TRIGEN 71631451Recalls & reported problems
Smith & Nephew, Inc. · META-TAN DRILL GUIDE · Class 2
All models of TRIGEN (381)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00885556399743), received between Nov 4, 2020 and Jan 30, 2026 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K122170); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Device Markings/Labelling Problem1
- Difficult or Delayed Separation1
- Mechanical Problem1
- Mechanics Altered1
- Off-Label Use1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Patient Involvement1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 30, 2026Mechanics Altered
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION (FEMUR FRACTURE IM NAIL) PROCEDURE, THE META-TAN DRILL GUIDE CONNECTION TO THE META TAN DROP WAS LOOSE AND WAS NOT A SNUG FIT. THIS MADE THE DROP UNSTABLE. THERE WAS A SURGICAL DELAY RETRYING TO TARGET THE JIG. THE PROCEDURE WAS RESUMED, AFTER A SIGNIFICANT DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24221004 ↗MalfunctionJun 30, 2025Off-Label Use; Difficult or Delayed Separation; Off-Label Use; Difficult or Delayed Separation
IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY, THE NAIL ENGAGED WITH ONE (1) META-TAN DRILL GUIDE OFF, SO SURGEON HAD TO BURR OFF ONE (1) PERCUTANEOUS GUIDE BOLT. THE PROCEDURE WAS RESUMED, AFTER A SIGNIFICANT DELAY, WITH THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22356710 ↗InjuryFeb 27, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT DURING SURGERY THE LAG SCREW MEASUREMENT WAS INCORRECT, INCORRECT SIZE LAG SCREW WAS SUBSEQUENTLY OPENED AND INSERTED. IT WAS THEN REMOVED AND A MORE APPROPRIATE SIZED LAG SCREW WAS INSERTED. THE PROCEDURE WAS COMPLETED, WITHOUT DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13615152 ↗InjuryAug 3, 2021Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE COMPRESSION SCREW COULD NOT GO THROUGH THE NAIL. IT IS UNKNOWN WHETHER THE META-TAN DRILL GUIDE, THE META-TAN DROP, THE META-TAN COMP SCREW DRILL OR THE META-TAN COMP SCREW STARTER DRILL CAUSED THIS PROBLEM. A CHANGE IN SURGICAL TECHNIQUE WAS REQUIRED TO COMPLETE THE PROCEDURE WITHOUT ANY DELAY. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12271541 ↗MalfunctionMay 26, 2021Break; Break
IT WAS REPORTED THAT, DURING SURGERY, A META-TAN DRILL GUIDE, A PERCUTANEOUS GUIDE BOLT, AND A CANNULATED IMPACTOR - LONG BROKE. DEVICE OUTSIDE THE PATIENT. THE PROCEDURE WAS ABLE TO BE COMPLETED WITH THE SAME DEVICE. IT IS UNKNOWN IF SURGERY WAS DELAYED. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11885782 ↗MalfunctionNov 4, 2020Break; Break
IT WAS REPORTED THAT THE META-TAN DRILL GUIDE HAS BECOME BROKEN. THIS WAS DISCOVERED AFTER AN OPERATION. THERE WAS NO CASE OR PATIENT AFFECTED.
Patient problems: No Patient Involvement
MAUDE report 10786299 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K122170. One submission often covers many models, so check the product and lot details in each record.
- Z-0351-2020 · Oct 7, 2019 · TerminatedLAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted sha
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page