Abbott Molecular Inc. · 13 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

7 open recalls name at least one Alinity m model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1,534
Deaths reported
0
Injuries reported
7
Recalls
14

We found 13 models of Alinity m by Abbott Molecular Inc. in the FDA's device registry (GUDID). Together they are named in 1,534 adverse event reports in the FDA's MAUDE database, received between May 5, 2020 and Sep 28, 2026 — classified as 7 injuries, 1,527 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (559 reports).

14 FDA recall records name at least one of these models by its device identifier, 7 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 10 reportsDec 2023: 6 reportsJan 2024: 28 reportsFeb 2024: 0 reportsMar 2024: 8 reportsApr 2024: 4 reportsMay 2024: 3 reportsJun 2024: 6 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 9 reportsOct 2024: 17 reportsNov 2024: 7 reportsDec 2024: 5 reportsJan 2025: 4 reportsFeb 2025: 10 reportsMar 2025: 14 reportsApr 2025: 19 reportsMay 2025: 23 reportsJun 2025: 13 reportsJul 2025: 16 reportsAug 2025: 22 reportsSep 2025: 5 reportsOct 2025: 6 reportsNov 2025: 9 reportsDec 2025: 7 reportsJan 2026: 9 reportsFeb 2026: 4 reportsMar 2026: 6 reportsApr 2026: 5 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 6 reportsSep 2026: 5 reportsOct 2026: 0 reports20242025202628
Last 36 months. The most recent month may be incomplete.

Alinity m recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Mar 20, 2025Open, Classified

    The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.

    Z-1665-2025 · Root cause: Device Design · Recall details · FDA record ↗

  2. Sep 12, 2024Open, Classified

    Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay controls can be traced to iron leaching into the Alinity m Lysis Solution from the lysis transfer pump in the Alinity m System.

    Z-0234-2025 · Root cause: Device Design · Recall details · FDA record ↗

  3. Sep 12, 2024Open, Classified

    Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay controls can be traced to iron leaching into the Alinity m Lysis Solution from the lysis transfer pump in the Alinity m System.

    Z-0233-2025 · Root cause: Device Design · Recall details · FDA record ↗

  4. Mar 20, 2024Open, Classified

    The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no retention bar." Abbott has identified an issue concerning the use of certain third-party Universal Transport Medium (UTM) tubes on the Alinity m System. This issue has been observed with Copan UTM-RT 3 mL Without Beads Transport & Preservation Medium for Viral Molecular Diagnostics Testing, Part Number (PN) 3C047N. However, it is possible that other third party manufactured tubes could experience the same issue.

    Z-1602-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  5. Oct 30, 2023Open, Classified

    Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinity m HCV AMP Kit.

    Z-0407-2024 · Root cause: Other · Recall details · FDA record ↗

  6. Nov 11, 2022Open, Classified

    The instrument was found to be installed without plunger assemblies in the clamp bar for an Amplification Detection Unit. After correction of the issue, the instrument was returned for use without calibration. This would result in a potential for delay of results and incorrect results.

    Z-1349-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  7. Aug 19, 2022Open, Classified

    Reports received of an increase of message code 9186 (internal control failed) when using the kit.

    Z-1679-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  8. Feb 11, 2022Terminated

    There is potential for the Reaction Vessels (RVs) detaching from the RV cap during transport by the pipettor robot.

    Z-0710-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  9. Dec 6, 2021Terminated

    There is an issue with the installation of updated camera firmware on the system.

    Z-0463-2022 · Root cause: Software design · Recall details · FDA record ↗

  10. Sep 2, 2021Terminated

    There is a potential for false positive results.

    Z-0008-2022 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  11. Sep 2, 2021Terminated

    There is a potential for false positive results.

    Z-0007-2022 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  12. Sep 2, 2021Terminated

    There is a potential for false positive results.

    Z-0006-2022 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  13. Sep 2, 2021Terminated

    There is a potential for false positive results.

    Z-0005-2022 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  14. Sep 2, 2021Terminated

    There is a potential for false positive results.

    Z-0004-2022 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

Most reported Alinity m problems

Device problems

  • False Positive Result559
  • Program or Algorithm Execution Problem302
  • Fluid/Blood Leak237
  • Device Dislodged or Dislocated194
  • Unable to Obtain Readings130
  • Mechanical Problem53
  • False Negative Result39
  • Incorrect, Inadequate or Imprecise Result or Readings27

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,512
  • No Known Impact Or Consequence To Patient9
  • Abrasion5
  • Skin Tears3
  • Head Injury1
  • Limb Fracture1

Alinity m models

13 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Alinity m adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Abbott Molecular Inc. products

All Abbott Molecular Inc. devices →