TRIGEN 71642685Recalls & reported problems
Smith & Nephew, Inc. · META-TAN LAG/COMPRESSION SCREW KIT 85MM/80MM · Class 2 · Implant
All models of TRIGEN (381)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556132982), received between May 12, 2021 and Sep 20, 2024 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Markings/Labelling Problem” (1 report).
4 other recalls cover devices cleared under the same FDA submission (K092748, K210980); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Device Markings/Labelling Problem1
- Off-Label Use1
- Unintended Movement1
Patient problems
- Implant Pain2
- Bone Fracture(s)1
- Respiratory Tract Infection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 20, 2024Off-Label Use; Adverse Event Without Identified Device or Use Problem; Off-Label Use; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT SUFFERED A FRACTURE IN THE LEFT FEMUR DUE TO A FALL. AN INTERNAL FIXATION WAS PERFORMED ON (B)(6) 2024 DURING WHICH THE SURGEON INSERTED AN INTERTAN 11.5MM X 20CM 125D (71675205) WITH META-TAN LAG/COMP KIT 85/80 (71642685). AFTER THE SURGERY, AT THE VISIT TO THE CLINIC ON (B)(6) 2024, THE PATIENT COMPLAINED OF PAIN IN HER LEFT THIGH AND HIP JOINT. AN X-RAY REVEALED HARDWARE CUTOUT AND A FEMORAL FRACTURE AT THE BASE OF THE NECK. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2024 TO REMOVE THE INTERTAN NAIL AND META-TAN SCREWS AND CONVERT TO A THR.
Patient problems: Bone Fracture(s); Implant Pain
MAUDE report 20274596 ↗InjuryMay 12, 2021Device Markings/Labelling Problem; Unintended Movement; Device Markings/Labelling Problem; Unintended Movement
IT WAS REPORTED THAT AFTER TRAUMA PROCEDURE FOR INTERNAL HIP FIXATION, THE PATIENT HAD POSTOPERATIVE PAIN ON THE HIP JOINT, COPD EXACERBATION AND LUNG INFECTION. X-RAY EXAMINATION INDICATED GOOD POSITIONING OF THE IMPLANT. TWO DAYS AFTER THE PROCEDURE, IT WAS DISCOVERED THAT TRIGEN META-TAN TROCH ANTEGR NAIL SCREWS WERE USE TO FIXATE A TRIGEN INTERTAN INTERTROCH ANTEGR NAIL, CAUSING THE IMPLANT TO BE UNSTABLE. A REVISION SURGERY HAD TO BE PERFORMED IN ORDER TO RETRIEVE THE WRONG SCREWS AND INSERT THE CORRECT ONES. DETAILS ABOUT THE SURGERY ARE UNKNOWN. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: Implant Pain; Respiratory Tract Infection
MAUDE report 11815676 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K092748, K210980. One submission often covers many models, so check the product and lot details in each record.
- Z-2456-2026 · Mar 16, 2026 · Open, ClassifiedTRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
- Z-1806-2025 · Mar 18, 2025 · Open, ClassifiedMETA-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
- Z-2443-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail
- Z-2444-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page