VASCU-PICC® MD17012601 (UDI 00884908142167)Recalls & reported problems
Medical Components, Inc. · 2.6F X 50CM DUAL LUMEN VASCU-PICC® BASIC SET · Class 2 · Implant
All models of VASCU-PICC (22)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00884908142167), received between Jan 27, 2026 and Sep 24, 2026 — classified as 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash4
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Bruise/Contusion1
- Extravasation1
- Pain1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjurySep 24, 2026Leak/Splash
PICC dressing noted to have drainage so primary RN asked for assistance changing dressing. Once old, soiled dressing was taken down, leaking noted to be coming from actual catheter and not from insertion site. MD notified and PICC removed. PICC team attempted to get a new PICC line but was unable to see a vein that she felt comfortable using. PIV placed instead. Brusing noted around the insertion site. Once PICC removed, further inspection confirmed that there was a hole in actual catheter originating from the white lumen (lumen that had sedation and pressors infusing).
Patient problems: Bruise/Contusion
MAUDE report 50884329 ↗InjuryJul 31, 2026Leak/Splash
This event involves a patient who was admitted to "[redacted]" Pediatric Intensive Care Unit (PICU) on "[redacted]" from an outside hospital for surgical management. On "[redacted]"-the day of admission, the patient underwent surgery for an open external incision and drainage of the retropharyngeal and mediastinal abscess via the left neck. On "[redacted]"-post-op day 1, the patient underwent a bedside procedure to insert a peripherally inserted central catheter (PICC) in the right upper extremity for antibiotic therapy without complication.On "[redacted]"-post-op day 7, the patient demonstrated wincing when the white port of the PICC line was flushed as per routine. Initially a 10 cubic centimeter (cc) syringe was used to flush the port but was met with resistance. A second attempt was made to flush the port with a 3cc syringe that was filled with heparin prior to flushing. The patient continued to wince, and leaking was noted at the insertion site. An Intravenous Team (IV) nurse was consulted to evaluate the site. Assessment of the PICC also indicated that the red port was not flushing. A decision was made to discontinue use of the line and to remove the PICC. On "[redacted]"-pos
Patient problems: Pain
MAUDE report 50096990 ↗InjuryJul 31, 2026Leak/Splash
This event involves a patient who was admitted to "[redacted]" Pediatric Intensive Care Unit (PICU) on "[redacted]" from an outside hospital for monitoring and management. On "[redacted]"-hospitalization day #45, the 2.6 Fr picc line was found cracked at the hub when the PICC RN came to assess and administered a tissue plasma activator (TPA), which is a thrombolytic agent, to the line. The TPA was administered to both lumens but after the dwell both lumens were still very sluggish with no blood return. The dressing was removed to further investigate and was found to be wet underneath. The line was then flushed with the dressing off and leaking was visible from catheter. On "[redacted]"-hospitalization day #47 a new PICC was placed by IV RN on "[redacted]"-that day at 1123 at the bedside over a guidewire at the same site as the previous PICC and midline. PICC breakage a result of user error, not device failure although there were still concerns regarding the product by the team.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50096989 ↗MalfunctionFeb 27, 2026Fracture
PICC LINE FRACTURE WHICH IS NON-REPAIRABLE. WILL NEED TO BE REPLACED, EXCHANGED UNDER CARDIAC ANEESTHESIA SEDATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24471158 ↗MalfunctionJan 27, 2026Leak/Splash
AT 1600 RN RETURNS PATIENT TO BED FROM MOMS' ARMS AFTER DRAWING BLOOD FROM PICC LINE AND CSF FROM EVD. MOM NOTICES A WET SPOT ON HER SHIRT LEADING TO THE DISCOVERY OF A NEWLY INFILTRATED LUE PICC. THIS LINE WAS PLACED (B)(6). FAULTY LINE REMOVED FROM PT. NEW PICC PLACED BY PICC TEAM SUCCESSFULLY AND CONFIRMED BY XRAY.
Patient problems: Extravasation
MAUDE report 24187815 ↗
Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page