Skip to content

DeVilbiss DV53DRecalls & reported problems

Devilbiss Healthcare LLC · IntelliPAP Standard · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885304001874), received between Aug 4, 2023 and Aug 4, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fire” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Fire1

Patient problems

  • Burn(s)1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 4, 2023Fire; Fire

    DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A CPAP BY AN END-USER, WHO REPORTED THAT THE POWER CORD FOR HIS UNIT "GOT HOT, SMOKING AND ON FIRE", AND HE SUFFERED BURNS TO HIS FINGERS. THE END-USER NOTED THE CPAP WAS PLUGGED INTO A "POWER STRIP" WITH NUMEROUS OTHER DEVICES. HE STATED THE EVENT OCCURRED WHEN HE PARTIALLY PULLED THE CPAP PLUG FROM THE STRIP AND THEN RE-INSERTED IT. THE END-USER DID NOT SEEK MEDICAL TREATMENT FOR HIS INJURY. DEVILBISS REQUESTED THE UNIT BE RETURNED FOR INVESTIGATION AND WILL FILE AN UPDATE IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

    Patient problems: Burn(s)

    MAUDE report 17462372 ↗

Other Ventilator, Non-Continuous (Respirator) devices

All Ventilator, Non-Continuous (Respirator) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page