bedside and confirm local monitoring devices continue to monitor. 2) Manually reboot PIC iX to continue central monitoring This notice should be passed on to all those who need to be aware within your organization or to any organization where PIC iX 4.x devices have been transferred. 5. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will install a software patch to your device. The patch will add defensive mitigations to force an automatic resumption of central patient monitoring without manual intervention. If central monitoring is disrupted and blank screen displays: 1) Initiate local downtime procedures to evaluate patients at bedside and confirm local monitoring devices continue to monitor. 2) Manually reboot PIC iX to continue central monitoring This notice should be passed on to all those who need to be aware within your organization or to any organization where PIC iX 4.x devices have been transferred. 5. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will install a software patch to your device. The patch will add defensive mitigations to force an automatic resumption of central patient monitoring without manual intervention. If you need any further information, please contact your local Philips representative, or call 1-800-722-9377.

Recall details

Recall number
Z-2608-2023
Recall class
Class II
Status
Open, Classified
Initiated by the company
Aug 17, 2023
Posted by the FDA
Sep 19, 2023
Recalling firm
Philips North America Llc
Quantity
960 units
Product code
MHX
FDA submissions
K153702
Distribution
Nationwide Foreign: Austria Bahrain Belgium Canada Czech Republic Denmark Finland France Germany Greece India Ireland Italy Japan Jordan Kenya Kuwait Lebanon Lithuania Netherlands Norway Poland Portugal Qatar Romania Slovakia South Africa South Korea Spain Sweden Sweden Switzerland United Kingdom Utd.Arab.Emir. Uzbekistan

Product description

Patient Information Center iX with Software Version Number 4.x in the following: (1) Patient Information Center iX Product Number: 866389; (2) Patient Information Center iX Expand Product Number: 866390; (3) PIC iX Essentials Product Number: 867093; (4) PIC iX Essentials Expand Product Number: 867154

Code information (models, lots, serial numbers)

UDI: (01)00884838104594(10)4.0.1 (01)00884838104594(10)4.0.2 (01)00884838112070(10)4.1.0 Software Version Number 4.x