Pro-PICC® CT MR17034101Recalls & reported problems
Medical Components, Inc. · 4F X 55CM SINGLE LUMEN Pro-PICC® CT BASIC IR SET · Class 2 · Implant
All models of Pro-PICC CT (12)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00884908033069), received between Mar 14, 2018 and May 8, 2024 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash4
- Difficult to Remove1
- Knotted1
Patient problems
- Extravasation2
- Pain2
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 8, 2024Leak/Splash; Leak/Splash
EXTRAVASATION OF SOLUTION THROUGH THE OSTIUM. THE PATIENT WAS PUNCTURED ON (B)(6) 2024. ON (B)(6) 2024, EXTRAVASATION WAS NOTICED DUE TO MOISTURE IN THE DRESSING. ON (B)(6) 2024, THE INSERTION NURSE ASSESSED AND REMOVED THE CATHETER. THERE WAS NO NEED FOR A NEW PUNCTURE. PATIENT WAS DISCHARGED HOME.
Patient problems: Extravasation
MAUDE report 19269183 ↗MalfunctionMay 8, 2024Leak/Splash; Leak/Splash
EXTRAVASATION OF SOLUTION THROUGH THE OSTIUM. ON (B)(6) 2024 THE PROCEDURE WAS CARRIED OUT. THE OCCURRENCE WAS NOTICED ON THE SAME NIGHT OF INSERTION, THROUGH THE NEED FOR NUMEROUS DRESSING CHANGES DUE TO HUMIDITY. ON (B)(6) 2024, THE CATHETER WAS EVALUATED BY THE INSERTION TEAM AND HAD TO BE REMOVED. NEW ACCESS WAS PERFORMED
Patient problems: Extravasation
MAUDE report 19269098 ↗MalfunctionJul 1, 2021Leak/Splash; Leak/Splash
CATHETER IS LEAKING AND MACERATED AFTER DAYS OF INSERTION.
Patient problems: Insufficient Information
MAUDE report 12098950 ↗MalfunctionJul 30, 2020Leak/Splash; Leak/Splash
AFTER IMPLANTATION IT WAS NOT POSSIBLE TO FLUSH AND NOT ASPIRATE BLOOD. WHEN THE CATHETER WAS FLUSHED, LIQUID CAME OUT FROM THE PUNCTURE SITE. THE DOCTOR PULLED THE PICC. THE PICC HAS A SPLIT AFTER 5-6 CM FROM THE PUNCTURE SITE TOWARDS THE HEART. THE DOCTOR IMPLANTED A NEW PICC FROM THE SAME PUNCTURE SITE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10344266 ↗InjuryMar 7, 2019Difficult to Remove; Difficult to Remove
DURING PLACEMENT OF THE PICC, A KNOT AROSE IN THE GUIDEWIRE. THE VESSEL WAS SMOOTHLY PUNCTURED AND THE GUIDEWIRE INTRODUCED, INITIALLY TAKING THE JUGULAR PATHWAY. THE GUIDEWIRE CURLED BACK IN SEVERAL DIRECTIONS AND EVENTUALLY A KNOT AROSE. THE WIRE GOT STUCK IN THE VEIN APPROXIMATELY 10 CM PROXIMAL FROM THE PUNCTURE SIDE. THE WIRE WAS EVENTUALLY PULLED FROM THE VESSEL, CAUSING DAMAGE TO THE BASILICA VEIN A NEW GUIDEWIRE AND DILATOR WERE OPENED AND USED TO PLACE THE PICC IN ANOTHER VEIN.
Patient problems: Pain
MAUDE report 8398444 ↗MalfunctionMar 14, 2018Knotted; Knotted
THE VESSEL WAS SMOOTHLY PUNCTURED AND THE GUIDEWIRE WAS INTRODUCED. THERE WAS A LITTLE BIT OF RESISTANCE. THEY TRIED TO RETRACT THE GUIDEWIRE IN ORDER TO REPOSITION THE NEEDLE A LITTLE BIT CLOSER TO THE VESSEL, BUT THE GUIDEWIRE WOULDN'T RETRACT. THE NEEDLE WAS REMOVED, BUT THE GUIDEWIRE WAS STUCK IN THE SUBCUTANEOUS FAT TISSUE OF THE PATIENT. A SMALL INCISION WAS MADE IN ORDER TO REMOVE THE KNOTTED GUIDEWIRE.
Patient problems: Pain
MAUDE report 7337240 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page