Skip to content

Navio PFSR120031Recalls & reported problems

Smith & Nephew, Inc. · Navio BBT Footpedal · Class 2

All models of Navio (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556628485), received between Apr 23, 2020 and Dec 14, 2023 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Loose or Intermittent Connection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loose or Intermittent Connection1
  • Self-Activation or Keying1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 14, 2023Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT, DURING A CORI ASSISTED UKA SURGERY, WHILE BURRING THE FEMUR, THE DRILL WOULD INTERMITTENTLY STOP. THE TEAM TRIED USING A NEW ANSPACH DRILL BUT THE SAME THING HAPPENED. THEY ALSO TRIED USING A DIFFERENT NAVIO HANDPIECE BUT THE SAME ISSUE CONTINUED TO OCCUR. THE SURGERY WAS COMPLETED, AFTER A NON-SIGNIFICANT DELAY, SWITCHING TO MANUAL INSTRUMENTATION. AFTER THE CASE THEY NOTICED THAT THE OR BED WAS LOCKED ON TOP OF THE NAVIO BBT FOOTPEDAL CORD. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18326689 ↗
  • InjuryApr 23, 2020Self-Activation or Keying; Self-Activation or Keying

    IT WAS REPORTED THAT BLUE BELT FOOT PEDAL REMAINS PERMANENTLY IN THE ACTIVE POSITION. SURGERY WAS COMPLETED WITH LEGION STANDARD INSTRUMENTATION. IT IS UNKNOWN IF THERE WAS A DELAY BUT NO PATIENT IMPACT WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9990045 ↗

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page