EVOS 71174928Recalls & reported problems
Smith & Nephew, Inc. · DEPTH GAUGE FOR 2.4MM AND 2.7MM SCREWS · Class 1
All models of EVOS (191)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00885556387696), received between Sep 17, 2019 and Sep 27, 2022 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break8
- Material Twisted/Bent2
- Fracture1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 27, 2022Fracture; Fracture
IT WAS REPORTED THAT, DURING A TRAUMA SURGERY THE HOOK OF A DEPTH GAUGE FOR 2.4MM AND 2.7MM SCREWS BROKE OFF. IT IS UNKNOWN HOW THE PROCEDURE WAS FINISHED AND IF THIS INCIDENT CAUSED A DELAY. NO HARM TO THE PATIENT OR FURTHER COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15491726 ↗MalfunctionMay 11, 2022Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED THAT THE TIP OF A SHORT DEPTH GAUGE 2.4 MM/2.7 MM SCREWS IS BROKEN AND BENT. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14352609 ↗MalfunctionFeb 27, 2022Break; Break
IT WAS REPORTED THAT THE DEPTH MEASURE OF AN EVOS DEPTH GAUGE FOR 2.4MM AND 2.7MM SCREWS BROKE. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13615275 ↗MalfunctionDec 21, 2021Break; Material Twisted/Bent; Break; Material Twisted/Bent
IT WAS REPORTED THAT, THE TIP OF THE DEPTH GAUGE FOR 2.4MM AND 2.7MM SCREWS WAS FOUND DURING INSPECTION BENT AND BROKEN, RENDERING IT UNUSABLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13056611 ↗MalfunctionAug 12, 2021Break; Break
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, A DEPTH GAUGE FOR 2.4MM AND 2.7MM SCREWS BROKE DURING PLATE PROCEDURE. INCIDENT OCCURRED WHILE INSTRUMENT WAS OUTSIDE THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12316036 ↗MalfunctionMar 29, 2021Break; Break
IT WAS REPORTED THAT, THE WIRE PORTION OF DEPTH GAUGE FOR 2.4MM AND 2.7MM SCREWS BROKE. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11586433 ↗MalfunctionMar 8, 2021Break; Material Twisted/Bent; Break; Material Twisted/Bent
IT WAS REPORTED THAT, THE MEASURING END OF A DEPTH GAUGE FOR 2.4MM AND 2.7MM SCREWS WAS BENT SEVERELY AND WHEN BENT BACK BROKE IN HALF. THIS HAPPENED AFTER PROCEDURE. NO FURTHER COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11431019 ↗MalfunctionFeb 28, 2021Break; Break
IT WAS REPORTED THAT THE DEPTH GAUGE BROKE. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY AND IF THERE WAS PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11391066 ↗MalfunctionSep 17, 2019Break; Break
IT WAS REPORTED THAT THE TIP OF THE DEPTH GAUGE BROKE OFF AND HAD TO BE EXTRACTED OFF THE PATIENT. NO DELAY REPORTED. NO PATIENT INJURIES REPORTED. AN S&N BACKUP WAS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9079870 ↗
FDA clearance and classification
The FDA regulates product code HTJ (gauge, depth) under 21 CFR 888.4300 as a Class I device. 21 CFR 888.4300 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the EVOS 71174928
Has the EVOS 71174928 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00885556387696), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the EVOS 71174928?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Sep 17, 2019 and Sep 27, 2022. In those reports, the most common specific device problems are “Break” (8), “Material Twisted/Bent” (2) and “Fracture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EVOS 71174928?
None of the reports are classified as deaths, 0 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the EVOS 71174928 FDA approved?
Devices of this type (product code HTJ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Smith & Nephew, Inc.
Who makes the EVOS 71174928, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand EVOS, model 71174928). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other GAUGE, DEPTH devices
- NA 319.006Synthes GmbH · GAUGE, DEPTH892 reports
- NA 03.133.080Synthes GmbH · GAUGE, DEPTH169 reports
- NA 03.010.072Synthes GmbH · GAUGE, DEPTH129 reports
- MULTILOC 03.019.017Synthes GmbH · GAUGE, DEPTH116 reports
- NA 319.01Synthes GmbH · GAUGE, DEPTH111 reports
- NA 319.004Synthes GmbH · GAUGE, DEPTH82 reports
- QUICKSET 2274-60-000Depuy (Ireland) · GAUGE, DEPTH68 reports
- NA 319.091Synthes GmbH · GAUGE, DEPTH64 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page