Medical Components, Inc. · 4F X 20CM SINGLE CT MIDLINE BASIC NURSING SET · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00884908113730), received between Jun 19, 2020 and Aug 7, 2026 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Material Twisted/Bent2
  • Appropriate Device Problem Term/Code Not Available1
  • Difficult to Remove1
  • Fracture1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient2
  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 7, 2026Difficult to Remove

    During Midline insertion guidewire lodged in the subcutaneous tissue. During the difficult removal process, a knot was observed in the guidewire.

    Patient problems: Insufficient Information

    MAUDE report 50172846 ↗
  • InjurySep 3, 2025Break; Break

    DEVICE WAS IMPLANT ON (B)(6) 2025. PATIENT SUFFERS FROM SEVERAL DISEASES AND UNDERWENT CT SCAN. THE IMAGES SHOWED A TUBULAR, HYPERDENSE MATERIAL IN THE PULMONARY ARTERIAL BRANCHES, LIKELY TO BE A CATHETER FRAGMENT. THE DEVICE WAS FOUND TO BE "TRUNKED" AND REMOVED. PATIENT WAS KEPT UNDER OBSERVATION FOR SEVEN DAYS WITHOUT ANY COMPLICATIONS BEING OBSERVED.

    Patient problems: Foreign Body In Patient

    MAUDE report 22952324 ↗
  • MalfunctionMar 1, 2023Material Deformation; Material Deformation

    ULTRASOUND GUIDED VEIN IDENTIFICATION BEFORE POSITIONING THE CATHETER. THE NEEDLE WAS INSERTED INTO THE VEIN. WHEN VENOUS BLOOD FLOWED BACK, THE GUIDEWIRE WAS INTRODUCED AND PROCEEDED INTO THE VEIN A FEW CENTIMETERS BUT THEN IT STOPPED. THE NEEDLE WAS REMOVED, WHILE THE GUIDE WAS STUCK UNDER THE SKIN AND DOES NOT COME OUT. TRACTION WAS PERFORMED UNTIL IT CAME OUT. COMPRESSIVE BANDAGE WAS APPLIED. THE GUIDEWIRE WAS KNOTTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16463241 ↗
  • MalfunctionMar 17, 2022Material Twisted/Bent; Material Twisted/Bent

    DURING MIDLINE INSERTION ON RIGHT BRACHIAL VEIN, GUIDEWIRE WELL POSITIONED WITH ULTRASOUND, AFTER PEEL-AWAY INSERTION, GUIDEWIRE REMOVAL WAS LIMITED BY OBSTACLES. ALL PEEL-AWAY PLUS GUIDEWIRE WAS REMOVED AND THE TIP OF GUIDEWIRE WAS KNOTTED. NEW PROCEDURE WAS STARTED, COMPRESSION TO REDUCE HEMATOMA, NEW SET WAS USED AND PLACED CORRECTLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13792183 ↗
  • MalfunctionMar 17, 2022Material Twisted/Bent; Material Twisted/Bent

    DURING MIDLINE INSERTION ON RIGHT BRACHIAL VEIN, GUIDEWIRE WELL POSITIONED WITH ULTRASOUND, AFTER PEEL-AWAY INSERTION, GUIDEWIRE REMOVAL WAS LIMITED BY OBSTACLES. ALL PEEL-AWAY PLUS GUIDEWIRE WAS REMOVED AND THE TIP OF GUIDEWIRE WAS COILED. AFTER SMALL BLEEDING STOP, NEW SET WAS USED AND PLACED CORRECTLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13792174 ↗
  • MalfunctionSep 8, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    DIFFICULT GUIDEWIRE REMOVAL. IT KNOTTED ON THE DISTAL TIP. IR PROCEDURE TO REMOVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12439233 ↗
  • MalfunctionJul 22, 2020Fracture; Fracture

    LINE FRACTURED WHILE IN USE, MEDICATION LEAKING FROM CRACK IN LINE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10308262 ↗
  • MalfunctionJun 19, 2020Break; Break

    GUIDEWIRE TIP KNOTTED AND BROKE INTO FOUR (4) PIECES WHICH REQUIRED SURGERY FOR REMOVAL.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10174159 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →