Skip to content

TRIGEN 71675386Recalls & reported problems

Smith & Nephew, Inc. · TRIGEN INTERTAN 10S 10MM X 20CM 130DEGREE · Class 2 · Implant

All models of TRIGEN (380)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00885556343692), received between Mar 20, 2018 and Nov 11, 2024 — classified as 7 injuries.

The most frequently reported specific device problem is “Fracture” (3 reports).

6 other recalls cover devices cleared under the same FDA submission (K040212, K241804); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Break2
  • Device Dislodged or Dislocated1
  • Migration1
  • Unintended Movement1

Patient problems

  • Failure of Implant5
  • Implant Pain1
  • Injury1
  • No Information1
  • Nonunion/Delayed-union Bone Fracture1
  • Pain1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 11, 2024Fracture; Fracture

    IT WAS REPORTED THAT AFTER AN INTERNAL FIXATION WAS PERFORMED ON (B)(6) 2024, PATIENT EXPERIENCED BREAKAGE OF THE NAIL. THIS ADVERSE EVENT WAS ADDRESSED BY PERFORMING A REVISION SURGERY ON AN UNKNOWN DATE, IN WHICH, THE DEVICE WAS EXCHANGED. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture; Pain

    MAUDE report 20649497 ↗
  • InjuryOct 11, 2024Fracture; Fracture

    IT WAS REPORTED THAT, AFTER RIGHT FEMUR OSTEOSYNTHESIS SURGERY HAD BEEN PERFORMED ON (B)(6) 2024 (DUE PER TROCHANTERIC AND SUBTROCHANTERIC FRACTURES OF THE RIGHT HIP), DURING REHABILITATION THE PATIENT EXPERIENCED AN EPISODE OF PAIN THAT OCCURRED DURING PHYSICAL THERAPY, SINCE THEN SOMETHING HAD BEEN WRONG WITH HIS LEG; HOWEVER, THE PATIENT HAD CONTINUED TO WALK WITH THE ROLLATOR. ON (B)(6) 2024 AN X-RAY CHECK REVEALED A FRACTURE OF THE INTERTAN NAIL AT THE LEVEL THE DISTAL LOCKING SCREW. THIS ADVERSE EVENT WAS REPORTED ON 20-SEP-2024 AND WAS SOLVED BY REVISION SURGERY ON (B)(6) 2024, IN WHICH THE INTERTAN 10S 10MM X 20CM 130D WAS EXPLANTED, AND A LONG NAIL WAS IMPLANTED. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 20434297 ↗
  • InjuryOct 20, 2021Break; Break

    IT WAS REPORTED THAT, AFTER AN INTERNAL FIXATION SURGERY (UNKNOWN DATE), WHERE A TRIGEN INTERTAN 10S 10MM X 20CM 130DEGREE, A INTERTAN SUBTROCHANTERIC LAG SCREW 11MM X 90MM AND A TRIGEN INTERNAL HEX CAPTURED SCREW 5.0MM X 35MM WERE IMPLANTED, THE TRIGEN INTERTAN 10S 10MM X 20CM 130DEGREE BROKEN INSIDE THE PATIENT JUST ABOVE THE DISTAL SCREW HOLE. IT IS UNKNOWN HOW THIS ADVERSE EVENT WAS TREATED AS WELL AS THE CURRENT HEALTH STATE OF THE PATIENT.

    Patient problems: Failure of Implant

    MAUDE report 12665443 ↗
  • InjuryAug 31, 2021Device Dislodged or Dislocated; Migration; Device Dislodged or Dislocated; Migration

    IT WAS REPORTED THAT, AFTER AN INTERTROCHANTERIC FRACTURE CORRECTION SURGERY WAS PERFORMED ON (B)(6) 2021, THE LAG/COMP SCREW KIT 90/85, THAT WAS ORIGINALLY INSERTED INTO A INTERTAN 10S 10MM X 20CM 130D, FAILED DUE TO A CUT OUT AND THE SCREWS BACKING OUT DESPITE BEING LOCKED IN THE CASE. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2021 TO TREAT THIS ADVERSE EVENT. THE PATIENT'S OUTCOME IS UNKNOWN.

