Why it was recalled

Right nails were anodized, marked, and labelled as left nails and vice versa

FDA root cause category: Under Investigation by firm

What the company is doing

The firm sent out a recall notification letter via email and overnight mail on 04/22/2022. The letter instructs the consignee to check inventory, return affected product, and complete the response form. Distributors were informed to notify their customers of the field action.

Recall details

Recall number
Z-1133-2022
Status
Open, Classified
Initiated by the company
Apr 22, 2022
Recalling firm
Smith & Nephew Orthopaedics GmbH
Quantity
17 units
Product code
JDS
FDA submissions
K040212
Distribution
Distributed nationwide to GA, OH, IN, TX, CO, MO, TN, CA, NC, PA and internationally to Canada, Australia.

Product description

INTERTAN 1.5 NAIL 13mm X 42cm 130DEG. Left INTERTAN 1.5 NAIL 13mm X 42cm 130DEG. Right

Code information (models, lots, serial numbers)

Product number: 71676627 (DI: 00885556131633), Batch Number: 21LT56988; 71676628 (DI: 00885556040027), Batch number: 21LT56989

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

82 devices cleared under K040212 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page