Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556216491), received between Aug 16, 2023 and Dec 3, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

5 other recalls cover devices cleared under the same FDA submission (K111025, K161264, K210980); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • Nonunion/Delayed-union Bone Fracture2
  • Ambulation or Postural Difficulties1
  • Bone Fracture(s)1
  • Failure of Implant1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 3, 2025Fracture; Fracture

    IT WAS REPORTED THAT, AFTER FEMORAL NAILING FOLLOWING A HIGH ENERGY INJURY ON (B)(6) 2023, THE PATIENT HAD A BONE DEFECT. THEN, ON (B)(6) 2024 TWO DISTAL SCREWS WERE REMOVED TO ALLOW DYNAMIZATION. THE PATIENT RECENTLY EXPERIENCE PAIN AND AN X-RAY SHOWED A BROKEN DISTAL SCREW. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY ON (B)(6) 2025, IN WHICH THE NAIL HAS BEEN REPLACE AND BONE GRAFT HAS BEEN PLACED TO ADDRESS THE BONE DEFECT. CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Nonunion/Delayed-union Bone Fracture; Pain

    MAUDE report 23704741 ↗
  • InjuryAug 16, 2023Fracture; Fracture

    IT WAS REPORTED THAT, AFTER UNDERGOING RIGHT FEMORAL INTRAMEDULLARY FRACTURE FIXATION ON (B)(6) 2021 AS A RESULT OF A BICYCLE ACCIDENT, THE PATIENT EXPERIENCED MIDDLE DISTAL INTERLOCKING SCREW BREAKAGE AND PROXIMAL DISTAL INTERLOCKING SCREW LOOSENING THAT CAUSED A 2 MM SHORTENING OF DISTAL SEGMENT, IN CONJUNCTION WITH RIGHT HYPERTROPHIC FEMORAL SHAFT NONUNION FRACTURE. THE PATIENT EXPERIENCED PAIN AND AMBULATION DIFFICULTIES THAT REQUIRED THE USE OF A WALKER. THESE COMPLICATIONS WERE TREATED BY PERFORMING A REVISION SURGERY ON (B)(6) 2022, IN WHICH THE INTERNAL FIXATION HARDWARE WAS REMOVED AND REPLACED WITH A COMPETITOR¿S INTRAMEDULLARY NAIL-SCREW SYSTEM (DEPUY SYNTHES). PATIENT¿S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Ambulation or Postural Difficulties; Bone Fracture(s); Failure of Implant; Nonunion/Delayed-union Bone Fracture

    MAUDE report 17557390 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K111025, K161264, K210980. One submission often covers many models, so check the product and lot details in each record.

  • Z-2456-2026 · Mar 16, 2026 · Open, Classified
    TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
  • Z-1806-2025 · Mar 18, 2025 · Open, Classified
    META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
  • Z-2443-2023 · Jul 24, 2023 · Open, Classified
    TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail
  • Z-2444-2023 · Jul 24, 2023 · Open, Classified
    TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail
  • Z-1132-2013 · Mar 6, 2013 · Terminated
    TRIGEN (TM) L-P SCREW, 4.5 MM X 37.5 MM, TI-6AL-4V, QTY: (1), STERILE R, REF 71645437, Smith & Nephew, Inc. Memphis, TN 38116 USA orthopedic surgery

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →