TRIGEN 71631452Recalls & reported problems

Smith & Nephew, Inc. · META-TAN DROP · Class 2

All models of TRIGEN (381)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00885556399798), received between Aug 3, 2021 and Jun 1, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Failure to Align” (1 report).

1 other recall cover devices cleared under the same FDA submission (K122170); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Align1
  • Flaked1
  • Material Twisted/Bent1
  • Mechanical Problem1
  • Naturally Worn1
  • Positioning Failure1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 1, 2026Use of Device Problem; Failure to Align

    IT WAS REPORTED THAT, DURING A META-TAN SURGERY, THE SURGICAL TEAM REPORTED THAT THE SYSTEM OF ONE (1) META-TAN DROP, ONE (1) META-TAN LAG SCREW DRILL SLEEVE AND ONE (1) META-TAN GUIDE PIN SLEEVE WAS MISALIGN, AS THEY INSERTED A GUIDE WIRE (GUIDE PIN 3.2MM X 343MM) AND WHEN THEY SCREWED THE CANNULATED LAG SCREW OVER THE WIRE, THE WIRE SNAPPED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25354010 ↗
  • MalfunctionJan 30, 2026Positioning Failure

    IT WAS REPORTED THAT DURING AN INTERNAL FIXATION (FEMUR FRACTURE IM NAIL) PROCEDURE, THE META-TAN DROP WOULD NOT ENGAGE THE META-TAN LAG SCREW DRILL SLEEVE PROPERLY, AND THEREFORE THE METATAN DRAG SCREW DRILL SLEEVE POTENTIALLY HAD AN UNSPECIFIED ISSUE. THE PROCEDURE WAS RESUMED, AFTER A SIGNIFICANT DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED. PATIENT CONDITION IS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24221374 ↗
  • MalfunctionJan 4, 2022Flaked; Material Twisted/Bent; Naturally Worn; Flaked; Material Twisted/Bent; Naturally Worn

    IT WAS REPORTED THAT A META-TAN DROP IS WORN, FADED, CHIPPED AND BENT. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13152372 ↗
  • InjuryAug 3, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE COMPRESSION SCREW COULD NOT GO THROUGH THE NAIL. IT IS UNKNOWN WHETHER THE META-TAN DRILL GUIDE, THE META-TAN DROP, THE META-TAN COMP SCREW DRILL OR THE META-TAN COMP SCREW STARTER DRILL CAUSED THIS PROBLEM. A CHANGE IN SURGICAL TECHNIQUE WAS REQUIRED TO COMPLETE THE PROCEDURE WITHOUT ANY DELAY. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12271576 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K122170. One submission often covers many models, so check the product and lot details in each record.

  • Z-0351-2020 · Oct 7, 2019 · Terminated
    LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted sha

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices