TRIGEN 71645035Recalls & reported problems
Smith & Nephew, Inc. · TRIGEN LOW PROFILE SCREW 5.0MM X 35MM · Class 2 · Implant
All models of TRIGEN (381)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00885556215692), received between Mar 17, 2019 and Nov 26, 2024 — classified as 8 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
5 other recalls cover devices cleared under the same FDA submission (K111025, K161264, K210980); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Adverse Event Without Identified Device or Use Problem2
- Crack1
- Difficult to Insert1
- Insufficient Device Problem Information1
- Loosening of Implant Not Related to Bone-Ingrowth1
- Migration or Expulsion of Device1
- Peeled/Delaminated1
Patient problems
- Failure of Implant3
- No Clinical Signs, Symptoms or Conditions3
- Bacterial Infection2
- Injury2
- Abrasion1
- Awareness During Anesthesia1
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 26, 2024Crack; Crack
IT WAS REPORTED THAT, DURING THR SURGERY, AS SURGEON WAS IMPLANTING THE FINAL DISTAL INTERLOCK TRIGEN LOW PROFILE SCREW 5.0MM X 35MM, RIGHT AS IT WAS FULLY DOWN, THE HEAD OF THE SCREW CRACKED. SURGEON DID NOT REMOVE THE SCREW AS IT WAS IN A SUFFICIENT POSITION TO LEAVE IN. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20784867 ↗MalfunctionJul 7, 2023Difficult to Insert; Peeled/Delaminated; Difficult to Insert; Peeled/Delaminated
IT WAS REPORTED THAT, DURING SURGERY AN INTERNAL FIXATION SURGERY, WHILE INSERTING ONE (1) TRIGEN LOW PROFILE SCREW 5.0MM X 35MM INTO THE SECOND DISTAL SCREW HOLE OF ONE (1) META-TAN 9MM X 30CM RIGHT, IT WAS FOUND THAT THE SCREW PASSED THE OUTSIDE BUT STOPPED AT INSIDE. REPEATED ATTEMPTS TO DRILL AGAIN DID NOT RESOLVE THE PROBLEM. IT WAS CONFIRMED THAT THERE WAS NO TECHNIQUE ERROR, AND THAT THE SCREW WAS NOT INSERTED. THE PROCEDURE WAS RESUMED, AFTER 10 MINUTES SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17283423 ↗InjuryJun 11, 2022Break; Break
IT WAS REPORTED THAT, DURING SURGERY, THE LEFT INTERTAN NAIL AND THE SINGLE SLIDING TRIGEN LOW PROFILE SCREW WERE PLACED. THE DISTAL LOCKING WAS PERFORMED IN THE HOLE AND THE TRIGEN LOW PROFILE SCREW STOPPED ADVANCING APPROXIMATELY 7MM TO REACH THE SECOND CORTEX. THE SCREW HEAD SEPARATED AND DID NOT ALLOW ADVANCEMENT. THE PHYSICIAN DECIDED TO RETRIEVE THE FRAGMENT WITH AN ADJUSTABLE PRESSURE FORCEPS BUT SCREW HEAD BROKE. THE SHAFT OF THE SCREW WAS LEFT INSIDE THE BONE PASSING THROUGH THE NAIL LOCKING HOLE. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE WITHOUT SIGNIFICANT DELAYS. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Failure of Implant; Foreign Body In Patient
MAUDE report 14670559 ↗InjuryDec 7, 2021Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT, AFTER AN INTERTAN SURGERY HAD BEEN PERFORMED ON (B)(6) 2021, THE PATIENT EXPERIENCED PAIN POST SURGERY. THE X-RAYS SHOWED BACKING OUT OF THE COMPRESSION SCREW USED WITH AN INTERTAN DEVICE WHICH WAS LOCKED WITH THE SET SCREW. ALSO, THE SURGICAL TEAM ARE STILL DECIDING ON APPROPRIATE TREATMENT FOR THE PATIENT. IT IS UNKNOWN HOW THE ADVERSE EVENT WAS SOLVED AND CURRENT HEALTH STATUS OF PATIENT.
