Skip to content

TRIGEN 71675511Recalls & reported problems

Smith & Nephew, Inc. · TRIGEN INTERTAN 10S 10MM X 36CM 125DEGREE LEFT · Class 2 · Implant

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885556344248), received between Oct 27, 2017 and Dec 13, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Failure to Align” (1 report).

6 other recalls cover devices cleared under the same FDA submission (K040212, K241804); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Failure to Align1

Patient problems

  • Nonunion/Delayed-union Bone Fracture2
  • Device Embedded In Tissue or Plaque1
  • Failure of Implant1
  • Injury1
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, FOLLOWING A LEFT IMN FIXATION SURGERY PERFORMED ON (B)(6) 2018 DUE TO A DISPLACED INTERTROCHANTERIC FRACTURE OF THE FEMUR, THE PATIENT REQUIRED CONVERSION TO A LEFT HIP THR DUE TO FRACTURE NONUNION ON (B)(6) 2018. THIS INFORMATION WAS COLLECTED RETROSPECTIVELY AS PART OF A POST-MARKET CLINICAL FOLLOW UP (PMCF) ACTIVITY AND IS PROVIDED IN AN ANONYMIZED FORMAT; THEREFORE, NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 20924858 ↗
  • InjuryNov 5, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER AN INTERNAL FIXATION PROCEDURE HAD BEEN PERFORMED ON THE PATIENT'S LEFT FEMUR ON (B)(6) 2016 TO TREAT A FEMORAL SHAFT AND INTERTROCHANTERIC FRACTURE, THE PATIENT EXPERIENCED PAINFUL NONUNION OF THE FEMORAL SHAFT FRACTURE AND BREAKAGE OF THE TWO (2) DISTAL INTERLOCKING SCREWS 46.7 WEEKS POSTOPERATIVELY (AS CONFIRMED PER X-RAY ASSESSMENT). A REVISION SURGERY WAS PERFORMED ON 23-AUG-2017 TO TREAT THIS ADVERSE EVENT. ADDITIONAL DETAILS REGARDING THE CLINICAL COURSE OF THE PATIENT WERE NOT PROVIDED. THIS INCIDENT WAS NOTICED IN A RETROSPECTIVE POST-MARKET CLINICAL FOLLOW UP ACTIVITY (PMCF)], ADDITIONAL INFORMATION IS NOT KNOWN, AND IT IS NOT POSSIBLE TO COLLECT IT, WE DO NOT HAVE THE OPPORTUNITY TO REVISIT NEGATIVE FEEDBACK TO RETRIEVE FURTHER INFORMATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: Device Embedded In Tissue or Plaque; Failure of Implant; Nonunion/Delayed-union Bone Fracture; Pain

    MAUDE report 15741080 ↗
  • InjuryOct 27, 2017Failure to Align; Failure to Align

    IT WAS REPORTED THAT WHEN INSERTING THE ANTI ROTATION BAR INTO THE FEMORAL HEAD THE OPERATORS NOTICED IT CONVERGED WITH THE 3.2MM X 343MM GUIDE PIN. THEY TRIED A NEW GUIDE PIN AS THEY DISCOVERED IT WAS BENT, ONLY TO HAVE THE SAME RESULT. WHEN THEY REMOVED THE NEW ONE IT WAS BENT TOO. THEY REMOVED THE WHOLE NAIL AND NOTICED THE SET SCREW HAD DROPPED POSSIBLY BENDING THE GUIDE PIN ON ENTRY. THEY WERE NOT SURE WHY THE SET SCREW HAD DROPPED BUT THERE WAS SOME FEELING THE 4.7MM HEX HAD BEEN INCORRECTLY USED TO TRY TO ATTACH THE NAIL TO THE JIG. SURGICAL DELAY OCCURED DUE TO THIS BUT TO NO ADVERSE OUTCOME TO THE PATIENT.

    Patient problems: Injury

    MAUDE report 6984408 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K040212, K241804. One submission often covers many models, so check the product and lot details in each record.

  • Z-1133-2022 · Apr 22, 2022 · Open, Classified
    INTERTAN 1.5 NAIL 13mm X 42cm 130DEG. Left INTERTAN 1.5 NAIL 13mm X 42cm 130DEG. Right
  • Z-0772-2021 · Nov 2, 2020 · Terminated
    Bone fixation nail. TRIGEN INTERTAN - Product Usage: indicated for fractures of the femur including simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long
  • Z-0097-2018 · Oct 24, 2017 · Open, Classified
    smith&nephew TRIGEN INTERTAN 10S, REF 71675384, 10 MM X 18 CM NAIL, 130o (degrees), SET SCREW PREASSEMBLED IN NAIL
  • Z-2258-2014 · Jul 2, 2014 · Terminated
    TRIGEN(R) INTERTAN(TM) 10MM X 18 CM NAIL, 125o, REF 71675201, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 18 CM NAIL, 125o, REF 71
  • Z-1637-2014 · Apr 4, 2014 · Terminated
    TRIGEN INTERTAN 1.5 NAIL: (1) REF 71676515, 10 MM X 10 CM, 125 degree, LEFT, (2) REF 71676556, 11.5 MM X 36 CM, 125 degree, RIGHT, (3) REF 71676558, 11.5 MM X 38 CM, 125 degree
  • Z-1636-2014 · Apr 4, 2014 · Terminated
    TRIGEN INTERTAN NAIL: (1) REF 71675201, 10 MM X 18 CM, 125 degree, (2) REF 71675202, 11.5 MM X 18 CM, 125 degree, (3) REF 71675203, 13 MM X 18 CM, 125 degree, (4) REF 71675204,

Other NAIL, FIXATION, BONE devices

All NAIL, FIXATION, BONE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.