TRIGEN 71642600Recalls & reported problems
Smith & Nephew, Inc. · META-TAN LAG/COMPRESSION SCREW KIT 100MM/95MM · Class 2 · Implant
All models of TRIGEN (381)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885556133088), received between Aug 6, 2020 and Aug 11, 2021 — classified as 3 injuries.
The most frequently reported specific device problem is “Break” (1 report).
4 other recalls cover devices cleared under the same FDA submission (K092748, K210980); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Improper or Incorrect Procedure or Method1
- Mechanical Jam1
- Osseointegration Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Failure of Implant1
- Foreign Body In Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryAug 11, 2021Mechanical Jam; Osseointegration Problem; Mechanical Jam; Osseointegration Problem
IT WAS REPORTED THAT, AFTER SURGERY HAD BEEN PERFORMED ON (B)(6) 2019, AND BEFORE A TRIGEN META TAN REMOVAL, IT WAS DISCOVERED THAT IS NOT POSSIBLE TO REMOVE THE LAG SCREW OUT DUE TO NO INSERTION OF A COMPRESSION SCREW. IT WAS REPORTED THAT THE EVENT WILL NOT BE SOLVED IN ORDER TO PROTECT THE PATIENT FROM RE-FRACTURING THE FEMUR. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12311048 ↗InjuryJul 1, 2021Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT AFTER AN INTERNAL FIXATION SURGERY, PERFORMED ON (B)(6) 2021, SCREWS OF META-TAN LAG/COMP KIT 100/95 ((B)(4)) WERE USED THROUGH AN INTERTAN 11.5MM X 20CM 125D ((B)(4)), LEADING TO EARLY IMPLANT FAILURE. WITH SOME FURTHER INVESTIGATION, IT WAS DISCOVERED THAT THERE WAS NO SALES REP PRESENT AT THE ORIGINAL OPERATION, AND THE ERROR WAS MADE BY A JUNIOR CLINICIAN AT POINT OF SURGERY. THE SURGEON AND HOSPITAL DO NOT BLAME THE IMPLANTS AT ALL, AND THE METATAN SCREWS HAVE NOW BEEN MORE CLEARLY LABELLED FOR FUTURE REFERENCE. THIS ISSUE WAS RESOLVED WITH A REVISION PERFORMED ON (B)(6) 2021.
Patient problems: Failure of Implant
MAUDE report 12105181 ↗InjuryAug 6, 2020Break; Break
IT WAS REPORTED THAT DURING A PLANNED SURGERY FOR NAIL REMOVAL, THE THREADED PART OF THE DEVICES DEFORMED/BROKE. THE TIPS FROM THE DEVICES DO NOT REMAIN IN THE PATIENT'S BODY, HOWEVER THE SLOTTED PART OF THE LAG SCREW ALSO BROKE OFF AND THE PIECES REMAIN INSIDE THE PATIENT BODY. IT WAS ATTEMPTED TO REMOVED THE PIECES USING AN EASY OUT BUT IT DID NOT MATE TO THE LAG SCREW. THE PROCEDURE WAS CONCLUDED WITHOUT REMOVING THE NAIL AND LAG SCREW. THE PIECES FROM LAG SCREW REMAINS. A 60 MIN DELAY WAS REPORTED. PATIENT OUTCOME IS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 10376753 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K092748, K210980. One submission often covers many models, so check the product and lot details in each record.
- Z-2456-2026 · Mar 16, 2026 · Open, ClassifiedTRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
- Z-1806-2025 · Mar 18, 2025 · Open, ClassifiedMETA-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
- Z-2443-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail
- Z-2444-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.