MANTIS Redux 48289999Recalls & reported problems
Stryker Corporation · LITE BLOCKER · Class 2 · Implant
All models of MANTIS Redux (6)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker MANTIS Redux spinal fixation blocker has 77 adverse event reports, describing no deaths and 42 injuries, often with the device migrating, dislodging, or slipping after implantation, sometimes requiring revision surgery. 35 reports mention malfunction, including a screw tab fracturing during a blocker insertion. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 76
- Deaths reported
- 0
- Injuries reported
- 41
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 76 adverse event reports that list this exact device (DI 04546540671431), received between Feb 23, 2016 and Mar 24, 2026 — classified as 41 injuries, 35 malfunctions.
The most frequently reported specific device problem is “Migration” (33 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Migration33
- Loose or Intermittent Connection19
- Device Dislodged or Dislocated15
- Device Slipped15
- Migration or Expulsion of Device10
- Mechanical Problem9
- Adverse Event Without Identified Device or Use Problem8
- Detachment of Device or Device Component5
Patient problems
- Injury42
- No Clinical Signs, Symptoms or Conditions21
- No Code Available15
- No Known Impact Or Consequence To Patient14
- Pain7
- Reaction6
- Failure of Implant4
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 25 most recent of 76 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 24, 2026Migration
A HEALTH PROFESSIONAL REPORTED THAT A MANTIS REDUX BLOCKER AND AN ES2 ROD MIGRATED POST-OPERATIVELY. REVISION SURGERY HAS NOT TAKEN PLACE NOR BEEN SCHEDULED. NO ADVERSE CONSEQUENCES WERE REPORTED. THIS REPORT REPRESENTS THE MANTIS REDUX BLOCKER MIGRATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24676733 ↗MalfunctionFeb 27, 2026Migration
A HEALTH PROFESSIONAL REPORTED THAT A MANTIS REDUX BLOCKER MIGRATED APPROXIMATELY TWO MONTHS POST-OPERATIVELY. REVISION SURGERY HAS NOT BEEN PERFORMED NOR SCHEDULED AT THIS TIME. NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24477451 ↗InjuryAug 4, 2025Migration; Migration
IT WAS FURTHER REPORTED THAT THE PATIENT "HAD BEEN MOVING AFTER THE SURGERY, NOT FOLLOWING THE DOCTOR'S INSTRUCTIONS TO LIMIT HIS ACTIVITIES".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22704640 ↗InjuryJun 17, 2025Migration; Migration
IT WAS REPORTED THAT A PATIENT UNDERWENT REVISION SURGERY APPROXIMATELY 6 MONTHS POST-OPERATIVELY TO ADDRESS A MIGRATED MANTIS REDUX BLOCKER.
Patient problems: No Clinical Signs, Symptoms or Conditions; Pain
MAUDE report 22243522 ↗MalfunctionNov 27, 2024Difficult to Insert; Migration; Difficult to Insert; Migration
IT WAS REPORTED THAT THE TAB OF AN ES2 AUGMENTABLE LONG BLADE POLYAXIAL SCREW FRACTURED OFF WHEN INSERTING A MANTIS REDUX BLOCKER, AND THAT THE BLOCKER WAS DISPLACED OUT OF THE SCREW INTRA-OPERATIVELY. THE BLOCKER WAS REMOVED FROM THE PATIENT, RESULTING IN A 20 MINUTE SURGICAL DELAY, AND THE SURGEON DECIDED TO LEAVE THE SCREW WITHOUT A BLOCKER INSERTED. THIS REPORT IS FOR THE MANTIS BLOCKER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20793754 ↗MalfunctionJul 22, 2024Migration; Migration
A COMPANY REPRESENTATIVE REPORTED THAT A MANTIS REDUX BLOCKER AT L2 MIGRATED POST-OPERATIVELY. NO ADVERSE CONSEQUENCES WERE REPORTED. THE MIGRATED BLOCKER WAS OBSERVED IN AN X-RAY TAKEN FOR A PLANNED REVISION SURGERY TO REMOVE ALL IMPLANTS AS THE PATIENT WAS REPORTED TO HAVE ACHIEVED GOOD BONE UNION. THE BLOCKER WAS EXPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19799126 ↗InjuryJun 11, 2024Migration; Migration
IT WAS REPORTED THAT A PATIENT WAS REVISED TO ADDRESS A MIGRATED MANTIS REDUX BLOCKER AT L3 ONE DAY AFTER IMPLANTATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19509648 ↗MalfunctionMay 14, 2024Migration; Migration
IT WAS REPORTED THAT APPROXIMATELY TWO WEEKS POST-OPERATIVELY A MANTIS REDUX BLOCKER AT LEFT L5 MIGRATED. REVISION SURGERY IS NOT SCHEDULED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19311124 ↗InjuryFeb 26, 2024Migration; Migration
A PATIENT WAS REVISED TO ADDRESS A MIGRATED MANTIS REDUX BLOCKER AT L3 APPROXIMATELY SEVEN WEEKS AFTER IMPLANTATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18785021 ↗InjuryMay 12, 2023Migration; Migration
A COMPANY REPRESENTATIVE REPORTED THAT A PATIENT WAS REVISED TO ADDRESS A FRACTURED XIA 3 POLYAXIAL SCREW AT RIGHT-SIDE S1 AND A MIGRATED MANTIS REDUX BLOCKER AT LEFT-SIDE S1. THE PATIENT REPORTED EXPERIENCING LOW BACK PAIN AND INSTABILITY APPROXIMATELY TWO MONTHS POST-OPERATIVELY, AND IMAGING IDENTIFIED THE FRACTURE AND MIGRATION. THIS REPORT IS FOR THE MANTIS REDUX BLOCKER.
