Skip to content

CD HORIZON® Spinal System 7540020Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SET SCREW 7540020 BREAK OFF TI · Class 2 · Implant

All models of CD HORIZON Spinal System (341)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic CD Horizon spinal system set screw has 262 adverse event reports, including 230 malfunctions, 32 injuries and no deaths. Reports mostly describe the screw slipping, failing to align, or breaking during spine surgery, and several describe threads stripping during insertion; most report no patient complications. There are no recalls that name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
257
Deaths reported
0
Injuries reported
32
Recalls naming it
0

Summary

The FDA's MAUDE database contains 257 adverse event reports that list this exact device (DI 00885074457093), received between Nov 15, 2016 and Sep 22, 2026 — classified as 32 injuries, 225 malfunctions.

The most frequently reported specific device problem is “Device Slipped” (90 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 6 reportsFeb 2025: 2 reportsMar 2025: 3 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 6 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 7 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 0 reportsApr 2026: 5 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Slipped90
  • Failure to Align76
  • Material Deformation26
  • Migration or Expulsion of Device22
  • Break19
  • Mechanical Problem7
  • Device Dislodged or Dislocated6
  • Mechanics Altered6

Patient problems

  • No Known Impact Or Consequence To Patient124
  • No Clinical Signs, Symptoms or Conditions113
  • No Code Available13
  • Failure of Implant10
  • No Consequences Or Impact To Patient10
  • Pain7

Recent adverse event reports

The 25 most recent of 257 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 22, 2026Product Quality Problem; Material Deformation

    Information was received from multiple sources (Manufacturer Representative, Distributor) regarding a device used in a Spinal therap y. It was reported that the set screw was received in damaged condition from the warehouse prior to supply to a hospital. There was no patient involvement in the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50850867 ↗
  • MalfunctionSep 17, 2026Material Deformation

    Information was received from multiple sources (Manufacturer Representative, Healthcare Provider, Distributor) regarding a patient h aving Transforaminal Lumbar Interbody Fusion due to Degenerative Disc Disease. It was reported that set screw thread was stripped during screw tightening and could not be used for the patient.. There were no patient symptoms reported. There were no further complic ations reported regarding the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50797847 ↗
  • MalfunctionApr 28, 2026Mechanical Problem

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A BREAK OFF SET SCREW USED FOR SPINAL THERAPY. IT WAS REPORTED THAT, PRIOR TO IMPLANT, MULTIPLE BREAK OFF SET SCREW UNITS HAD ATTACHED TO THE KEY, MAKING IT IMPOSSIBLE TO BREAK. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25015276 ↗
  • MalfunctionApr 27, 2026Mechanical Problem

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A BREAK OFF SET SCREW USED FOR SPINAL THERAPY. IT WAS REPORTED THAT, PRIOR TO IMPLANT, MULTIPLE BREAK OFF SET SCREW UNITS HAD ATTACHED TO THE KEY, MAKING IT IMPOSSIBLE TO BREAK. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24996880 ↗
  • MalfunctionApr 24, 2026Mechanical Problem

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A BREAK OFF SET SCREW USED FOR SPINAL THERAPY. IT WAS REPORTED THAT, PRIOR TO IMPLANT, MULTIPLE BREAK OFF SET SCREW UNITS HAD ATTACHED TO THE KEY, MAKING IT IMPOSSIBLE TO BREAK. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24981815 ↗
  • MalfunctionApr 23, 2026Mechanical Problem

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A BREAK OFF SET SCREW USED FOR SPINAL THERAPY. IT WAS REPORTED THAT, PRIOR TO IMPLANT, MULTIPLE BREAK OFF SET SCREW UNITS HAD ATTACHED TO THE KEY, MAKING IT IMPOSSIBLE TO BREAK. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24963773 ↗
  • MalfunctionApr 21, 2026Mechanical Problem

