STRIDE PFSR100931Recalls & reported problems
Smith & Nephew, Inc. · Stride Femoral Component Insertion/Removal Impactor Head · Class 2
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00885556627334), received between Nov 6, 2020 and Dec 22, 2021 — classified as 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break3
- Crack3
- Fracture2
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Patient Involvement2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryDec 22, 2021Break; Break
IT WAS REPORTED THAT DURING A NAVIO ASSISTED UKR SURGERY, THE STRIDE FEMORAL INS/REM IMPACTOR HEAD WAS BEING USED INSIDE THE PATIENT WHEN A PIECE OF THE PLASTIC CRACKED. THE PROCEDURE WAS COMPLETED, WITHOUT DELAY, USING THE SAME DEVICE. AFTER THE SURGERY, THE STRIDE FEMORAL INS/REM IMPACTOR HEAD WAS FULLY BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13060371 ↗InjuryAug 19, 2021Fracture; Fracture
IT WAS REPORTED THAT, DURING A NAVIO ASSISTED UKR SURGERY, THE STRIDE FEMORAL INS/REM IMPACTOR HEAD SPLIT IN HALF WHILE IMPACTING. NO DELAYS WERE REPORTED. THE PROCEDURE WAS FINISHED WITH A SMITH AND NEPHEW BACK UP DEVICE WITHOUT DELAYS. THE PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12342989 ↗InjuryMay 24, 2021Crack; Crack
IT WAS REPORTED THAT, UPON IMPACTING THE FEMUR IN A NAVIO-ASSISTED UKR SURGERY, THE STRIDE FEMORAL INS/REM IMPACTOR HEAD CRACKED IN HALF INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE (WRAPPED WITH COBAN) WITHOUT SIGNIFICANT DELAYS. THE PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11872967 ↗MalfunctionApr 19, 2021Fracture; Fracture
IT WAS REPORTED THAT, BEFORE A NAVIO UKR PROCEDURE, IT WAS FOUND THAT A STRIDE FEMORAL INSERTION/REMOVAL IMPACTOR HEAD WAS BROKEN. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. THERE WAS NO PATIENT AND NO DEBRIS INVOLVED AT THE MOMENT OF THE ISSUE. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11683778 ↗MalfunctionDec 15, 2020Crack; Crack
IT WAS REPORTED THAT DURING AN INSTRUMENT TRAY REPROCESSING, IT WAS FOUND THAT THE STRIDE FEMORAL IMPACTOR WAS CRACKED. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11008941 ↗MalfunctionDec 15, 2020Crack; Crack
IT WAS REPORTED THAT DURING AN INSTRUMENT TRAY REPROCESSING, IT WAS FOUND THAT THE STRIDE FEMORAL IMPACTOR WAS CRACKED. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11008898 ↗MalfunctionNov 6, 2020Break; Break
IT WAS REPORTED THAT THESE STRIDE FEMORAL INS/REM IMPACTOR HEAD ARE BROKEN. IT IS UNKNOWN WHERE THIS EVENT OCCURS. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Patient Involvement
MAUDE report 10804791 ↗MalfunctionNov 6, 2020Break; Break
IT WAS REPORTED THAT THESE STRIDE FEMORAL INS/REM IMPACTOR HEAD ARE BROKEN. IT IS UNKNOWN WHERE THIS EVENT OCCURS. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Patient Involvement
MAUDE report 10804740 ↗
Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- LinkSymphoKnee 880-307/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls2 reports
- Scorpio 73-3710Howmedica Osteonics Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls1 report
- Scorpio 73-3110Howmedica Osteonics Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls0 reports
- Arthrex® AR-503-TTTEArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall14 reports
- Arthrex® AR-503-TTTJArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall9 reports
- Arthrex® AR-503-TTTGArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- ENGAGE 1-10011-300Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- ENGAGE 1-10003-400Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall6 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.