Affiniti CVx Diagnostic Ultrasound System 795190 (UDI 00884838106598)Recalls & reported problems
Philips Ultrasound LLC · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00884838106598), received between Feb 4, 2022 and Jul 18, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Unintended System Motion” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended System Motion3
- Device Damaged by Another Device1
- Unexpected Shutdown1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 18, 2024Device Damaged by Another Device; Device Damaged by Another Device
A CUSTOMER REPORTED AN AFFINITI ULTRASOUND SYSTEM MONITOR DETACHED FROM THE ARTICULATING ARM SUPPORT. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE FIELD SERVICE ENGINEER REPLACED THE ARTICULATING MONITOR ARM TO RESOLVE THE ISSUE. AN INVESTIGATION IS UNDERWAY TO DETERMINE THE ROOT CAUSE OF THE ISSUE. RESULTS OF THE INVESTIGATION WILL BE INCLUDED IN A FOLLOW UP REPORT UPON ITS COMPLETION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19771075 ↗MalfunctionJul 24, 2023Unintended System Motion; Unintended System Motion
A CUSTOMER REPORTED THAT THE CONTROL PANEL ARM WAS SWIVELING DURING TRANSPORT OF AN AFFINITI CVX ULTRASOUND SYSTEM. THE DEVICE WAS NOT IN CLINICAL USE AT TIME OF DISCOVERY. NO PATIENT OR USER WAS HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17379711 ↗MalfunctionMay 26, 2023Unintended System Motion; Unintended System Motion
THE CUSTOMER REPORTED THE CONTROL PANEL COULD NOT BE LOCKED IN PLACE ON THEIR AFFINITI ULTRASOUND SYSTEM WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17011730 ↗MalfunctionDec 29, 2022Unexpected Shutdown; Unexpected Shutdown
A CUSTOMER REPORTED THAT AN AFFINITI CVX ULTRASOUND SYSTEM WAS SHUTTING DOWN ON ITS OWN. IT IS UNKNOWN IF THIS OCCURRED DURING A CRITICAL PROCEDURE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16073829 ↗MalfunctionFeb 4, 2022Unintended System Motion; Unintended System Motion
A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR AFFINITI CVX ULTRASOUND SYSTEM¿S CONTROL PANEL WOULD NOT LOCK INTO POSITION PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. A PHILIPS SERVICE ENGINEER ADJUSTED AND SECURED THE TRIGGER CABLE TO REPAIR THE SYSTEM. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE ULTRASOUND SYSTEM HAS BEEN RETURNED TO SERVICE WITH NO ADDITIONAL ISSUES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13456054 ↗
Other System, Image Processing, Radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page