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A BREAK OFF SET SCREW USED FOR SPINAL THERAPY. IT WAS REPORTED THAT, PRIOR TO IMPLANT, MULTIPLE BREAK OFF SET SCREW UNITS HAD ATTACHED TO THE KEY, MAKING IT IMPOSSIBLE TO BREAK. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24935466 ↗
  • MalfunctionFeb 8, 2026Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT DURING A TLIF PROCEDURE WITH PEDICLE SCREW PLACEMENT AT LEVELS L2-L5 FOR A PATIENT WITH A PRE-OPERATIVE DIAGNOSIS OF LUMBOSACRAL PLEXOPATHY, THE SET SCREW BROKE BEFORE FINAL TIGHTENING. A NEW SET SCREW WAS USED TO SECURE THE CONSTRUCT AND COMPLETE THE SURGERY. NO FRAGMENTS OF THE IMPLANT OR INSTRUMENT REMAINED IN THE PATIENT. THE PRODUCT WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24288215 ↗
  • MalfunctionFeb 4, 2026Break; Product Quality Problem

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING AN EVENT WHICH OCCURRED DURING A L2 - L5 PEDICLE SCREW PLACEMENT, TLIF IN A PATIENT DIAGNOSED WITH LUMBOSACRAL PLEXOPATHY. IT WAS REPORTED THAT THE SET SCREW BROKE BEFORE FINAL TIGHTENING DUE TO WHICH ANOTHER SET SCREW WAS USED TO SECURE THE ROD AND COMPLETE THE SURGERY. EVENT OCCURRED DUE TO DEFECTIVE PRODUCT. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24256786 ↗
  • MalfunctionFeb 2, 2026Device Slipped

    INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, DISTRIBUTOR) REGARDING AN EVENT WHICH OCCURRED DURING P ROCEDURE OF REMOVAL OF THE TUMOR AND FIXATION OF THE IMPLANT IN A PATIENT DIAGNOSED WITH CERVICAL AND THORACIC SPINE TUMORS. IT WAS REPORTED THAT DURING A NORMAL SURGICAL PROCEDURE, WHEN TIGHTENING THE SET SCREW, THE SET SCREW AND SCREW SLIPPED, INDICATING THAT THE SET SCREW OCCURRED SLIPPERY THREAD. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24229679 ↗
  • MalfunctionDec 4, 2025Material Deformation; Material Deformation

    INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, DISTRIBUTOR) REGARDING A PATIENT HAVING SPINAL DEFORMITY CORRECTION PROCEDURE FOR THORACOLUMBAR SCOLIOSIS. IT WAS REPORTED THAT DIFFICULTIES AROSE WHILE INSERTING THE RODS DUE TO THE SEVERITY OF THE SCOLIOSIS. THREAD DAMAGE (STRIPPED THREADS) OCCURRED IN LOCKING SCREWS, WHICH WERE SUBSEQUENTLY DEEMED UNUSABLE AND REPLACED WITH NEW ONES. THERE WERE NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THE REPORTED EVENT OCCURRED DURING NORMAL USAGE OF DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23713662 ↗
  • MalfunctionDec 4, 2025Material Deformation; Material Deformation

    INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, DISTRIBUTOR) REGARDING A PATIENT HAVING SPINAL DEFORMITY CORRECTION PROCEDURE FOR THORACOLUMBAR SCOLIOSIS. IT WAS REPORTED THAT DIFFICULTIES AROSE WHILE INSERTING THE RODS DUE TO THE SEVERITY OF THE SCOLIOSIS. THREAD DAMAGE (STRIPPED THREADS) OCCURRED IN LOCKING SCREWS, WHICH WERE SUBSEQUENTLY DEEMED UNUSABLE AND REPLACED WITH NEW ONES. THERE WERE NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THE REPORTED EVENT OCCURRED DURING NORMAL USAGE OF DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23713653 ↗
  • MalfunctionDec 4, 2025Material Deformation; Material Deformation

    INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, DISTRIBUTOR) REGARDING A PATIENT HAVING SPINAL DEFORMITY CORRECTION PROCEDURE FOR THORACOLUMBAR SCOLIOSIS. IT WAS REPORTED THAT DIFFICULTIES AROSE WHILE INSERTING THE RODS DUE TO THE SEVERITY OF THE SCOLIOSIS. THREAD DAMAGE (STRIPPED THREADS) OCCURRED IN LOCKING SCREWS, WHICH WERE SUBSEQUENTLY DEEMED UNUSABLE AND REPLACED WITH NEW ONES. THERE WERE NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THE REPORTED EVENT OCCURRED DURING NORMAL USAGE OF DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23713647 ↗
  • MalfunctionDec 4, 2025Material Deformation; Material Deformation

    INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, DISTRIBUTOR) REGARDING A PATIENT HAVING SPINAL DEFORMITY CORRECTION PROCEDURE FOR THORACOLUMBAR SCOLIOSIS. IT WAS REPORTED THAT DIFFICULTIES AROSE WHILE INSERTING THE RODS DUE TO THE SEVERITY OF THE SCOLIOSIS. THREAD DAMAGE (STRIPPED THREADS) OCCURRED IN LOCKING SCREWS, WHICH WERE SUBSEQUENTLY DEEMED UNUSABLE AND REPLACED WITH NEW ONES. THERE WERE NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THE REPORTED EVENTOCCURRED DURING NORMAL USAGE OF DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23713620 ↗
  • MalfunctionDec 4, 2025Material Deformation; Material Deformation

    INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, DISTRIBUTOR) REGARDING A PATIENT HAVING SPINAL DEFORMITY CORRECTION PROCEDURE FOR THORACOLUMBAR SCOLIOSIS. IT WAS REPORTED THAT DIFFICULTIES AROSE WHILE INSERTING THE RODS DUE TO THE SEVERITY OF THE SCOLIOSIS. THREAD DAMAGE (STRIPPED THREADS) OCCURRED IN LOCKING SCREWS, WHICH WERE SUBSEQUENTLY DEEMED UNUSABLE AND REPLACED WITH NEW ONES. THERE WERE NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THE REPORTED EVENT OCCURRED DURING NORMAL USAGE OF DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23713615 ↗
  • MalfunctionDec 3, 2025Device Slipped; Device Slipped

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING POSTERIOR LUMBAR SPINE THERAPY. IT WAS REPORTED THAT THE CLINICIAN HAD A THREAD SLIP WHEN SCREWING THE PEDICLE SCREW INTO THE SET SCREW, WHICH MADE IT IMPOSSIBLE TO SCREW IT IN. LATER, TWO NEW SET SCREWS WERE REPLACED DURING THE OPERATION.. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23702973 ↗
  • MalfunctionDec 2, 2025Device Slipped; Device Slipped

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING POSTERIOR LUMBAR SPINE THERAPY. IT WAS REPORTED THAT THE CLINICIAN HAD A THREAD SLIP WHEN SCREWING THE PEDICLE SCREW INTO THE SET SCREW, WHICH MADE IT IMPOSSIBLE TO SCREW IT IN. LATER, TWO NEW SET SCREWS WERE REPLACED DURING THE OPERATION.. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23692968 ↗
  • InjuryAug 20, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT THE SET SCREWS WERE LOOSE AND THERE WAS LOOSENING OF THE STRUCTURE. THE PATIENT WENT REVISION SURGERY ON (B)(6) 2025 AS A RESULT OF THE EVENT AND SCREWS WERE EXPLANTED TO INSERT SOLERA 5.5/6.0 IMPLANT. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22847255 ↗
  • InjuryAug 20, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT THE SET SCREWS WERE LOOSE AND THERE WAS LOOSENING OF THE STRUCTURE. THE PATIENT WENT REVISION SURGERY ON (B)(6) 2025 AS A RESULT OF THE EVENT AND SCREWS WERE EXPLANTED TO INSERT SOLERA 5.5/6.0 IMPLANT. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22847247 ↗
  • InjuryAug 20, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT THE SET SCREWS WERE LOOSE AND THERE WAS LOOSENING OF THE STRUCTURE. THE PATIENT WENT REVISION SURGERY ON (B)(6) 2025 AS A RESULT OF THE EVENT AND SCREWS WERE EXPLANTED TO INSERT SOLERA 5.5/6.0 IMPLANT. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22847243 ↗
  • InjuryAug 20, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT THE SET SCREWS WERE LOOSE AND THERE WAS LOOSENING OF THE STRUCTURE. THE PATIENT WENT REVISION SURGERY ON (B)(6) 2025 AS A RESULT OF THE EVENT AND SCREWS WERE EXPLANTED TO INSERT SOLERA 5.5/6.0 IMPLANT. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22847230 ↗
  • MalfunctionAug 5, 2025Defective Device; Defective Device

