- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00884908033250), received between Jul 1, 2019 and Aug 1, 2023 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Leak/Splash3
- Break2
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 1, 2023Leak/Splash; Leak/Splash
A LEAKING FROM THE CATHETER HUB WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17439074 ↗MalfunctionOct 31, 2022Break; Break
IMPLANTATION OF A PICC CATHETER WAS PERFORMED. AFTER INTRODUCING THE PICC, AN ATTEMPT WAS MADE TO REMOVE THE GUIDE WIRE WITH A LOT OF RESISTANCE. WHEN REMOVING THE GUIDE WIRE, A FRACTURE WAS IDENTIFIED WITH EXTRAVASATION OF SERUM IN THE CATHETER BODY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15698941 ↗InjurySep 2, 2022Break; Break
ATTEMPTED AT PUNCTURE OF PICC PERFORMED ACCORDING TO MEDICAL REQUEST. VENOUS NETWORK EVALUATED WITH USG AID AND LEFT BASIC VEIN SELECTED FOR PUNCTURE. STARTED THE GUIDED PROCEDURE WITH USG. NEEDLE PUNCTURE PERFORMED AND AFTER VENOUS RETURN INSERTED APPROXIMATELY 4 CM FROM THE GUIDE WIRE AND THERE WAS RESISTANCE, OPTED FOR THE REMOVAL OF THE GUIDE WIRE TO REPOSITION THE NEEDLE, BUT WHEN ATTEMPTING TO REMOVE THE GUIDE WIRE, IT DID NOT COME OUT, NEEDLE PULLED OUT AND REMOVED BUT THE GUIDE WIRE IS NOT, YET IT WENT OUT. MEDICAL COMMUNICATION WHO EVALUATED AND WAS NOT ABLE TO REMOVE THE GUIDE WIRE EVEN AFTER PERFORMING A PIC ON THE SKIN. I ALSO INFORM THAT THE SITE HAS NO ACTIVE BLEEDING, SHOWING THAT THE GUIDE WIRE POSSIBLE WAS IN THE SUBCUTANEOUS TISSUE. AFTER PERFORMING X-RAY, IT SHOWED A FOLD IN THE TIP OF THE GUIDE WIRE. VASCULAR PHYSICIAN PERFORMS REMOVAL BY TRACTION AND AFTER A NEW X-RAY, THE PRESENCE OF A REMAINING FRAGMENT IN THE SUBCUTANEOUS TISSUE WAS EVIDENCED. PATIENT WILL BE SUBMITTED TO DISSECTION FOR FRAGMENT REMOVAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15348186 ↗MalfunctionJul 22, 2021Leak/Splash; Leak/Splash
CATHETER INSERTED ON (B)(6) 2021, SHOWED EXTERNALIZATION OF APPROXIMATELY 4CM AND RUPTURE WITH MEDICATION OVERFLOW ON (B)(6) 2021, EVIDENCED DURING DRESSING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12210242 ↗MalfunctionMar 24, 2021Leak/Splash; Leak/Splash
TWO DAYS AFTER INSERTION DURING FIRST INFUSION SESSION A LEAK WAS NOTED IN THE EXTENSION LINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11562461 ↗MalfunctionJul 1, 2019Fracture; Fracture
CATHETER RUPTURED 2MM BEYOND THE 26CM MARK. REASON FOR PASSAGE OF THE PICC TO ANTIBIOTIC THERAPY (ROCEFIN)
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8748421 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page