Alinity m 09N43-095Recalls & reported problems
Abbott Molecular Inc. · Alinity m EBV AMP Kit · Class 2
All models of Alinity m (12)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00884999049505), received between Feb 27, 2023 and Jan 20, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “False Negative Result” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- False Negative Result1
- False Positive Result1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 20, 2026False Negative Result
THE CUSTOMER REPORTED FALSE NOT DETECTED RESULTS ON THE ALINITY M EBV AMP KIT. THE CUSTOMER REPORTED 1 PATIENT HAD 2 FALSE NOT DETECTED RESULTS: SAMPLE ID (SID) (B)(6) WAS "NOT DETECTED" FOR EBV-PLASMA PROTOCOL ON (B)(6) 2025. IN ADDITION SID (B)(6) (VIRAL LOAD OF 1479 IU/ML FOR WHOLE BLOOD) WAS "NOT DETECTED" FOR EBV-PLASMA ON (B)(6) 2025 AS WELL. THE PATIENT HAS A CHRONIC T-CELL EBV INFECTION AND HAS BEEN POSITIVE WITH COMPARABLE VALUES SINCE (B)(6) 2025 (CONFIRMED THROUGH REFERENCE LAB IN PARIS). THE CUSTOMER IS ALLEGING THAT ALL NOT DETECTED RESULTS FOR THAT PATIENT ARE QUESTIONED AS THE PATIENT HAS A KNOWN EBV INFECTION CONSTANTLY SHOWING SINCE (B)(6) 2025 (ALSO IN SEROLOGY). BASED ON THIS INFORMATION PROVIDED, THE CUSTOMER IS ALLEGING DISCREPANT FALSE NOT DETECTED RESULTS ON EBV-PLASMA (LIST NUMBER 09N43-91) FOR SID (B)(6) AND SID (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24121824 ↗MalfunctionFeb 27, 2023False Positive Result; False Positive Result
THE CUSTOMER REPORTED DISCREPANT (FALSE DETECTED) ALINITY EBV RESULTS WITH LOWER LIMIT OF QUANTIFICATION (LLOQ). SAMPLE ID, ALINITY RESULT, DATE, MAYO RESULT : (B)(6). MAYO RESULTS WERE GENERATED USING AN ALTERNATIVE PLATFORM (ROCHE MOLECULAR). THERE WAS NO IMPACT TO PATIENT MANAGEMENT FOR THE SAMPLE IDS LISTED ABOVE. MAYO RESULTS WERE RELEASED TO PROVIDER. THE REPORTED EVENT WAS CONSIDERED FALSE DETECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16450192 ↗
FDA clearance and classification
Cleared through 510(k) K212778 on Jul 15, 2022 (decision: Substantially Equivalent). Submitted by Abbott Molecular, Inc. as “Alinity m EBV AMP Kit (List No. 09N43-095), Alinity m EBV CTRL Kit (List No. 09N43-085), Alinity m EBV CAL Kit (List No. 09N43-075)”. The FDA review took 317 days. FDA record ↗
The FDA regulates product code QLX (nucleic acid amplification test for the quantitation of epstein-barr virus (ebv) dna) under 21 CFR 866.3183 as a Class II device. 21 CFR 866.3183 ↗
Common questions about the Alinity m 09N43-095
Has the Alinity m 09N43-095 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00884999049505), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Alinity m 09N43-095?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Feb 27, 2023 and Jan 20, 2026. In those reports, the most common specific device problems are “False Negative Result” (1) and “False Positive Result” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Alinity m 09N43-095?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Alinity m 09N43-095 FDA approved?
It was cleared through 510(k) K212778 on Jul 15, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Alinity m 09N43-095, and is it still sold?
The FDA's device database lists Abbott Molecular Inc. as the company responsible for it (brand Alinity m, model 09N43-095). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.