Patient problems: Pain
MAUDE report 16919798 ↗InjuryMay 3, 2023Migration; Migration
IT WAS REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY 2.5 MONTHS POST-OPERATIVELY TO ADDRESS A MIGRATED MANTIS REDUX BLOCKER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16865147 ↗InjuryDec 20, 2022Migration; Migration
IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION SURGERY TO ADDRESS A MIGRATED MANTIS REDUX BLOCKER (NOTICED VIA A CT SCAN) APPROXIMATELY 11 DAYS AFTER IMPLANTATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16014774 ↗InjurySep 24, 2022Migration; Migration
IT WAS REPORTED THAT A PATIENT WAS REVISED TO ADDRESS FOUR MIGRATED MANTIS REDUX BLOCKERS APPROXIMATELY TWO MONTHS AFTER IMPLANTATION. THIS REPORT CAPTURES THE THIRD OF THE FOUR MIGRATED BLOCKERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15481149 ↗InjurySep 24, 2022Migration; Migration
IT WAS REPORTED THAT A PATIENT WAS REVISED TO ADDRESS FOUR MIGRATED MANTIS REDUX BLOCKERS APPROXIMATELY TWO MONTHS AFTER IMPLANTATION. THIS REPORT CAPTURES THE FOURTH OF THE FOUR MIGRATED BLOCKERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15480793 ↗InjurySep 24, 2022Migration; Migration
IT WAS REPORTED THAT A PATIENT WAS REVISED TO ADDRESS FOUR MIGRATED MANTIS REDUX BLOCKERS APPROXIMATELY TWO MONTHS AFTER IMPLANTATION. THIS REPORT CAPTURES THE SECOND OF THE FOUR MIGRATED BLOCKERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15480714 ↗InjurySep 24, 2022Migration; Migration
IT WAS REPORTED THAT A PATIENT WAS REVISED TO ADDRESS FOUR MIGRATED MANTIS REDUX BLOCKERS APPROXIMATELY TWO MONTHS AFTER IMPLANTATION. THIS REPORT CAPTURES THE FIRST OF THE FOUR MIGRATED BLOCKERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15480690 ↗InjuryAug 24, 2021Migration; Migration
IT WAS REPORTED THAT ON CT IMMEDIATELY AFTER SURGERY, SHOWED THAT THE SCREW BLOCKER OF THE LEFT S2 WAS COMING LOOSE. THE PATIENT WAS TAKEN BACK INTO THE SURGICAL SUITE AND ANESTHESIA WAS ADDED, THE WOUND WAS REOPENED, AND THE SCREW AND BLOCKER WERE REPLACED WITH NEW ONES. THIS LED TO A 20-30 MINUTE DELAY. THIS REPORT WILL COVER THE BLOCKER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12362979 ↗MalfunctionMay 24, 2021Use of Device Problem; Migration; Use of Device Problem; Migration
IT WAS REPORTED THAT DURING THE PROCEDURE, A MANTIS REDUX BLOCKER WAS INADVERTENTLY LEFT IN THE PATIENT. AN XRAY WAS PERFORMED AND AN ADDITIONAL SET SCREW WAS OBSERVED. REVISION IS SCHEDULED IN THE FUTURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11872852 ↗InjuryApr 2, 2021Material Separation; Migration; Material Separation; Migration
IT WAS REPORTED THAT SIX DAYS AFTER THE INITIAL SURGERY, AN ES2 BLOCKER MIGRATED FROM ITS RESPECTIVE SCREW. THE SURGEON PLANS ON REVISING THE PATIENT. UPDATE: PATIENT WAS REVISED ON (B)(6) 2021.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11610891 ↗MalfunctionOct 19, 2020Migration; Migration
IT WAS REPORTED THAT A PATIENT RECEIVED REVISION SURGERY APPROXIMATELY FOUR MONTHS POST OPERATIVELY DUE TO A LOOSE SCREW. DURING THE REVISION SURGERY, THE SURGEON FOUND THAT AN ES2 INTEGRATED BLADE SCREW TULIP DISENGAGED AND A MANTIS REDUX BLOCKER MIGRATED. THE SCREWS WERE THEN REPLACED AND THE CONSTRUCT WAS EXTENDED. THIS REPORT REPRESENTS THE MANTIS REDUX BLOCKER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10701964 ↗InjurySep 29, 2020Migration; Migration
A PHYSICIAN REPORTED THAT A MANTIS REDUX BLOCKER MIGRATED TWO DAYS AFTER IMPLANTATION AND THAT, "THE BLOCKER MAY HAVE BEEN SLANTED WITH RESPECT TO THE ROD WHEN THE BLOCKER WAS INSERTED." THE PATIENT HAS BEEN REVISED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10602833 ↗InjuryAug 10, 2020Adverse Event Without Identified Device or Use Problem; Migration; Adverse Event Without Identified Device or Use Problem; Migration
IT WAS REPORTED THAT A PATIENT PRESENTED TO THE HOSPITAL WITH "COMPLAINTS". UPON REVIEW OF XRAY, IT WAS OBSERVED THAT TWO XIA 3 TITANIUM BLOCKERS HAD COME LOOSE FROM THE SCREW HEAD. REVISION SURGERY HAS BEEN PERFORMED. THIS REPORT CAPTURES THE SECOND OF TWO BLOCKERS.
