Skip to content

TRIGEN 71645027Recalls & reported problems

Smith & Nephew, Inc. · TRIGEN LOW PROFILE SCREW 5.0MM X 27.5MM · Class 2 · Implant

All models of TRIGEN (381)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00885556215548), received between Jan 4, 2018 and Nov 26, 2025 — classified as 9 injuries.

The most frequently reported specific device problem is “Break” (3 reports).

5 other recalls cover devices cleared under the same FDA submission (K111025, K161264, K210980); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Fracture3
  • Adverse Event Without Identified Device or Use Problem2
  • Insufficient Device Problem Information1
  • Material Twisted/Bent1

Patient problems

  • Bone Fracture(s)3
  • Device Embedded In Tissue or Plaque3
  • Injury2
  • Pain1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 26, 2025Fracture; Material Twisted/Bent; Fracture; Material Twisted/Bent

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, WHEN INSERTING ONE (1) TRIGEN LOW PROFILE SCREW 5.0MM X 27.5MM, THE SCREW HEAD BECAME STRIPPED PREVENTING ROTATION. ATTEMPTS TO REMOVE THE SCREW USING PLIERS RESULTED IN FRACTURE OF THE SCREW HEAD, MAKING EXTRACTION DIFFICULT, THE FRACTURED HEAD WAS REMOVED. THE SURGEON DECIDED TO LEAVE THE SCREW INSIDE THE BODY. THE PROCEDURE WAS RESUMED, AFTER A 2 HOUR SURGICAL DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 23657646 ↗
  • InjuryJan 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING SURGERY, AFTER INSERTING THE TRIGEN LOW PROFILE SCREW 5.0MM X 27.5MM, AN INCOMPLETE BONE FRACTURE WAS IDENTIFIED. THE PROCEDURE WAS RESUMED WITHOUT DELAY, USING THE SAME DEVICE.

    Patient problems: Bone Fracture(s)

    MAUDE report 21174990 ↗
  • InjuryApr 8, 2020Fracture; Fracture

    IT WAS REPORTED THAT A FEMALE PATIENT HAD AN INTERTAN SURGERY. AFTER THE SURGERY SURGEON FOUND OUT THERE WAS FRACTURE AROUND DISTAL SCREWS, HE THINKS THE DIAMETER OF THE PART UNDER THE SCREW HEAD IS WIDE AND THE DIAMETER OF DRILL IS NOT FIT FOR IT.

    Patient problems: Injury

    MAUDE report 9938481 ↗
  • InjuryApr 8, 2020Fracture; Fracture

    IT WAS REPORTED THAT A FEMALE PATIENT HAD AN INTERTAN SURGERY. AFTER THE SURGERY SURGEON FOUND OUT THERE WAS FRACTURE AROUND DISTAL SCREWS, HE THINKS THE DIAMETER OF THE PART UNDER THE SCREW HEAD IS WIDE AND THE DIAMETER OF DRILL IS NOT FIT FOR IT.

    Patient problems: Injury

    MAUDE report 9938149 ↗
  • InjuryNov 5, 2019Break; Break

    IT WAS REPORTED THAT AFTER SURGERY THE SURGEON FOUND A SHADOW IN A X-RAY IMAGE. AFTER RECONFIRMING THE IMAGE, HE FOUND SOMETHING REMAINED. THEN THE SAME SURGERY WAS PERFORMED AGAIN USING OTHER IMPLANTS/DEVICES. AFTER PULLING OUT THE NAIL PREVIOUSLY IMPLANTED, A METALLIC PIECE WAS REMOVED FROM THE MEDULLARY SPACE. THEN REMOVING THE METAL MATERIALS AS MUCH AS POSSIBLE, THE SURGERY WAS COMPLETED BY IMPLANTING OTHER NAIL. BACK-UPS AVAILABLE. DELAY GREATER THAN 30 MIN.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 9281441 ↗
  • InjuryOct 30, 2019Break; Break

    IT WAS REPORTED THAT AFTER SURGERY THE SURGEON FOUND A SHADOW IN A X-RAY IMAGE. AFTER RECONFIRMING THE IMAGE, HE FOUND SOMETHING REMAINED. THEN THE SAME SURGERY WAS PERFORMED AGAIN USING OTHER IMPLANTS/DEVICES. AFTER PULLING OUT THE NAIL PREVIOUSLY IMPLANTED, A METALLIC PIECE WAS REMOVED FROM THE MEDULLARY SPACE. THEN REMOVING THE METAL MATERIALS AS MUCH AS POSSIBLE, THE SURGERY WAS COMPLETED BY IMPLANTING OTHER NAIL. BACK-UPS AVAILABLE. DELAY GREATER THAN 30 MIN.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 9255828 ↗
  • InjuryFeb 11, 2019Break; Break

    IT WAS REPORTED THAT PATIENT WHO HAD INTERTAN OPERATION ON (B)(6) 2017 AT THIS HOSPITAL HAD PAIN AND TOOK X-RAY AT OTHER HOSPITAL, THEN IT WAS NOTED THAT THE NAIL WAS BROKEN. REVISION SURGERY (BHR) WIL BE PERFORMED.

    Patient problems: Pain

    MAUDE report 8329830 ↗
  • InjuryDec 19, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO FRACTURE. THE FRACTURE STARTED AT DISTAL SCREW SITE. PRIMARY SURGERY WAS PERFORMED ON (B)(6) 2018. (ABOUT 1 WEEK AGO) THE COMPLAINT IMPLANTS WERE NOT AVAILABLE.

    Patient problems: Bone Fracture(s)

    MAUDE report 8181554 ↗
  • InjuryJan 4, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IF WAS REPORTED A PARTIAL BONE FRACTURE DURING SURGERY WHILE INSERTING THE LPS AT THE DISTAL FEMORAL

    Patient problems: Bone Fracture(s)

    MAUDE report 7166063 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K111025, K161264, K210980. One submission often covers many models, so check the product and lot details in each record.

  • Z-2456-2026 · Mar 16, 2026 · Open, Classified
    TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
  • Z-1806-2025 · Mar 18, 2025 · Open, Classified
    META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
  • Z-2444-2023 · Jul 24, 2023 · Open, Classified
    TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail
  • Z-2443-2023 · Jul 24, 2023 · Open, Classified
    TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail
  • Z-1132-2013 · Mar 6, 2013 · Terminated
    TRIGEN (TM) L-P SCREW, 4.5 MM X 37.5 MM, TI-6AL-4V, QTY: (1), STERILE R, REF 71645437, Smith & Nephew, Inc. Memphis, TN 38116 USA orthopedic surgery

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates