JOURNEY 74016408Recalls & reported problems
Smith & Nephew, Inc. · JOURNEY II UK RESECTION ANTERIOR/POSTERIOR BLOCK LEFT MEDIAL/RIGHT LATERAL SZ 8 · Class 2
All models of JOURNEY (566)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556689998), received between Mar 23, 2021 and Jul 20, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Device Reprocessing Problem” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K190085); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Reprocessing Problem1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 20, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT, DURING A TKA SURGERY, A DRILL COULD NOT DRILLED THROUGH A JRNY II UNI RESECT FINISH TRL LM/RL SZ 8 AND A LARGE AMOUNT OF METAL LINT WAS GENERATED. THE DRILLING WAS CONDUCTED BY FREEHAND. IT IS UNKNOWN IF ANY POTENTIAL METAL DEBRIS ENTERED THE SURGICAL SITE AND IF AN ADDITIONAL INTERVENTION WAS NECESSARY TO RETRIEVE THE METAL LINT. SURGERY WAS FINISHED, AFTER A NON-SIGNIFICANT DELAY. NO HARM TO THE PATIENT OR FURTHER COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15064596 ↗MalfunctionMar 23, 2021Device Reprocessing Problem; Device Reprocessing Problem
IT WAS REPORTED THAT THE LUG HOLE OF THE FINISHING GUIDE HAS METAL SHAVINGS INSIDE FROM DRILL BIT ((B)(4)) AND IS NOT FUNCTIONAL. THIS FAILURE WAS NOTICED DURING PROCEDURE AND THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. NO SURGICAL DELAY OR INJURY TO THE PATIENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11552215 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K190085. One submission often covers many models, so check the product and lot details in each record.
- Z-1393-2023 · Mar 27, 2023 · Open, ClassifiedJOURNEY II UNI XLPE TIBIA INSERT MEDIAL SZ 7-8 8MM
- Z-2132-2020 · Jan 28, 2020 · TerminatedJOURNEY II UNI RESECTION PREP TRIAL, LEFT MEDIAL/ RIGHT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of moti
- Z-2133-2020 · Jan 28, 2020 · TerminatedJOURNEY II UNI RESECTION PREP TRIAL, RIGHT MEDIAL/ LEFT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of moti
Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- LinkSymphoKnee 880-307/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls2 reports
- Scorpio 73-3710Howmedica Osteonics Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls1 report
- Scorpio 73-3110Howmedica Osteonics Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls0 reports
- Arthrex® AR-503-TTTEArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall14 reports
- Arthrex® AR-503-TTTJArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall9 reports
- Arthrex® AR-503-TTTGArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- ENGAGE 1-10011-300Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- ENGAGE 1-10003-400Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