TRIGEN 71645027Recalls & reported problems
Smith & Nephew, Inc. · TRIGEN LOW PROFILE SCREW 5.0MM X 27.5MM · Class 2 · Implant
All models of TRIGEN (381)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00885556215548), received between Jan 4, 2018 and Nov 26, 2025 — classified as 9 injuries.
The most frequently reported specific device problem is “Break” (3 reports).
5 other recalls cover devices cleared under the same FDA submission (K111025, K161264, K210980); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Fracture3
- Adverse Event Without Identified Device or Use Problem2
- Insufficient Device Problem Information1
- Material Twisted/Bent1
Patient problems
- Bone Fracture(s)3
- Device Embedded In Tissue or Plaque3
- Injury2
- Pain1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryNov 26, 2025Fracture; Material Twisted/Bent; Fracture; Material Twisted/Bent
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, WHEN INSERTING ONE (1) TRIGEN LOW PROFILE SCREW 5.0MM X 27.5MM, THE SCREW HEAD BECAME STRIPPED PREVENTING ROTATION. ATTEMPTS TO REMOVE THE SCREW USING PLIERS RESULTED IN FRACTURE OF THE SCREW HEAD, MAKING EXTRACTION DIFFICULT, THE FRACTURED HEAD WAS REMOVED. THE SURGEON DECIDED TO LEAVE THE SCREW INSIDE THE BODY. THE PROCEDURE WAS RESUMED, AFTER A 2 HOUR SURGICAL DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 23657646 ↗InjuryJan 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING SURGERY, AFTER INSERTING THE TRIGEN LOW PROFILE SCREW 5.0MM X 27.5MM, AN INCOMPLETE BONE FRACTURE WAS IDENTIFIED. THE PROCEDURE WAS RESUMED WITHOUT DELAY, USING THE SAME DEVICE.
Patient problems: Bone Fracture(s)
MAUDE report 21174990 ↗InjuryApr 8, 2020Fracture; Fracture
IT WAS REPORTED THAT A FEMALE PATIENT HAD AN INTERTAN SURGERY. AFTER THE SURGERY SURGEON FOUND OUT THERE WAS FRACTURE AROUND DISTAL SCREWS, HE THINKS THE DIAMETER OF THE PART UNDER THE SCREW HEAD IS WIDE AND THE DIAMETER OF DRILL IS NOT FIT FOR IT.
Patient problems: Injury
MAUDE report 9938481 ↗InjuryApr 8, 2020Fracture; Fracture
IT WAS REPORTED THAT A FEMALE PATIENT HAD AN INTERTAN SURGERY. AFTER THE SURGERY SURGEON FOUND OUT THERE WAS FRACTURE AROUND DISTAL SCREWS, HE THINKS THE DIAMETER OF THE PART UNDER THE SCREW HEAD IS WIDE AND THE DIAMETER OF DRILL IS NOT FIT FOR IT.
Patient problems: Injury
MAUDE report 9938149 ↗InjuryNov 5, 2019Break; Break
IT WAS REPORTED THAT AFTER SURGERY THE SURGEON FOUND A SHADOW IN A X-RAY IMAGE. AFTER RECONFIRMING THE IMAGE, HE FOUND SOMETHING REMAINED. THEN THE SAME SURGERY WAS PERFORMED AGAIN USING OTHER IMPLANTS/DEVICES. AFTER PULLING OUT THE NAIL PREVIOUSLY IMPLANTED, A METALLIC PIECE WAS REMOVED FROM THE MEDULLARY SPACE. THEN REMOVING THE METAL MATERIALS AS MUCH AS POSSIBLE, THE SURGERY WAS COMPLETED BY IMPLANTING OTHER NAIL. BACK-UPS AVAILABLE. DELAY GREATER THAN 30 MIN.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9281441 ↗InjuryOct 30, 2019Break; Break
IT WAS REPORTED THAT AFTER SURGERY THE SURGEON FOUND A SHADOW IN A X-RAY IMAGE. AFTER RECONFIRMING THE IMAGE, HE FOUND SOMETHING REMAINED. THEN THE SAME SURGERY WAS PERFORMED AGAIN USING OTHER IMPLANTS/DEVICES. AFTER PULLING OUT THE NAIL PREVIOUSLY IMPLANTED, A METALLIC PIECE WAS REMOVED FROM THE MEDULLARY SPACE. THEN REMOVING THE METAL MATERIALS AS MUCH AS POSSIBLE, THE SURGERY WAS COMPLETED BY IMPLANTING OTHER NAIL. BACK-UPS AVAILABLE. DELAY GREATER THAN 30 MIN.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9255828 ↗InjuryFeb 11, 2019Break; Break
IT WAS REPORTED THAT PATIENT WHO HAD INTERTAN OPERATION ON (B)(6) 2017 AT THIS HOSPITAL HAD PAIN AND TOOK X-RAY AT OTHER HOSPITAL, THEN IT WAS NOTED THAT THE NAIL WAS BROKEN. REVISION SURGERY (BHR) WIL BE PERFORMED.
Patient problems: Pain
MAUDE report 8329830 ↗InjuryDec 19, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO FRACTURE. THE FRACTURE STARTED AT DISTAL SCREW SITE. PRIMARY SURGERY WAS PERFORMED ON (B)(6) 2018. (ABOUT 1 WEEK AGO) THE COMPLAINT IMPLANTS WERE NOT AVAILABLE.
Patient problems: Bone Fracture(s)
MAUDE report 8181554 ↗InjuryJan 4, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IF WAS REPORTED A PARTIAL BONE FRACTURE DURING SURGERY WHILE INSERTING THE LPS AT THE DISTAL FEMORAL
Patient problems: Bone Fracture(s)
MAUDE report 7166063 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111025, K161264, K210980. One submission often covers many models, so check the product and lot details in each record.
- Z-2456-2026 · Mar 16, 2026 · Open, ClassifiedTRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
- Z-1806-2025 · Mar 18, 2025 · Open, ClassifiedMETA-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
- Z-2444-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail
- Z-2443-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail
- Z-1132-2013 · Mar 6, 2013 · TerminatedTRIGEN (TM) L-P SCREW, 4.5 MM X 37.5 MM, TI-6AL-4V, QTY: (1), STERILE R, REF 71645437, Smith & Nephew, Inc. Memphis, TN 38116 USA orthopedic surgery
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page