Patient problems: Failure of Implant; Pain
MAUDE report 12953204 ↗InjurySep 14, 2021Break; Break
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, A TRIGEN LOW PROFILE SCREW 5.0MM X 35MM WAS HARD TO INSERT COMPLETELY, THE SURGEON APPLIED STRONG FORCE AND CAUSED THE SCREW HEAD BROKE OFF (IT WAS FIXED TO THE SCREW DRIVER SO IT DID NOT FELL INSIDE THE PATIENT). THREADED PART OF THE SCREW REMAINED INSIDE THE PATIENT. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12471991 ↗InjuryMay 13, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, DUE TO A FALL ON (B)(6) 2020, A PATIENT EXPERIENCED A RIGHT TIBIAL FRACTURE WHICH REQUIRED A TEMPORARY TREATMENT WITH AN EXTERNAL FIXATOR ON (B)(6) 2020. THIS EXTERNAL FIXATOR WAS REMOVED ON (B)(6) 2020 AND AFTER A RIGHT TIBIAL INTERNAL INFIXATION SURGERY WITH TRIGEN LOW PROFILE SCREW 5.0MM X 35MM PERFORMED THE SAME DAY, THE PATIENT EXPERIENCED SECRETION AND WOUND HEALING PROBLEMS WITH THE PRESENCE OF STAPHYLOCOCCUS AUREUS. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2021 TO TREAT THIS ADVERSE EVENT. THE TRIGEN LOW PROFILE SCREW 5.0MM X 35MM (X2), INT HEX CAP SCR 5.0MM X 35MM, TRIGEN LOW PROFILE SCREW 5.0MM X 40MM AND TRIGEN LOW PROFILE SCREW 5.0MM X 55MM WERE EXPLANTED DURING THIS REVISION SURGERY. IT IS UNKNOWN IF THE EXPLANTED DEVICE WAS SUBSTITUTED WITH A SMITH AND NEPHEW IMPLANT. THE PATIENT OUTCOME IS UNKNOWN.
Patient problems: Awareness During Anesthesia; Bacterial Infection
MAUDE report 11817078 ↗InjuryMay 13, 2021Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT, DUE TO A FALL ON (B)(6) 2020, A PATIENT EXPERIENCED A RIGHT TIBIAL FRACTURE WHICH REQUIRED A TEMPORARY TREATMENT WITH AN EXTERNAL FIXATOR ON (B)(6) 2020. THIS EXTERNAL FIXATOR WAS REMOVED ON (B)(6) 2020 AND AFTER A RIGHT TIBIAL INTERNAL INFIXATION SURGERY WITH TRIGEN LOW PROFILE SCREW 5.0MM X 35MM PERFORMED THE SAME DAY, THE PATIENT EXPERIENCED SECRETION AND WOUND HEALING PROBLEMS WITH THE PRESENCE OF STAPHYLOCOCCUS AUREUS. A REVISION SURGERY WAS PERFORMED ON 07-APR-2021 TO TREAT THIS ADVERSE EVENT. THE TRIGEN LOW PROFILE SCREW 5.0MM X 35MM (X2), INT HEX CAP SCR 5.0MM X 35MM, TRIGEN LOW PROFILE SCREW 5.0MM X 40MM AND TRIGEN LOW PROFILE SCREW 5.0MM X 55MM WERE EXPLANTED DURING THIS REVISION SURGERY. IT IS UNKNOWN IF THE EXPLANTED DEVICE WAS SUBSTITUTED WITH A SMITH AND NEPHEW IMPLANT. THE PATIENT OUTCOME IS UNKNOW.
Patient problems: Abrasion; Bacterial Infection
MAUDE report 11817058 ↗InjuryApr 19, 2021Break; Break
IT WAS REPORTED THAT, AFTER AN INTERNAL FIXATION SURGERY THAT HAD BEEN PERFORMED ON 12-FEB-2021, THE INTERTAN 10S 10MM X 18CM 125D NAIL FRACTURED ON 16-FEB-2021. A REVISION SURGERY WAS PERFORMED ON 24-MAR-2021 TO EXPLANT THE NAIL SYSTEM. PATIENT OUTCOME IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 11685228 ↗MalfunctionNov 13, 2019Break; Break
IT WAS REPORTED THAT THE ITEM BROKE AT THE POINT OF THE LAG/COMPRESSION SCREW COMBO. IT BROKE INSIDE THE PATIENT, AND ALL PIECES WERE RECOVERED. BACK-UP USED TO COMPLETE THE PROCEDURE. NO DELAY OR INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9318110 ↗InjuryMay 14, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT REVISION SURGERY WAS PERFORMED DUE TO UNSPECIFIED REASONS.
Patient problems: Injury
MAUDE report 8609098 ↗InjuryMar 17, 2019Break; Break
IT WAS REPORTED THAT AN UPCOMING REVISION SURGERY IS SCHEDULED FOR THE PATIENT DUE TO BREAKAGE OF THE NAIL. DETAILS OF REVISION ARE UNAVAILABLE. FOUR DEVICES WERE EXPLANTED, ACCORDING TO X-RAYS ATTACHED AND DEVICES RECEIVED.
Patient problems: Injury
MAUDE report 8427361 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111025, K161264, K210980. One submission often covers many models, so check the product and lot details in each record.
- Z-2456-2026 · Mar 16, 2026 · Open, ClassifiedTRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
- Z-1806-2025 · Mar 18, 2025 · Open, ClassifiedMETA-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
- Z-2443-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail
- Z-2444-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail
- Z-1132-2013 · Mar 6, 2013 · TerminatedTRIGEN (TM) L-P SCREW, 4.5 MM X 37.5 MM, TI-6AL-4V, QTY: (1), STERILE R, REF 71645437, Smith & Nephew, Inc. Memphis, TN 38116 USA orthopedic surgery
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page