Patient problems: Injury; No Clinical Signs, Symptoms or Conditions
MAUDE report 10384493 ↗InjuryAug 10, 2020Migration; Migration
IT WAS REPORTED THAT A PATIENT PRESENTED TO THE HOSPITAL WITH "COMPLAINTS". UPON REVIEW OF XRAY, IT WAS OBSERVED THAT TWO XIA 3 TITANIUM BLOCKERS HAD COME LOOSE FROM THE SCREW HEAD. REVISION SURGERY HAS BEEN PERFORMED. THIS REPORT CAPTURES THE FIRST OF TWO BLOCKERS.
Patient problems: Failure of Implant; Injury
MAUDE report 10384492 ↗InjuryMay 22, 2020Migration; Migration
IT WAS REPORTED THAT TWO MANTIS SCREWS AND A MANTIS REDUX BLOCKER MIGRATED POST-OPERATIVELY. TWO MONTHS AFTER INITIAL SURGERY, THE PATIENT EXPERIENCED A BONE FRACTURE AFTER EXCESSIVE HIP MOVEMENT/ROTATION THAT CAUSED THE TWO SCREWS AND BLOCKER TO BECOME LOOSE. THE PATIENT DID NOT FUSE AND REVISION SURGERY WAS PERFORMED. THIS REPORT REPRESENTS THE BLOCKER.
Patient problems: Injury
MAUDE report 10083172 ↗InjuryMar 20, 2020Migration; Migration
IT WAS REPORTED THAT A MANTIS REDUX BLOCKER MIGRATED POST-OPERATIVELY. THE PATIENT BEGAN TO HEAR NOISE AND EXPERIENCE PAIN 2 MONTHS AFTER INITIAL SURGERY. REVISION SURGERY WAS PERFORMED.
Patient problems: Injury; Pain
MAUDE report 9861800 ↗
FDA clearance and classification
The FDA regulates product code MNH (orthosis, spondylolisthesis spinal fixation) under 21 CFR 888.3070 as a Class II device. 21 CFR 888.3070 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the MANTIS Redux 48289999
Has the MANTIS Redux 48289999 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04546540671431), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the MANTIS Redux 48289999?
The FDA's MAUDE database lists 76 adverse event reports for this device, received between Feb 23, 2016 and Mar 24, 2026. In those reports, the most common specific device problems are “Migration” (33), “Loose or Intermittent Connection” (19) and “Device Dislodged or Dislocated” (15); the most common patient problems are “Injury” (42), “Pain” (7) and “Reaction” (6). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the MANTIS Redux 48289999?
None of the reports are classified as deaths, 41 as injuries and 35 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the MANTIS Redux 48289999 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the MANTIS Redux 48289999, and is it still sold?
The FDA's device database lists Stryker Corporation as the company responsible for it (brand MANTIS Redux, model 48289999). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Orthosis, spondylolisthesis spinal fixation devices
- CD HORIZON® Spinal System 7540020Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation257 reports
- EXPEDIUM TOP NOTCH 179771555Depuy Spine, LLC · Orthosis, spondylolisthesis spinal fixation41 reports
- EXPEDIUM 179712650Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation31 reports
- XIA 4 5 48130000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation29 reports
- EXPEDIUM VERSE 199721530Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation29 reports
- XIA 4.5 System - XIA CT 48850000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation27 reports
- CD HORIZON® Spinal System 7576545Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation26 reports
- DENALI® Spinal System 101-10001VB Spine LLC · Orthosis, spondylolisthesis spinal fixation24 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page