    Patient problems: Failure of Implant

    MAUDE report 12400003 ↗
  • InjuryAug 12, 2020Break; Break

    IT WAS REPORTED THAT, ABOUT ONE MONTH AFTER A TRAUMA SURGERY IN WHICH AN INTERTAN 10S 10MM X 20CM 130D NAIL HAD BEEN IMPLANTED, IT WAS DISCOVERED THAT THE NAIL HAD BROKEN AT THE DISTAL LOCKING HOLE. NO FALL HAS BEEN ASSOCIATED TO THE INJURY, THE PATIENT UNDERWENT REVISION SURGERY TO CORRECT THE ISSUE. IT IS UNKNOWN THE CURRENT HEALTH STATUS OF THE PATIENT.

    Patient problems: Failure of Implant

    MAUDE report 10399893 ↗
  • InjuryMar 31, 2020Fracture; Fracture

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED TO CORRECT AN INTERTAN NAIL BREAKAGE DUE TO STRESS AND FATIGUE.

    Patient problems: Injury

    MAUDE report 9902829 ↗
  • InjuryMar 20, 2018Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT PATIENT WENT FOR A MANDATORY CLINIC CHECK UP AND SURGEON REQUEST A X-RAY FOR THE PATIENT TO SEE HOW THE FRACTURE IS HEALING . NOTED FROM THE X-RAY THAT THE COMPRESSION SCREW IS BACKING OUT WHEREAS THE TOP SCREW IS STILL IN INTACT. ACCORDING TO SURGEON, PATIENT IS WALKING IN AWKWARD POSTURE . NO PAIN SO FAR. COMPRESSION SCREW BACK OUT OF 1 CM.

    Patient problems: No Information

    MAUDE report 7355630 ↗

FDA clearance and classification

Cleared through 510(k) K241804 on Aug 22, 2024 (decision: Substantially Equivalent). Submitted by Smith & Nephew, Inc. as “TRIGEN INTERTAN 10S Nail System”. The FDA review took 62 days. FDA record ↗

Cleared through Special 510(k) K040212 on Feb 20, 2004 (decision: Substantially Equivalent). Submitted by Smith & Nephew, Inc. as “Trigen Intertan”. The FDA review took 21 days. FDA record ↗

The FDA regulates product code JDS (nail, fixation, bone) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K040212, K241804. One submission often covers many models, so check the product and lot details in each record.

  • Z-1133-2022 · Apr 22, 2022 · Open, Classified
    INTERTAN 1.5 NAIL 13mm X 42cm 130DEG. Left INTERTAN 1.5 NAIL 13mm X 42cm 130DEG. Right
  • Z-0772-2021 · Nov 2, 2020 · Terminated
    Bone fixation nail. TRIGEN INTERTAN - Product Usage: indicated for fractures of the femur including simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long
  • Z-0097-2018 · Oct 24, 2017 · Open, Classified
    smith&nephew TRIGEN INTERTAN 10S, REF 71675384, 10 MM X 18 CM NAIL, 130o (degrees), SET SCREW PREASSEMBLED IN NAIL
  • Z-2258-2014 · Jul 2, 2014 · Terminated
    TRIGEN(R) INTERTAN(TM) 10MM X 18 CM NAIL, 125o, REF 71675201, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 18 CM NAIL, 125o, REF
  • Z-1637-2014 · Apr 4, 2014 · Terminated
    TRIGEN INTERTAN 1.5 NAIL: (1) REF 71676515, 10 MM X 10 CM, 125 degree, LEFT, (2) REF 71676556, 11.5 MM X 36 CM, 125 degree, RIGHT, (3) REF 71676558, 11.5 MM X 38 CM, 125 degree
  • Z-1636-2014 · Apr 4, 2014 · Terminated
    TRIGEN INTERTAN NAIL: (1) REF 71675201, 10 MM X 18 CM, 125 degree, (2) REF 71675202, 11.5 MM X 18 CM, 125 degree, (3) REF 71675203, 13 MM X 18 CM, 125 degree, (4) REF 71675204,

Common questions about the TRIGEN 71675386

Has the TRIGEN 71675386 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00885556343692), based on the FDA recall file dated Nov 27, 2026. However, 6 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the TRIGEN 71675386?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between Mar 20, 2018 and Nov 11, 2024. In those reports, the most common specific device problems are “Fracture” (3), “Break” (2) and “Device Dislodged or Dislocated” (1); the most common patient problems are “Failure of Implant” (5), “Implant Pain” (1) and “Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the TRIGEN 71675386?

None of the reports are classified as deaths, 7 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the TRIGEN 71675386 FDA approved?

It was cleared through Special 510(k) K040212 on Feb 20, 2004; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the TRIGEN 71675386 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the TRIGEN 71675386, and is it still sold?

The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand TRIGEN, model 71675386). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other NAIL, FIXATION, BONE devices

All NAIL, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page