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING POSTERIOR OPEN THORACOLUMBAR SPINE PROCEDURE, POSTERIOR OPEN DIRECT DECOMPRESSION FOR LUMBAR POSTERIOR DISC HERNIATION. IT WAS REPORTED THAT WHEN SCREWING THE SET SCREW DURING THE OPERATION, THERE WAS AN OBVIOUS DEFECT IN THE PRODUCT THREAD AND THE SET SCREW COULD NOT BE SCREWED. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22710899 ↗
  • MalfunctionAug 5, 2025Defective Device; Defective Device

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING POSTERIOR OPEN THORACOLUMBAR SPINE PROCEDURE, POSTERIOR OPEN DIRECT DECOMPRESSION FOR LUMBAR POSTERIOR DISC HERNIATION. IT WAS REPORTED THAT WHEN SCREWING THE SET SCREW DURING THE OPERATION, THERE WAS AN OBVIOUS DEFECT IN THE PRODUCT THREAD AND THE SET SCREW COULD NOT BE SCREWED. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22710872 ↗
  • MalfunctionApr 22, 2025Device Slipped; Material Deformation; Device Slipped; Material Deformation

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH PRE-OPERATIVE DIAGNOSIS FOR LUMBAR SPINE DEGENERATION. PROCEDURE OR TECHNIQUE USED WAS PLIF (POSTERIOR LUMBAR). PATIENT HAD A MEDICAL HISTORY OF SMOKING. IT WAS REPORTED THAT DURING THE IMPLANTATION PROCESS OF THE SET SCREW, OCCURRED DEFORMATION SLIPPERY THREAD, SO THAT THE LOCKING COULD NOT BE COMPLETED, AND THE IMPLANTATION PROCESS WAS LIMITED. THERE WAS NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21880474 ↗
  • MalfunctionMar 15, 2025Material Deformation; Material Deformation

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING SPINAL PRODUCT THAT WILL BE USED IN POSTERIOR LUMBAR INTERBODY FUSION SPINAL CANAL DECOMPRESSION SITE FOR LUMBAR SPINAL STENOSIS. IT WAS REPORTED THAT DURING THE IMPLANTATION PROCESS OF THE SET SCREW, THE THREAD WAS SLIPPERY DUE TO DEFORMATION AND THE LOCKING COULD NOT BE COMPLETED, SO THE IMPLANTATION PROCESS WAS LIMITED. THERE WERE NO FURTHER COMPLICATIONS OR SYMPTOMS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21613315 ↗

FDA clearance and classification

Cleared through Special 510(k) K052187 on Aug 22, 2005 (decision: Substantially Equivalent). Submitted by Medtronic Sofamor Danek as “Modification To: Cd Horizon Spinal System”. The FDA review took 11 days. FDA record ↗

The FDA regulates product code MNH (orthosis, spondylolisthesis spinal fixation) under 21 CFR 888.3070 as a Class II device. 21 CFR 888.3070 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the CD HORIZON® Spinal System 7540020

Has the CD HORIZON® Spinal System 7540020 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00885074457093), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the CD HORIZON® Spinal System 7540020?

The FDA's MAUDE database lists 257 adverse event reports for this device, received between Nov 15, 2016 and Sep 22, 2026. In those reports, the most common specific device problems are “Device Slipped” (90), “Failure to Align” (76) and “Material Deformation” (26); the most common patient problems are “Failure of Implant” (10) and “Pain” (7). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CD HORIZON® Spinal System 7540020?

None of the reports are classified as deaths, 32 as injuries and 225 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CD HORIZON® Spinal System 7540020 FDA approved?

It was cleared through Special 510(k) K052187 on Aug 22, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the CD HORIZON® Spinal System 7540020 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the CD HORIZON® Spinal System 7540020, and is it still sold?

The FDA's device database lists Medtronic Sofamor Danek, Inc. as the company responsible for it (brand CD HORIZON® Spinal System, model 7540020). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Orthosis, spondylolisthesis spinal fixation devices

All Orthosis, spondylolisthesis spinal fixation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page