Skip to content

TRIGEN 71642605Recalls & reported problems

Smith & Nephew, Inc. · META-TAN LAG/COMPRESSION SCREW KIT 105MM/100MM · Class 2 · Implant

All models of TRIGEN (381)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556133071), received between Jul 23, 2021 and Feb 5, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Fracture” (1 report).

4 other recalls cover devices cleared under the same FDA submission (K092748, K210980); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1
  • Improper or Incorrect Procedure or Method1
  • Migration or Expulsion of Device1

Patient problems

  • Failure of Implant1
  • Laceration(s)1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 5, 2025Fracture; Migration or Expulsion of Device; Fracture; Migration or Expulsion of Device

    IT WAS REPORTED THAT, AFTER AN INTERNAL FIXATION WAS PERFORMED IN (B)(6) 2023, THE FRACTURE DYNAMIZED, LEAVING THE LAG SCREW PROUD OVER THE LATERAL FEMUR. THIS ADVERSE EVENT WAS SOLVED BY A REVISION SURGERY ON (B)(6) 2025, IN WHICH THE LAG SCREW WAS REMOVED AND A SHORTED SCREW WAS INSERTED. DURING THE REMOVAL PROCEDURE, THE HEAD OF THE LAG SCREW BROKE OFF. DRILLING ALONG THE INFERIOR LAG SCREW TRACT WAS PERFORMED TO REMOVE ANY RESIDUAL SURROUNDING BONE. THE SCREW WAS ULTIMATELY REMOVED WITH VICE GRIPS. PATIENT IS DOING WELL.

    Patient problems: Failure of Implant; Laceration(s)

    MAUDE report 21314195 ↗
  • InjuryJul 23, 2021Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT, AFTER A INTRAMEDULLARY SURGERY ON (B)(6) 2021, A REVISION SURGERY WAS MADE ON (B)(6) 2021 BECAUSE AN INCORRECT METATAN INTERGRATED LAG WAS SCREWED INTO THE INTERTAN CONSTRUCT (BOXES BETWEEN INTERTAN & METATAN LAG SCREWS ARE ALMOST IDENTICAL TO LOOK AT). CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12216949 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K092748, K210980. One submission often covers many models, so check the product and lot details in each record.

  • Z-2456-2026 · Mar 16, 2026 · Open, Classified
    TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
  • Z-1806-2025 · Mar 18, 2025 · Open, Classified
    META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
  • Z-2443-2023 · Jul 24, 2023 · Open, Classified
    TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail
  • Z-2444-2023 · Jul 24, 2023 · Open, Classified
    TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates