DeVilbiss iFill 535IRecalls & reported problems
Devilbiss Healthcare LLC · Portable Oxygen Station · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885304000860), received between Dec 2, 2016 and Dec 2, 2016 — classified as 1 malfunction.
The most frequently reported specific device problem is “Maintenance Does Not Comply To Manufacturers Recommendations” (1 report).
1 other recall cover devices cleared under the same FDA submission (K053240); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Maintenance Does Not Comply To Manufacturers Recommendations1
Patient problems
- No Patient Involvement1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 2, 2016Maintenance Does Not Comply To Manufacturers Recommendations; Maintenance Does Not Comply To Manufacturers Recommendations
THE COMPLAINT WAS REPORTED FROM A LONG TERM CARE FACILITY IN FRANCE. THERE IS NO ADVERSE EVENT OR INJURY ASSOCIATED WITH THE COMPLAINT. A NURSE WAS REMOVING/DISCONNECTING A FULL IFILL OXYGEN CYLINDER FROM THE IFILL OXYGEN TRANSFILL UNIT, AT WHICH TIME SHE HEARD A LOUD SOUND AND SAW A BRIEF FLAME COMING FROM THE OXYGEN CONNECTING PORT. THE NURSE IS UNSURE IF THE IFILL UNIT WAS RUNNING OR HAD STOPPED AT THE TIME SHE DISCONNECTED CYLINDER BUT IT NOTED THE IFILL UNIT WAS PLUGGED INTO THE WALL. THE NURSE STATED SHE HEARD A LOUD SOUND AND SAW A BRIEF FLAME, WHICH CAUSED THE NURSE TO DROP THE CYLINDER TO THE FLOOR. ONCE THE CYLINDER WAS ON THE FLOOR THE EVENT ENDED; SHE REPORTED NO OTHER ISSUES WITH THE CYLINDER OR THE IFILL UNIT. THE CYLINDER AND TRANSFILL DEVICE WERE QUARANTINED AND THE OXYGEN SUPPLIER CONTACTED. THE FACILITY ACKNOWLEDGES THE ROUTINE USE OF CLEANING AGENT ON THE IFILL CYLINDER THAT CONTAINS HIGH CONCENTRATIONS OF PROPRANOLOL AND ETHANOL. THE CLEANING AGENT IS HIGHLY FLAMMABLE AND IS CONTRAINDICATED FOR USE WITH OXYGEN EQUIPMENT. THERE WERE NO INJURIES OR DAMAGE RELATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6142736 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K053240. One submission often covers many models, so check the product and lot details in each record.
- Z-0038-2007 · Aug 24, 2006 · TerminatedDeVilbiss iFill Personal Oxygen Station model number 535D.
Other Generator, Oxygen, Portable devices
- DeVilbiss 525DSDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls14 deaths287 reports
- Inogen One G5 Oxygen Concentrator - Bluetooth 1260 BluetoothInogen, Inc. · Generator, Oxygen, Portable3 deaths235 reports
- EverFlo EverFlo w/O2 Percentage Indicator, 230V, EURespironics, Inc. · Generator, oxygen, portable5 deaths112 reports
- Inogen One G3 Oxygen Concentrator 1050Inogen, Inc. · Generator, Oxygen, Portable2 deaths111 reports
- DeVilbiss 125DDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls1 death3 reports
- EverFlo EVERFLO NON-O2 Percentage Indicator, 120V, US,CARespironics, Inc. · Generator, oxygen, portable19 deaths80 reports
- EverFlo EVERFLO w/Oxygen Percentage Indicator, 120V US, CARespironics, Inc. · Generator, oxygen, portable2 deaths71 reports
- SimplyGo Mini SimplyGo Mini, USRespironics, Inc. · Generator, oxygen, portable1 recall1 death5 reports