Skip to content

JOURNEY 74014461Recalls & reported problems

Smith & Nephew, Inc. · JOURNEY II EXTRAMEDULLARY ALIGNMENT TUBE · Class 1

All models of JOURNEY (566)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00885556237892), received between Jan 21, 2020 and Aug 6, 2021 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break2
  • Connection Problem1
  • Incomplete or Inadequate Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2021Break; Break

    IT WAS REPORTED THAT, THE LOCKING MECHANISM OF A JRNY II ALIGNMENT TUBE IS BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12291547 ↗
  • MalfunctionJul 15, 2021Break; Break

    IT WAS REPORTED THAT, DURING SET-UP OF TOTAL KNEE ARTHROPLASTY SURGERY, A JOURNEY II ALIGNMENT TUBE WAS NOTICED TO BE BROKEN. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A SMITH & NEPHEW BACK-UP DEVICE INSTEAD. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12171081 ↗
  • MalfunctionMay 20, 2021Incomplete or Inadequate Connection; Incomplete or Inadequate Connection

    IT WAS REPORTED THAT, DURING A SET UP FOR A TKA SURGERY, THE JII XR TIB NON-SPIKED FIX ROD CASE (B)(4) WAS NOT ABLE TO FULLY GO THROUGH THE JRNY II ALIGNMENT TUBE CASE (B)(4) AND WOULD GET STUCK. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. NO PATIENT INJURIES AND NO OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11858146 ↗
  • MalfunctionJan 21, 2020Connection Problem; Connection Problem

    IT WAS REPORTED THAT DURING SURGERY, THE TWO INSTRUMENTS WON'T COME APART FOR CLEANING. IT CAUSED A DELAY OF 30 MINUTES TO AN HOUR. THERE WAS AN S&N BACKUP DEVICE AVAILABLE. NO PATIENT INJURIES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9610647 ↗

Other Guide, surgical, instrument devices

All Guide, surgical, instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page